Development of a New Radiation Shield for the Face and Neck of IVR Physicians.

Toshimitsu Sato1, Yoichi Eguchi2, Chika Yamazaki1, Takanobu Hino1, Toshikazu Saida3, Koichi Chida2,4.   

Abstract

Interventional radiology (IVR) procedures are associated with increased radiation exposure and injury risk. Furthermore, radiation eye injury (i.e., cataract) in IVR staff have also been reported. It is crucial to protect the eyes of IVR physicians from X-ray radiation exposure. Many IVR physicians use protective Pb eyeglasses to reduce occupational eye exposure. However, the shielding effects of Pb eyeglasses are inadequate. We developed a novel shield for the face (including eyes) of IVR physicians. The novel shield consists of a neck and face guard (0.25 mm Pb-equivalent rubber sheet, nonlead protective sheet). The face shield is positioned on the left side of the IVR physician. We assessed the shielding effects of the novel shield using a phantom in the IVR X-ray system; a radiophotoluminescence dosimeter was used to measure the radiation exposure. In this phantom study, the effectiveness of the novel device for protecting against radiation was greater than 80% in almost all measurement situations, including in terms of eye lens exposure. A large amount of scattered radiation reaches the left side of IVR physicians. The novel radiation shield effectively protects the left side of the physician from this scattered radiation. Thus, the device can be used to protect the face and eyes of IVR physicians from occupational radiation exposure. The novel device will be useful for protecting the face (including eyes) of IVR physicians from radiation, and thus could reduce the rate of radiation injury. Based on the positive results of this phantom study, we plan to perform a clinical experiment to further test the utility of this novel radiation shield for IVR physicians.

Entities:  

Keywords:  X-ray examination; disaster medicine; face shield; fluoroscopically guided interventional procedures; fluoroscopy; interventional radiology (IVR); percutaneous coronary intervention (PCI); protective apron; radiation dose; radiation protection and safety

Year:  2022        PMID: 36004878      PMCID: PMC9404996          DOI: 10.3390/bioengineering9080354

Source DB:  PubMed          Journal:  Bioengineering (Basel)        ISSN: 2306-5354


1. Introduction

Interventional radiology (IVR) procedures are increasingly being performed because of the significant advantages for patients [1,2,3,4,5,6]. However, IVR procedures are associated with increased radiation exposure and injury risk in both patients and IVR staff [7,8,9,10]. Many studies have evaluated the radiation dose to patients and IVR staff, and methods to reduce exposure [11,12,13,14,15,16,17,18]. We also previously evaluated exposure of patients and staff to radiation in our IVR laboratory [19,20,21,22,23,24,25,26]. In 2011, the International Commission on Radiological Protection significantly reduced the limit of occupational exposure of the eyes to radiation, from 150 to 100 mSv/5 years (i.e., 20 mSv/year) [27]. Furthermore, radiation eye injury (i.e., cataract) in IVR staff has also been reported [28,29]. It is crucial to protect the eyes of IVR physicians from X-ray radiation exposure [30,31,32,33,34,35]. Therefore, evaluation of the exposure of the eyes of IVR physicians to occupational radiation, and related protection, is important [36,37,38,39,40]. Lead (Pb) eyeglasses are useful for shielding the eyes against radiation [41,42,43]. Many IVR physicians use protective Pb eyeglasses to reduce occupational eye exposure. Despite the diversity in the thickness and shape of Pb eyeglasses, none offer complete protection against radiation exposure to the eyes of IVR physicians [44,45,46]. Therefore, we developed a unique face radiation shield that also protects the eyes. The device was designed to protect the neck and the left side of the face, including the left eye, of IVR physicians. The purpose of this phantom study was to evaluate the radiation-protective effects of the novel shield in an IVR X-ray system.

2. Materials and Methods

2.1. Development of the Novel Radiation Shield

Figure 1 shows the novel radiation shield for IVR physicians. The device consists of a neck guard and face shield designed using a 0.25 mm Pb-equivalent rubber sheet (nonlead protective sheet, Figure 2). Pb-equivalent rubber sheeting is easy to handle and often used in personal protective aprons. The device is lightweight (0.65 kg). The neck guard and face shield are firmly connected and have adequate stability. The face shield was designed to mainly protect the left side of IVR physicians from scattered radiation.
Figure 1

Photograph of the novel shield: (a) frontal view; (b) lateral view. The neck guard and face shield are fastened together to create a single device, which cannot be disassembled. The face shield is attached to the left side of the neck guard and protects the left side of the physicians’ neck and face from radiation. The novel shield was designed so that it does not obstruct IVR physicians’ field of vision.

Figure 2

Schematic of the novel shield: (a) frontal view; (b) lateral view. The novel shield consists of a neck guard and face shield, which together comprise a single unit to promote stability and prevent misalignment. The shield is firmly attached behind the neck of the IVR physicians using Velcro to protect them from scattered radiation from the left side. The device consists of a neck guard and face shield designed using a 0.25 mm Pb-equivalent rubber sheet (nonlead protective sheet).

2.2. Phantom Study

We conducted a phantom study at Yamagata University Hospital, Japan. Figure 3 displays the experimental setup used to simulate the typical settings for IVR procedures.
Figure 3

Experimental setup used for our phantom study (e.g., LAO60): (a) without novel shielding device; (b) with novel shielding device.

A digital cine angiography X-ray unit (an “under-tube” X-ray tube system) with a 16-in mode flat-panel detector (FPD) was used. Digital cine acquisitions were performed at 30 frames/s with a total duration of 150 s (30 s × 5). An automatic control system was used to set the X-ray exposure settings (i.e., kilovoltage and milliamperage) (Table 1).
Table 1

X-ray exposure setup used in our study.

Tube-Viewing AnglesTube Kilovoltage(kV)Tube Milliamperage (mA)Additional Copper Filter (mm)
60° left anterior oblique743200.3
30° right anterior oblique833200.3
Posteroanterior743200.3
60° left anterior oblique+30° craniocaudal744000.3
30° right anterior oblique+30° caudocranial793200.3
We set the focus-to-image receptor (i.e., FPD) distance to 120 cm, and the height of the patient table to 92 cm. Five standard tube-viewing angles were used to simulate the typical settings for percutaneous coronary intervention (PCI) and cardiac catheterization: posteroanterior (PA), 60° left anterior oblique (LAO), 30° right anterior oblique (RAO), 30° RAO + 30° caudocranial (cranial), and 60° LAO + 30° craniocaudal (caudal). A trunk phantom (PBU-60) was used to simulate the patient (Figure 3). A head phantom (THRA1) was used to simulate the IVR physician (Figure 3); it was placed 70 cm horizontally and 40 cm vertically from the central radiation beam on the patient table. This position is similar to that used by physicians during PCI at our hospital. The height of the head phantom was 165 cm; therefore, the eye of the phantom was approximately 150 cm above the floor. We did not use a ceiling-protecting Pb plate.

2.3. Dosimetry

Scattered radiation from the trunk phantom representing the patient was measured using radiophotoluminescence dosimeters (RPLDs; GD-302M), with and without the novel radiation shield. Dose Ace FGD-1000 was used as the measurement/readout system. RPLDs were placed on the surface of the head phantom representing the physician at 24 locations, including the left (No. ③) and right (No. ㉑) eyes (Figure 4).
Figure 4

The 24 measurement points on the head of the phantom simulating the physician: (a) frontal view; (b) left lateral view; (c) right lateral view. Twenty-four dosimeters were attached to the points marked on the phantom’s surface (left eye: No. ③, right eye: No. ㉑). The distance between the measurement points was 3 cm.

The background radiation dose was subtracted from the measurements, and the doses were calibrated. The average of three measurements was recorded for each X-ray viewing angle. Based on the doses measured with (Dwith) and without (Dwithout) the novel radiation shield, we calculated the effectiveness of the radiation protection of the shield as: (Dwithout − Dwith)/Dwithout × 100%.

3. Results

Table 2 summarizes the results of our phantom study of the novel radiation shield. The scattered radiation doses were highest and lowest for the LAO views (LAO 60° and LAO 60° + CAU 30°) and RAO views (RAO 30° and RAO 30° + CRA 30°), respectively, for all measurements acquired without the novel radiation shield.
Table 2

Summary of the phantom study.

Posteroanterior60° Left Anterior Oblique30° Right Anterior Oblique60° Left Anterior Oblique +30° Craniocaudal30° Right Anterior Oblique +30° Caudocranial
1 MP2 Without3 With 4 PE 2 Without3 With 4 PE 2 Without3 With 4 PE 2 Without3 With 4 PE 2 Without3 With 4 PE
(μGy)(μGy) (%) (μGy)(μGy) (%) (μGy)(μGy) (%) (μGy)(μGy) (%) (μGy)(μGy) (%)
6918671 90.3 12,6001764 86.0 3058786 74.3 15,5121250 91.9 4147673 83.8
7673716 90.7 13,4511894 85.9 3207809 74.8 16,4601352 91.8 4289591 86.2
8065639 92.1 14,4701861 87.1 3442785 77.2 17,5851319 92.5 4495508 88.7
7821706 91.0 13,5101839 86.4 3304770 76.7 15,7991283 91.9 4491653 85.5
8320718 91.4 14,1751867 86.8 3490778 77.7 16,6141284 92.3 4568587 87.1
8473640 92.4 15,1451917 87.3 3645760 79.2 17,8211304 92.7 4786484 89.9
8368658 92.1 13,7961640 88.1 3342675 79.8 16,1851215 92.5 4840636 86.9
8686620 92.9 14,5101596 89.0 3637657 81.9 17,1031161 93.2 4865516 89.4
9096588 93.5 16,6671789 89.3 4038674 83.3 19,2161281 93.3 5168421 91.9
8074592 92.7 13,7571558 88.7 3167566 82.1 15,8331159 92.7 4621574 87.6
8533590 93.1 14,8001569 89.4 3620609 83.2 17,0661088 93.6 4924490 90.1
9534573 94.0 16,9661777 89.5 4080699 82.9 19,2781216 93.7 5319449 91.6
7731569 92.6 13,2582080 84.3 2879561 80.5 14,9371819 87.8 4509563 87.5
8406581 93.1 14,7642477 83.2 3274620 81.1 15,8371771 88.8 4664547 88.3
8772530 94.0 15,9163059 80.8 3730629 83.1 17,3481869 89.2 4968450 90.9
6353627 90.1 11,2671555 86.2 2714847 68.8 13,6991130 91.7 3735680 81.8
6669618 90.7 12,3381634 86.8 2875757 73.7 14,6771153 92.1 3765553 85.3
6283607 90.3 12,9981733 86.7 2822769 72.8 15,0901228 91.9 3891504 87.0
5652610 89.2 95151392 85.4 2470968 60.8 12,0941025 91.5 3380726 78.5
5536558 89.9 10,0101385 86.2 2585763 70.5 11,758971 91.7 3387556 83.6
3563393 89.0 6026940 84.4 1905568 70.2 6548640 90.2 2493400 84.0
1498298 80.1 1693631 62.7 1050610 41.9 1827439 76.0 1346458 65.9
3183369 88.4 4071754 81.5 1782605 66.0 4146510 87.7 2155434 79.8
4118405 90.2 6742935 86.1 2298611 73.4 6144616 90.0 2778438 84.2

1 MP: Measurement point. 2 Without: Doses measured without the novel radiation shield (The average of three measurements was recorded). 3 With: Doses measured with the novel radiation shield (The average of three measurements was recorded). 4 PE: Protective effect, (Dwithout − Dwith)/Dwithout × 100%.

The scattered radiation doses were higher for the left side (No. ①–⑮) compared to the right side (No. ⑲–㉔) of the face. Figure 5 depicts the protective effect of the novel radiation shield. The radiation protection effectiveness of the novel radiation shield was greater than 80% at almost all measurement points, except RAO 30°, at which the effectiveness was slightly lower. The average radiation protection effectiveness of the novel device for the five viewing angles were 87.5% and 83.6% for the left (No. ③) and right (No. ㉑) eyes, respectively.
Figure 5

Protective effect of the novel radiation shield in the phantom study. Measurement point: Twenty-four dosimeters were attached to the points marked on the phantom’s surface (left eye: No. ③, right eye: No. ㉑) (See Figure 4). PA: posteroanterior, LAO60: 60° left anterior oblique, RAO30: 30° right anterior oblique, LAO60+CAU30: 60° left anterior oblique + 30° craniocaudal (caudal), RAO30+CRA30: 30° right anterior oblique + 30° caudocranial (cranial).

4. Discussion

It is crucial to evaluate exposure of patients and healthcare workers to radiation during radiological examinations, especially IVR [47,48,49,50,51,52]. Despite the importance of protecting IVR physicians from occupational radiation exposure, no ideal radiation shield exists [53,54,55,56]. Although many devices protecting against radiation are available, none offer complete protection, especially for IVR physicians [54,57,58]. We developed a novel radiation shield to protect the face of IVR physicians (Figure 1 and Figure 2). This device is lightweight and comfortable to wear and has a unique design that protects the face (including the eyes) of IVR physicians. To provide stability and prevent misalignment, the face shield is firmly connected to the neck guard as a single component (Figure 1 and Figure 2). The device also allows IVR physicians to have a full field of vision. The face shield is connected to the left side of the face because most occupational radiation exposure to IVR physicians occurs from that side. At almost all measurement points, the radiation protection of the shield was greater than 80%, which confirms its usefulness for IVR physicians. However, slightly lower effectiveness (<80%) was observed for the RAO view and No. ㉒. Thus, the protective effects of the device were slightly reduced in the RAO view compared to the other views. However, compared to the left side of the face, the doses of radiation delivered to the right side are nonetheless small, such that the device would still be effective for protecting IVR physicians from occupational radiation exposure. Similarly, the protection at No. ㉑ (right eye) was relatively low (i.e., 70.5%) at RAO30; however, this is unlikely to be a problem because the radiation doses delivered to this area are also small. Radiation exposure to physicians is greater in the LAO compared to the RAO view because of the higher levels of scattered radiation (from the patient to the physician) in the former view. Occupational radiation protection of the eyes is crucial for IVR physicians, and Pb eyeglasses are often used for this purpose. Lightweight and comfortable Pb eyeglasses (0.07 mm Pb-equivalent) are often preferred by IVR physicians because of the prolonged duration of IVR procedures. However, the radiation-shielding effect of 0.07 mm Pb-equivalent eyeglasses is inadequate (45–60%). Although the radiation-shielding effect of 0.75 mm Pb-equivalent eyeglasses (~80%) is superior to that of 0.07 mm Pb-equivalent eyeglasses, the latter glasses are heavy and uncomfortable, which makes them unsuitable for use by IVR physicians. Our novel shield provides eye radiation protection of above 80% on average (left eye, ③: 87.5%, right eye, ㉑: 83.6%), which is superior to that of Pb eyeglasses. Generally, the distance between the left side of the IVR physician and the scattered radiation source (i.e., the patient) is small, such that more scattered radiation is received by the left than the right side of the physician [42,43,59]. Therefore, our novel radiation shield was developed to protect the left side of the IVR physician’s head. IVR physicians are also potentially at higher risk of radiation-induced brain tumors compared to the general population [60,61,62]. Roguin et al. reported a higher rate of tumors on the left compared to the right side of the brain in IVR physicians, which they attributed to the higher radiation dose to the left side of the head (because it is nearer to the primary X-ray beam and exposed to more scattered radiation) [63]. The novel shield was designed to protect particularly the left side of the head of IVR physicians, and thus may reduce the risk of radiation-induced brain tumors. Currently, the novel shield is available only in a single size; small and large sizes may also be needed. The novel shield protects only the face and neck of IVR physicians. Therefore, other radiation shields (e.g., a protective apron) are also required. Further studies comparing the eye-protective effect of our novel radiation shield with that of protective Pb glasses (using the same radioactive source in the same environment) may be needed. This study using phantoms introduces our novel shield for the face and neck of IVR physicians, but further investigation is required in clinical settings to fully test the shield.

5. Conclusions

We performed a phantom study to investigate the protective effects against radiation of a novel shield for the face and eyes of IVR physicians and found it to be highly effective (>80% protection) under almost all measurement conditions. The novel shield can reduce the radiation dose by more than 80% without the use of Pb eyeglasses and offers equivalent or superior protection compared to Pb eyeglasses. A large amount of scattered radiation reaches the left side of IVR physicians. The novel radiation shield effectively protects the left side of the physician from this scattered radiation. Thus, the device can be used to protect the face and eyes of IVR physicians from occupational radiation exposure. The novel device will be useful for protecting the face (including eyes) of IVR physicians from radiation, and thus could reduce the rate of radiation injury. Based on the positive results of this phantom study, we plan to perform a clinical experiment to further test the utility of this novel radiation shield for IVR physicians.
  60 in total

1.  Invasive Cardiologists Are Exposed to Greater Left Sided Cranial Radiation: The BRAIN Study (Brain Radiation Exposure and Attenuation During Invasive Cardiology Procedures).

Authors:  Ryan R Reeves; Lawrence Ang; John Bahadorani; Jesse Naghi; Arturo Dominguez; Vachaspathi Palakodeti; Sotirios Tsimikas; Mitul P Patel; Ehtisham Mahmud
Journal:  JACC Cardiovasc Interv       Date:  2015-08-17       Impact factor: 11.195

2.  Diagnostic reference levels and achievable doses for common computed tomography examinations: Results from the Japanese nationwide dose survey.

Authors:  Yuta Matsunaga; Koichi Chida; Yuya Kondo; Kenichi Kobayashi; Masanao Kobayashi; Kazuyuki Minami; Shoichi Suzuki; Yasuki Asada
Journal:  Br J Radiol       Date:  2018-10-24       Impact factor: 3.039

3.  The necessity of follow-up for radiation skin injuries in patients after percutaneous coronary interventions: radiation skin injuries will often be overlooked clinically.

Authors:  Mamoru Kato; Koichi Chida; Tadaya Sato; Hajime Oosaka; Tetsuo Tosa; Masahito Munehisa; Ken Kadowaki
Journal:  Acta Radiol       Date:  2012-09-27       Impact factor: 1.990

4.  OCCUPATIONAL RADIATION EXPOSURE DOSE OF THE EYE IN DEPARTMENT OF CARDIAC ARRHYTHMIA PHYSICIAN.

Authors:  Mamoru Kato; Koichi Chida; Takato Ishida; Fumiaki Sasaki; Hideto Toyoshima; Hajime Oosaka; Ken Terata; Yoshihisa Abe; Toshibumi Kinoshita
Journal:  Radiat Prot Dosimetry       Date:  2019-12-31       Impact factor: 0.972

5.  Brain and neck tumors among physicians performing interventional procedures.

Authors:  Ariel Roguin; Jacob Goldstein; Olivier Bar; James A Goldstein
Journal:  Am J Cardiol       Date:  2013-02-16       Impact factor: 2.778

Review 6.  Eye protection in interventional procedures.

Authors:  Beth A Schueler; Kenneth A Fetterly
Journal:  Br J Radiol       Date:  2021-10-01       Impact factor: 3.629

Review 7.  Optimizing patient radiation dose in intervention procedures.

Authors:  Koichi Chida; Mamoru Kato; Haruo Saito; Tadashi Ishibashi; Shoki Takahashi; Masahiro Kohzuki; Masayuki Zuguchi
Journal:  Acta Radiol       Date:  2010-02       Impact factor: 1.990

8.  Randomized Controlled Trial of Radiation Protection With a Patient Lead Shield and a Novel, Nonlead Surgical Cap for Operators Performing Coronary Angiography or Intervention.

Authors:  Ashraf Alazzoni; Chris L Gordon; Jaffer Syed; Madhu K Natarajan; Michael Rokoss; Jon-David Schwalm; Shamir R Mehta; Tej Sheth; Nicholas Valettas; James Velianou; Shaheen Pandie; Darar Al Khdair; Michael Tsang; Brandi Meeks; Kiersten Colbran; Ed Waller; Shun Fu Lee; Tamara Marsden; Sanjit S Jolly
Journal:  Circ Cardiovasc Interv       Date:  2015-08       Impact factor: 6.546

9.  Evaluation of Radiation Protection Methods for Assistant Staff during CT Imaging in High-energy Trauma: Lens Dosimetry with a Phantom Study.

Authors:  Joji Ota; Hajime Yokota; Tatsuya Kawasaki; Junichi Taoka; Hideyuki Kato; Koichi Chida; Yoshitada Masuda; Takashi Uno
Journal:  Health Phys       Date:  2021-06-01       Impact factor: 1.316

10.  Occupational eye dose correlation with neck dose and patient-related quantities in interventional cardiology procedures.

Authors:  Hiroki Ishii; Koichi Chida; Ko Satsurai; Yoshihiro Haga; Yuji Kaga; Mitsuya Abe; Yohei Inaba; Masayuki Zuguchi
Journal:  Radiol Phys Technol       Date:  2022-01-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.