The effect of variable arterial transducer level on the accuracy of pulse contour waveform-derived measurements in critically ill patients.

Huai-Wu He1, Da-Wei Liu2, Yun Long1, Xiao-Ting Wang1, Mei-Ling Zhao3, Xiao-Li Lai4.   

Abstract

We know that a 10 cm departure from the reference level of pressure transducer position is equal to a 7.5 mmHg change of invasive hemodynamic pressure monitoring in a fluid-filled system. However, the relationship between the site level of a variable arterial pressure transducer and the pulse contour-derived parameters has yet to be established in critically ill patients. Moreover, the related quantitative analysis has never been investigated. Forty-two critically ill patients requiring PiCCO-Plus cardiac output monitoring were prospectively studied. The phlebostatic axis was defined as the zero reference level; the arterial pressure transducer was then vertically adjusted to different positions (+5, +10, +15, +20, -20, -15, -10, -5 cm) of departure from the zero reference site. The pulse contour waveform-derived parameters were recorded at each position. Elevation of the pressure transducer caused significantly positive changes in the continuous cardiac index (+CCI), stroke volume index (+SVI), and stroke volume variation (+SVV), and negative changes in the rate of left ventricular pressure rise during systole (-dP/dtmax), the systemic vascular resistance index (-SVRI), and vice versa. At the 5 cm position, the SVRI changes reached statistical significance with error. At the 10 cm position, the changes in CCI and dP/dtmax reached statistical significance with error, while the change in SVV reached statistical significance at 15 cm. The change rate of CCI was more than 5 % at the 15 cm position and approximately 10 % at the 20 cm position. On average, for every centimeter change of the transducer, there was a corresponding 0.014 L/min/m(2) CCI change and 0.36 % change rate, a 1.41 mmHg/s dP/dtmax change and 0.13 % change rate, and a 25 dyne/s/cm(5) SVRI change and 1.2 % change rate. The variation of arterial transducer position can result in inaccurate measurement of pulse contour waveform-derived parameters, especially when the transducer's vertical distance is more than 10 cm from the phlebostatic axis. These findings have clinical implications for continuous hemodynamic monitoring.

Entities:  

Keywords:  Arterial blood pressure transducer; Continuous hemodynamic monitoring; Critical care; Phlebostatic axis; Pulse contour waveform-derived measurements

Mesh:

Year:  2015        PMID: 26298710     DOI: 10.1007/s10877-015-9756-x

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  24 in total

1.  Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability.

Authors:  Oliver Gödje; Kerstin Höke; Alwin E Goetz; Thomas W Felbinger; Daniel A Reuter; Bruno Reichart; Reinhard Friedl; Andreas Hannekum; Ulrich J Pfeiffer
Journal:  Crit Care Med       Date:  2002-01       Impact factor: 7.598

2.  Recommendation of a clinical impulse response analysis for catheter calibration-dumping coefficient and natural frequency are incomplete parameters for clinical evaluation.

Authors:  Hiroaki Watanabe; Shin-ichi Yagi; Akiyoshi Namiki
Journal:  J Clin Monit Comput       Date:  2006-03-07       Impact factor: 2.502

Review 3.  Evidence-based practice habits: transforming research into bedside practice.

Authors:  Carol A Rauen; Mary Beth Flynn Makic; Elizabeth Bridges
Journal:  Crit Care Nurse       Date:  2009-04       Impact factor: 1.708

4.  Comparison of direct blood pressure measurement at the radial and dorsalis pedis arteries during surgery in the horizontal and reverse Trendelenburg positions.

Authors:  T Parry; N Hirsch; N Fauvel
Journal:  Anaesthesia       Date:  1995-06       Impact factor: 6.955

5.  Central venous oxygen saturation for the diagnosis of low cardiac output in septic shock patients.

Authors:  A Perner; N Haase; J Wiis; J O White; A Delaney
Journal:  Acta Anaesthesiol Scand       Date:  2009-09-07       Impact factor: 2.105

6.  The precision of PiCCO® measurements in hypothermic post-cardiac arrest patients.

Authors:  T Tagami; S Kushimoto; R Tosa; M Omura; J Hagiwara; H Hirama; H Yokota
Journal:  Anaesthesia       Date:  2012-03       Impact factor: 6.955

7.  Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults.

Authors:  A E Wilson; K Bermingham-Mitchell; N Wells; K Zachary
Journal:  Am J Crit Care       Date:  1996-07       Impact factor: 2.228

8.  Assessment of left ventricular function by pulse wave analysis in critically ill patients.

Authors:  Sabino Scolletta; Laurent Bodson; Katia Donadello; Fabio S Taccone; Alessandro Devigili; Jean-Louis Vincent; Daniel De Backer
Journal:  Intensive Care Med       Date:  2013-03-09       Impact factor: 17.440

9.  Passive leg raising for predicting fluid responsiveness: importance of the postural change.

Authors:  Julien Jabot; Jean-Louis Teboul; Christian Richard; Xavier Monnet
Journal:  Intensive Care Med       Date:  2008-09-16       Impact factor: 17.440

10.  Pulmonary artery pressure measurement in patients with elevated pressures: effect of backrest elevation and method of measurement.

Authors:  K Dobbin; S Wallace; J Ahlberg; M Chulay
Journal:  Am J Crit Care       Date:  1992-09       Impact factor: 2.228

View more
  3 in total

Review 1.  Journal of Clinical Monitoring and Computing 2016 end of year summary: cardiovascular and hemodynamic monitoring.

Authors:  Bernd Saugel; Karim Bendjelid; Lester A Critchley; Steffen Rex; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2017-01-07       Impact factor: 2.502

Review 2.  Passive Leg Raising in Intensive Care Medicine.

Authors:  Huai-Wu He; Da-Wei Liu
Journal:  Chin Med J (Engl)       Date:  2016-07-20       Impact factor: 2.628

3.  The Effect of Changing Arterial Transducer Position on Stroke Volume Measurements Using FloTrac System Version 4.0: A Pilot Experimental Study.

Authors:  Jiro Ito; Daisuke Kawakami; Takeshi Morimoto; Hiroshi Ueta; Takahiro Shimozono; Hiroyuki Mima
Journal:  Crit Care Explor       Date:  2021-06-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.