Patient and personnel exposure during CT fluoroscopy-guided interventional procedures.

R D Nawfel1, P F Judy, S G Silverman, S Hooton, K Tuncali, D F Adams.   

Abstract

PURPOSE: To estimate patient dose and personnel exposure from phantom measurements during computed tomographic (CT) fluoroscopy, to use the estimates to provide users with dose information, and to recommend methods to reduce exposure.
MATERIALS AND METHODS: Surface dose was estimated on a CT dosimetric phantom by using thermoluminescent dosimetric (TLD) and CT pencil chamber measurements. Scatter exposure was estimated from scattered radiation measured at distances of 10 cm to 1 m from the phantom. Scatter exposures measured with and without placement of a lead drape on the phantom surface adjacent to the scanning plane were compared.
RESULTS: Phantom surface dose rates ranged from 2.3 to 10. 4 mGy/sec. Scattered exposure rates for a commonly used CT fluoroscopic technique (120 kVp, 50 mA, 10-mm section thickness) were 27 and 1.2 microGy/sec at 10 cm and 1 m, respectively, from the phantom. Lead drapes reduced the scattered exposure by approximately 71% and 14% at distances of 10 and 60 cm from the scanning plane, respectively.
CONCLUSION: High exposures to patients and personnel may occur during CT fluoroscopy-guided interventions. Radiation exposure to patients and personnel may be reduced by modifying CT scanning techniques and by limiting fluoroscopic time. In addition, scatter exposure to personnel may be substantially reduced by placing a lead drape adjacent to the scanning plane.

Entities:  

Mesh:

Year:  2000        PMID: 10887246     DOI: 10.1148/radiology.216.1.r00jl39180

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  45 in total

1.  Robotically driven interventions: a method of using CT fluoroscopy without radiation exposure to the physician.

Authors:  Stephen B Solomon; Alexandru Patriciu; Mark E Bohlman; Louis R Kavoussi; Dan Stoianovici
Journal:  Radiology       Date:  2002-10       Impact factor: 11.105

2.  Cost comparison analysis of low-field (0.23 T) MRI- and CT-guided bone biopsies.

Authors:  J Alanen; L Keski-Nisula; R Blanco-Sequeiros; O Tervonen
Journal:  Eur Radiol       Date:  2003-06-25       Impact factor: 5.315

3.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

4.  Counterpoint: Conventional Fluoroscopy-Guided Selective Cervical Nerve Root Block-A Safe, Effective, and Efficient Modality in the Hands of an Experienced Proceduralist.

Authors:  F W Ott; R Pluhm; K Ozturk; A M McKinney; J B Rykken
Journal:  AJNR Am J Neuroradiol       Date:  2020-06       Impact factor: 3.825

5.  Selective lumbar nerve root blocks with CT fluoroscopic guidance: technique, results, procedure time, and radiation dose.

Authors:  Andrew L Wagner
Journal:  AJNR Am J Neuroradiol       Date:  2004-10       Impact factor: 3.825

6.  Respiratory motion compensation with tracked internal and external sensors during CT-guided procedures.

Authors:  Jörn Borgert; S Krüger; H Timinger; J Krücker; N Glossop; A Durrani; A Viswanathan; B J Wood
Journal:  Comput Aided Surg       Date:  2006-05

7.  Cost comparison of low-field (0.23 T) MRI-guided laser ablation and surgery in the treatment of osteoid osteoma.

Authors:  J Ronkainen; R Blanco Sequeiros; O Tervonen
Journal:  Eur Radiol       Date:  2006-04-25       Impact factor: 5.315

Review 8.  Computed tomography-guided percutaneous abscess drainage in coloproctology: review of the literature.

Authors:  R Golfieri; A Cappelli
Journal:  Tech Coloproctol       Date:  2007-08-03       Impact factor: 3.781

9.  [Percutaneous liver biopsy. Overview of different techniques].

Authors:  J Kettenbach; M Blum; K El-RaBadi; H Langenberger; B Happel; J Berger; A Ba-Ssalamah
Journal:  Radiologe       Date:  2005-01       Impact factor: 0.635

10.  CT fluoroscopy: novel application for the treatment of ventricular pathologies.

Authors:  Kieran Murphy; David A Nussbaum; Philippe Gailloud
Journal:  Neuroradiology       Date:  2007-02-16       Impact factor: 2.804

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.