Radiation dose of interventional radiology system using a flat-panel detector.

Koichi Chida1, Yohei Inaba, Haruo Saito, Tadashi Ishibashi, Shoki Takahashi, Masahiro Kohzuki, Masayuki Zuguchi.   

Abstract

OBJECTIVE: Currently, cardiac interventional radiology equipment has tended toward using flat-panel detectors (FPDs) instead of image intensifiers (IIs) because FPDs offer better imaging performance. However, the radiation dose from an FPD in cardiac interventional radiology is not clear. The purpose of our study was to measure the radiation doses during cineangiography and fluoroscopy of many cardiac radiology systems that use FPDs or IIs, in clinical settings.
MATERIALS AND METHODS: This study examined 20 radiology systems in 15 cardiac catheterization laboratories (11 used FPD and nine used II). The entrance surface doses with digital cineangiography and fluoroscopy were compared for the 20 systems using acrylic plates (20-cm thick) and a skin dose monitor.
RESULTS: For fluoroscopy, the average entrance surface doses of the 20-cm-thick acrylic plates were identical for FPD (average +/- SD, 16.63 +/- 7.89 mGy/min; range, 5.7-26.4 mGy/min; maximum/minimum, 4.63) and II (17.81 +/- 12.52 mGy/min; range, 6.5-42.2 mGy/min; maximum/minimum, 6.49) (p = 0.799). For digital cineangiography, the average entrance surface dose of the 20-cm-thick acrylic plate was slightly lower with FPD (29.68 +/- 16.40 mGy/10 s; range, 8.9-58.5 mGy/10 s; maximum/minimum, 6.57) than with II (38.50 +/- 33.71 mGy/10 s; range, 15.2-117.1 mGy/10 s; maximum/minimum, 7.70), although the difference was not significant (p = 0.487).
CONCLUSION: We found that the average entrance doses of cineangiography and fluoroscopy in FPD systems were not significantly different from those in II systems. Hence, FPDs did not inherently reduce the radiation dose, although FPDs possess good detective quantum efficiency. Therefore, to reduce the radiation dose of cardiac interventional radiology systems, even FPD systems, practical measures are necessary.

Mesh:

Year:  2009        PMID: 19933664     DOI: 10.2214/AJR.09.2747

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  22 in total

1.  Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children.

Authors:  Meike Weis; Claudia Hagelstein; Theo Diehm; Stefan O Schoenberg; K Wolfgang Neff
Journal:  Pediatr Radiol       Date:  2016-02

2.  A Rotatable Quality Control Phantom for Evaluating the Performance of Flat Panel Detectors in Imaging Moving Objects.

Authors:  Yoshihiro Haga; Koichi Chida; Yohei Inaba; Yuji Kaga; Taiichiro Meguro; Masayuki Zuguchi
Journal:  J Digit Imaging       Date:  2016-02       Impact factor: 4.056

3.  Comparison of dose at an interventional reference point between the displayed estimated value and measured value.

Authors:  Koichi Chida; Yohei Inaba; Yoshiaki Morishima; Masaaki Taura; Ayako Ebata; Isao Yanagawa; Ken Takeda; Masayuki Zuguchi
Journal:  Radiol Phys Technol       Date:  2011-06-04

4.  Effectiveness of a novel real-time dosimeter in interventional radiology: a comparison of new and old radiation sensors.

Authors:  Yohei Inaba; Masaaki Nakamura; Koichi Chida; Masayuki Zuguchi
Journal:  Radiol Phys Technol       Date:  2018-10-10

5.  The effectiveness of additional lead-shielding drape and low pulse rate fluoroscopy in protecting staff from scatter radiation during cardiac resynchronization therapy (CRT).

Authors:  Yoshiaki Morishima; Koichi Chida; Yoshiaki Katahira
Journal:  Jpn J Radiol       Date:  2018-10-15       Impact factor: 2.374

6.  Comprehensive assessment of patient image quality and radiation dose in latest generation cardiac x-ray equipment for percutaneous coronary interventions.

Authors:  Amber J Gislason-Lee; Claire Keeble; Daniel Egleston; Josephine Bexon; Stephen M Kengyelics; Andrew G Davies
Journal:  J Med Imaging (Bellingham)       Date:  2017-05-02

Review 7.  Coronary imaging techniques with emphasis on CT and MRI.

Authors:  Mathieu Lederlin; Jean-Benoit Thambo; Valérie Latrabe; Olivier Corneloup; Hubert Cochet; Michel Montaudon; François Laurent
Journal:  Pediatr Radiol       Date:  2011-11-30

8.  Investigation of reference levels and radiation dose associated with abdominal EVAR (endovascular aneurysm repair) procedures across several European Centres.

Authors:  E Tuthill; L O'Hora; M O'Donohoe; S Panci; P Gilligan; D Campion; R Trenti; E Fox; D Catania; L Rainford
Journal:  Eur Radiol       Date:  2017-05-18       Impact factor: 5.315

9.  Effectiveness of a New Lead-Shielding Device and Additional Filter for Reducing Staff and Patient Radiation Exposure During Videofluoroscopic Swallowing Study Using a Human Phantom.

Authors:  Yoshiaki Morishima; Koichi Chida; Yoshikazu Muroya; Yoshiya Utsumi
Journal:  Dysphagia       Date:  2017-09-18       Impact factor: 3.438

10.  Impact of latest generation cardiac interventional X-ray equipment on patient image quality and radiation dose for trans-catheter aortic valve implantations.

Authors:  Amber J Gislason-Lee; Claire Keeble; Christoper J Malkin; Daniel Egleston; Josephine Bexon; Stephen M Kengyelics; Daniel Blackman; Andrew G Davies
Journal:  Br J Radiol       Date:  2016-09-29       Impact factor: 3.039

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.