Agreement between IOLMaster® 500 and Pentacam® HR for keratometry assessment in type 2 diabetic and non-diabetic patients.

João N Beato1,2, João Esteves-Leandro1, David Reis3, Rita Matos1, Manuel Falcão1,2, Vítor Rosas1, Ângela Carneiro1,2, Fernando Falcão-Reis1,2.   

Abstract

AIM: To evaluate inter-device agreement of anterior keratometry obtained by the IOLMaster® 500 and Pentacam® HR in type 2 diabetic and non-diabetic patients.
METHODS: Corneal measurements were sequentially performed in 60 diabetes mellitus (DM) and 48 age and sex-matched controls undergoing cataract surgery. Variables recorded included flat and steep keratometry, mean keratometry (Km), astigmatism magnitude, axis location, J0 and J45 components. Bland-Altman plots and intraclass correlation coefficients were used for examination of agreement. Subgroup analyses were performed for astigmatism magnitude, diabetes duration, hemoglobin A1c (HbA1c) levels and diabetic retinopathy (DR) stage.
RESULTS: Agreement for Km and astigmatism magnitude were considered good and moderate, with 95% limits of agreement (LoA) of -1.09 to 1.23 diopters (D) and -0.83 to 0.86 D in DM group, respectively; and -0.59 to 0.72 D and -0.98 to 0.75 D in non-DM group, respectively. In contrast, the 95% LoA for corneal axis exceeded the clinically relevant margins in both groups. In the total sample, only 41 eyes (38%) had a smaller than 5-degree difference. Diabetes duration, HbA1c levels and DR stage were not found to significantly affect agreement. Logistic regression showed that higher corneal power (P=0.021) and astigmatism magnitude (P=0.011) were associated with a decreased risk of having a difference in axis location greater than 10-degrees.
CONCLUSION: In both groups, IOLMaster and Pentacam agree well for corneal power and moderately for astigmatism. However, axis location disagreement is frequent in eyes with flatter corneas and small amounts of astigmatism. International Journal of Ophthalmology Press.

Entities:  

Keywords:  astigmatism; cataract; diabetes mellitus; diabetic retinopathy; keratometry

Year:  2020        PMID: 32566503      PMCID: PMC7270259          DOI: 10.18240/ijo.2020.06.10

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  2 in total

1.  A comparison of IOLMaster 500 and IOLMaster 700 in the measurement of ocular biometric parameters in cataract patients.

Authors:  Jianhong Jiang; Xiaojing Pan; Mingming Zhou; Xiaoyun Wang; Hai Zhu; Dongfang Li
Journal:  Sci Rep       Date:  2022-07-27       Impact factor: 4.996

2.  Agreement between Two Swept-Source Optical Coherence Tomography Biometers and a Partial Coherence Interferometer.

Authors:  Ji Young Moon; Soo Chang Cho; Hyun Jin Kim; Roo Min Jun; Kyung Eun Han
Journal:  Korean J Ophthalmol       Date:  2022-06-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.