BACKGROUND: Axial length and anterior chamber depth have been measured clinically using conventional ultrasound biometry. Recently, a non-contact device has become available to measure these parameters. This study evaluated the repeatability and accuracy of this device.
METHODS: The axial length and anterior chamber depth were measured by two practitioners on a group of young subjects using the IOLMaster followed by a conventional ultrasound biometer operated by a third practitioner. The accommodation was controlled in ultrasound biometry through a full correction on the non-measured eye and a distant fixation target.
RESULTS: There was good repeatability and accuracy of axial length assessment. The mean difference between the IOLMaster and ultrasound biometry was -0.099 mm, with 95% limits of agreement between 0.66 and -0.85 mm. The axial length was slightly shorter from the IOLMaster and the difference was not significant. The anterior chamber depth was repeatable but was shown to be deeper than the ultrasound results. The mean difference in anterior chamber depth between the IOLMaster and ultrasound biometry was 0.15 mm, with 95% limits of agreement between 0.34 and -0.03 mm. It is suggested that the former device is not measuring the axial anterior chamber depth.
CONCLUSIONS: The IOLMaster is a non-contact 'optical' A-scan which is simple to use and good for axial length assessment. The anterior chamber depth assessment should be further evaluated.

Mesh:

Year:  2001        PMID: 11727876     DOI: 10.1046/j.1475-1313.2001.00611.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  34 in total

1.  IOLMaster biometry: refractive results of 100 consecutive cases.

Authors:  H Eleftheriadis
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

2.  Lens thickness assessment: anterior segment optical coherence tomography versus A-scan ultrasonography.

Authors:  Nikoo Hamzeh; Sasan Moghimi; Golshan Latifi; Massood Mohammadi; Nassim Khatibi; Shan C Lin
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

3.  Factors affecting refractive outcome after cataract surgery in patients with a history of acute primary angle closure.

Authors:  Hyo Seok Lee; Jung Won Park; Sang Woo Park
Journal:  Jpn J Ophthalmol       Date:  2013-10-29       Impact factor: 2.447

4.  The effects of ocular magnification on Spectralis spectral domain optical coherence tomography scan length.

Authors:  Irene Ctori; Stephen Gruppetta; Byki Huntjens
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-01-10       Impact factor: 3.117

5.  Assessment of eye length changes in accommodation using dynamic extended-depth OCT.

Authors:  Yu-Cherng Chang; Keke Liu; Carolina de Freitas; Alex Pham; Florence Cabot; Siobhan Williams; Ethan Adre; Giovanni Gregori; Marco Ruggeri; Sonia H Yoo; Arthur Ho; Jean-Marie Parel; Fabrice Manns
Journal:  Biomed Opt Express       Date:  2017-04-26       Impact factor: 3.732

6.  Optical Coherence Tomography Predictors of Risk for Progression to Non-Neovascular Atrophic Age-Related Macular Degeneration.

Authors:  Karim Sleiman; Malini Veerappan; Katrina P Winter; Michelle N McCall; Glenn Yiu; Sina Farsiu; Emily Y Chew; Traci Clemons; Cynthia A Toth
Journal:  Ophthalmology       Date:  2017-08-26       Impact factor: 12.079

7.  Axial length measurement using partial coherence interferometry in myopic children: repeatability of the measurement and comparison with refractive components.

Authors:  Shuhei Kimura; Satoshi Hasebe; Manabu Miyata; Ichiro Hamasaki; Hiroshi Ohtsuki
Journal:  Jpn J Ophthalmol       Date:  2007-04-06       Impact factor: 2.447

8.  Changes in macular thickness and depth of anterior chamber in patients after filtration surgery.

Authors:  Gulsara Karasheva; Winfried Goebel; Thomas Klink; Wolfgang Haigis; Franz Grehn
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-02-20       Impact factor: 3.117

9.  Validation of optical coherence tomography-based crystalline lens thickness measurements in children.

Authors:  Bret M Lehman; David A Berntsen; Melissa D Bailey; Karla Zadnik
Journal:  Optom Vis Sci       Date:  2009-03       Impact factor: 1.973

10.  Repeatability of on- and off-axis eye length measurements using the lenstar.

Authors:  Krystal L Schulle; David A Berntsen
Journal:  Optom Vis Sci       Date:  2013-01       Impact factor: 1.973

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.