Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting.

W Buhre1, A Weyland, S Kazmaier, G G Hanekop, M M Baryalei, M Sydow, H Sonntag.   

Abstract

OBJECTIVE: To investigate the precision and accuracy of continuous pulse contour cardiac output (PCCO) compared with intermittent transcardiopulmonary (TCPCO) and pulmonary artery thermodilution (TDCO) measurements in patients undergoing minimally invasive coronary bypass surgery (MIDCAB).
DESIGN: Prospective, controlled, clinical study.
SETTING: University hospital. PARTICIPANTS: Twelve patients undergoing MIDCAB.
INTERVENTIONS: Thirty-six measurements of PCCO and thermodilution cardiac output (CO) were simultaneously performed after the start of surgery, during bypass grafting, and at the end of surgery. TCPCO and TDCO were simultaneously assessed by three injections of ice-cold saline randomly spread over the respiratory cycle. The pulse contour device was initially calibrated with an additional set of aortic thermodilution measurements.
MEASUREMENTS AND MAIN RESULTS: Absolute values of CO ranged between 1.6 and 9.2 L/min. A close agreement among the three techniques was observed at all measurements. Mean bias between PCCO and TDCO and TCPCO was 0.003 L/min (2 SD of differences between methods = 1.26 L/min) and 0.27 L/min (2 SD of differences between methods = 1.16 L/min), respectively. The correlation coefficients were r2 = 0.90 for TCPCO versus PCCO and r2 = 0.88 for TDCO versus PCCO.
CONCLUSION: The results of the present study show that compared with thermodilution CO, pulse contour analysis enables accurate measurement of continuous CO in patients undergoing MIDCAB.

Entities:  

Mesh:

Year:  1999        PMID: 10468257     DOI: 10.1016/s1053-0770(99)90216-1

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  28 in total

1.  Non-contact left ventricular endocardial mapping in cardiac resynchronisation therapy.

Authors:  P D Lambiase; A Rinaldi; J Hauck; M Mobb; D Elliott; S Mohammad; J S Gill; C A Bucknall
Journal:  Heart       Date:  2004-01       Impact factor: 5.994

2.  Comparison between cardiac output measured by the pulmonary arterial thermodilution technique and that measured by the femoral arterial thermodilution technique in a pediatric animal model.

Authors:  M Rupérez; J López-Herce; C García; C Sánchez; E García; D Vigil
Journal:  Pediatr Cardiol       Date:  2003-12-23       Impact factor: 1.655

3.  Methods of monitoring shock.

Authors:  Ednan K Bajwa; Atul Malhotra; B Taylor Thompson
Journal:  Semin Respir Crit Care Med       Date:  2004-12       Impact factor: 3.119

4.  Correlation between cardiac output measured by the femoral arterial thermodilution technique pulmonary arterial and that measured by contour pulse analysis in a paediatric animal model.

Authors:  Jesús López-Herce; Marta Rupérez; César Sánchez; Cristina García; Elena García
Journal:  J Clin Monit Comput       Date:  2006-02-28       Impact factor: 2.502

Review 5.  [Pulmonary artery catheter in anaesthesiology and intensive care medicine].

Authors:  E E C de Waal; L de Rossi; W Buhre
Journal:  Anaesthesist       Date:  2006-06       Impact factor: 1.041

6.  Erroneous continuous cardiac output by calibrated pulse contour analysis.

Authors:  Jörn Grensemann; Frank Wappler; Samir G Sakka
Journal:  J Clin Monit Comput       Date:  2013-04-19       Impact factor: 2.502

Review 7.  Newer methods of cardiac output monitoring.

Authors:  Yatin Mehta; Dheeraj Arora
Journal:  World J Cardiol       Date:  2014-09-26

Review 8.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

9.  [Accuracy of pulse contour cardiac index measurements during changes of preload and aortic impedance].

Authors:  T W Felbinger; M S Goepfert; T Goresch; A E Goetz; D A Reuter
Journal:  Anaesthesist       Date:  2005-08       Impact factor: 1.041

10.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.