Comparison of different techniques of anterior chamber depth and keratometric measurements.

Uri Elbaz1, Yaniv Barkana, Yariv Gerber, Isaac Avni, David Zadok.   

Abstract

PURPOSE: To compare measurements of anterior chamber depth (ACD) and corneal curvature obtained with the Pentacam to other commonly used devices.
DESIGN: Prospective, nonrandomized clinical trial.
METHODS: Measurements of ACD and keratometry were prospectively obtained in 22 eyes of 11 subjects with the Pentacam compared with measurements of the ultrasound (US) A-scan and IOLMaster (ACD), and with those of automated keratometry (AK) and IOLMaster (keratometry) at the Assaf Harofeh Medical Center.
RESULTS: The mean interdevice differences in keratometry for Pentacam vs AK, Pentacam vs IOLMaster, and AK vs IOLMaster were -0.046 diopters, -0.471 diopters, and -0.424 diopters, respectively. Measurements of keratometry by the IOLMaster differed statistically significantly from those of Pentacam (P < .01) and AK (P < .01). For measurement of keratometry, 95% limits of agreement were -1.321 to 1.229 diopters for the Pentacam and AK, -1.478 to 0.536 diopters for the Pentacam and IOLMaster, and -1.026 to 0.178 diopters for the AK and IOLMaster. The mean interdevice differences in ACD for the Pentacam vs US, Pentacam vs IOLMaster, and US vs IOLMaster were 0.103, 0.099, and -0.004 mm, respectively. Measurements of ACD by the Pentacam differed statistically significantly from those of US (P < .05) and IOLMaster (P < .01). For measurement of ACD, 95% limits of agreement were -0.197 to 0.404 mm for the Pentacam and US, -0.169 to 0.367 mm for the Pentacam and IOLMaster, and -0.332 to 0.324 mm for the US and IOLMaster.
CONCLUSIONS: For some clinical applications, measurements of ACD and corneal curvature with the tested machines may differ greatly and therefore may not be interchangeable.

Entities:  

Mesh:

Year:  2006        PMID: 17101110     DOI: 10.1016/j.ajo.2006.08.031

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  24 in total

1.  Anterior chamber depth in normal subjects by rotating scheimpflug imaging.

Authors:  Matthew T Feng; Michael W Belin; Renato Ambrósio; Satinder P S Grewal; Wang Yan; Mohamed S Shaheen; Charles McGhee; Naoyuki Maeda; Tobias H Neuhann; H Burkhard Dick; Saleh A Alageel; Andreas Steinmueller
Journal:  Saudi J Ophthalmol       Date:  2011-04-30

2.  The effect of head tilt on keratometric measurement using the IOLMaster.

Authors:  C Y Park; J R Do; S H Kim; C-Y Lim; R S Chuck
Journal:  Eye (Lond)       Date:  2013-09-20       Impact factor: 3.775

3.  Keratometry with five different techniques: a study of device repeatability and inter-device agreement.

Authors:  Shiva Mehravaran; Soheila Asgari; Sara Bigdeli; Ashkan Shahnazi; Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2014-02-23       Impact factor: 2.031

4.  Precision of 5 different keratometry devices.

Authors:  Jonas Vejvad Nørskov Laursen; Peter Jeppesen; Thomas Olsen
Journal:  Int Ophthalmol       Date:  2015-03-29       Impact factor: 2.031

Review 5.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

6.  Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer.

Authors:  Leyla Asena; Sirel Gür Güngör; Ahmet Akman
Journal:  Int Ophthalmol       Date:  2016-06-07       Impact factor: 2.031

7.  Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation.

Authors:  Ahmet Kirgiz; Kurşat Atalay; Havva Kaldirim; Kubra Serefoglu Cabuk; Mehmet Orcun Akdemir; Muhittin Taskapili
Journal:  Int Ophthalmol       Date:  2016-09-01       Impact factor: 2.031

8.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

9.  Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers.

Authors:  Ireneusz Grulkowski; Jonathan J Liu; Jason Y Zhang; Benjamin Potsaid; Vijaysekhar Jayaraman; Alex E Cable; Jay S Duker; James G Fujimoto
Journal:  Ophthalmology       Date:  2013-06-04       Impact factor: 12.079

10.  Study of distribution of white-to-white corneal diameter and anterior chamber depth in study population obtained with optical biometry using intraocular lens (IOL) master.

Authors:  Kritika Singh; Sandeep Gupta; P S Moulick; Neeraj Bhargava; Alok Sati; Gagandeep Kaur
Journal:  Med J Armed Forces India       Date:  2018-08-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.