It is important to evaluate the radiation eye dose (3 mm dose equivalent, Hp (3)) received by physicians during computed tomography fluoroscopy (CTF)-guided biopsy, as physicians are close to the source of scattered radiation. In this study, we measured the radiation eye dose in Hp (3) received by one physician during CTF in a timeframe of 18 months using a direct eye dosimeter, the DOSIRISTM. The physician placed eye dosimeters above and under their lead (Pb) eyeglasses. We recorded the occupational radiation dose received using a neck dosimeter, gathered CT dose-related parameters (e.g., CT-fluoroscopic acquisition number, CT-fluoroscopic time, and CT-fluoroscopic mAs), and performed a total of 95 procedures during CTF-guided biopsies. We also estimated the eye dose (Hp (3)) received using neck personal dosimeters and CT dose-related parameters. The physician eye doses (right and left side) received in terms of Hp (3) without the use of Pb eyeglasses for 18 months were 2.25 and 2.06 mSv, respectively. The protective effect of the Pb eyeglasses (0.5 mm Pb) on the right and left sides during CTF procedures was 27.8 and 37.5%, respectively. This study proved the existence of significant correlations between the eye and neck dose measurement (right and left sides, R2 = 0.82 and R2 = 0.55, respectively) in physicians. In addition, we found significant correlations between CT-related parameters, such as CT-fluoroscopy mAs, and radiation eye doses (right and left sides, R2 = 0.50 and R2 = 0.52, respectively). The eye dose of Hp (3) received in CTF was underestimated when evaluated using neck dosimeters. Therefore, we suggest that the physician involved in CTF use a direct eye dosimeter such as the DOSIRIS for the accurate evaluation of their eye lens dose.Entities:
Keywords: CTF-guided biopsy; eye lens; radiation eye dose; radiation protection glasses; scattered radiation
Mesh:
Year: 2022 PMID: 35202201 PMCID: PMC8878526 DOI: 10.3390/tomography8010036
Source DB: PubMed Journal: Tomography ISSN: 2379-1381
Figure 1(a) The physician wore a protective apron (0.35 mm Pb equivalent), neck guard (0.25 mm Pb equivalent), and Pb glasses during CTF, as well as (b) eye dosimeters (DOSIRIS) above and under the Pb eyeglasses.
Summary of our 18-month study on patients undergoing CTF-guided biopsies (95 procedures).
| Characteristics | Ave. ± SD | Median | Range |
|---|---|---|---|
| Age (years) | 51.8 ± 20.8 | 55.5 | 5.0–88.0 |
| Body Mass Index | 23.2 ± 3.9 | 22.7 | 15.8–34.3 |
| CT-fluoroscopic acquisition number | 26.2 ± 15.5 | 22.0 | 8.0–96.0 |
| CT-fluoroscopic time (s) | 20.0 ± 11.8 | 17.1 | 5.9–72.1 |
| CT-fluoroscopic mAs | 464.1 ± 327.4 | 376.0 | 118.0–2105.0 |
| CTDI vol (mGy) | 10.2 ± 5.3 | 7.7 | 3.3–29.6 |
| DLP (mGy*cm) | 215.1 ± 88.5 | 186.7 | 76.2–539.9 |
| Target depth (mm) | 60.5 ± 21.6 | 56.5 | 14.3–119.5 |
Ave.: average; SD: standard deviation; CTDI: CT dose index; DLP: dose length product.
Total physician doses (eye and neck) received in 18 months during CTF procedures (95 procedures).
| (μSv) | Right Side | Left Side |
|---|---|---|
| Eye dose (DOSIRIS), Hp (3) | 2250 | 2060 |
| Neck dose (Glass Badge), Hp (3) | 1160 | 1020 |
Protective effect of Pb eyeglasses used for 9 months during CTF procedures (47 procedures).
| (μSv) | Outside of Eyeglasses | Inside of Eyeglasses | Protective Effect (%) |
|---|---|---|---|
| Right eye dose (DOSIRIS), Hp (3) | 720 | 520 | 27.8 |
| Left eye dose (DOSIRIS), Hp (3) | 800 | 500 | 37.5 |
Figure 2Relationship between the eye dosimeter (DOSIRIS, Hp (3)) and neck dosimeter (Glass Badge, Hp (3)), taken monthly during CTF-guided biopsies (95 procedures). (a) Right side dosimeter; (b) left side dosimeter.
Figure 3Relationship between the eye dosimeter measurements (DOSIRIS, Hp (3)) inside and outside the Pb eyeglasses, taken monthly during CTF-guided biopsies (47 procedures). (a) Right side dosimeter; (b) left side dosimeter.
Determination coefficient (R2) between physician doses and CT dose-related parameters (95 procedures).
| Determination Coefficient (R2) | CT-Acquisitions No. | CT-Fluoroscopic Time (s) | CT-Fluoroscopic mAs |
|---|---|---|---|
| Right eye dose (DOSIRIS), Hp (3) | 0.463 | 0.441 | 0.497 |
| Left eye dose (DOSIRIS), Hp (3) | 0.507 | 0.486 | 0.524 |
| Right neck dose (Glass Badge), Hp (3) | 0.788 | 0.799 | 0.745 |
| Left neck dose (Glass Badge), Hp (3) | 0.666 | 0.675 | 0.592 |