CT fluoroscopy shielding: decreases in scattered radiation for the patient and operator.

Ziv Neeman1, Sergio A Dromi, Shawn Sarin, Bradford J Wood.   

Abstract

PURPOSE: High-radiation exposure occurs during computed tomographic (CT) fluoroscopy. Patient and operator doses during thoracic and abdominal interventional procedures were studied in the present experiment, and a novel shielding device to reduce exposure to the patient and operator was evaluated.
MATERIALS AND METHODS: With a 16-slice CT scanner in CT fluoroscopy mode (120 kVp, 30 mA), surface dosimetry was performed on adult and pediatric phantoms. The shielding was composed of tungsten antimony in the form of a lightweight polymer sheet. Doses to the patient were measured with and without shielding for thoracic and abdominal procedures. Doses to the operator were recorded with and without phantom, gantry, and table shielding in place. Double-layer lead-free gloves were used by the operator during the procedures.
RESULTS: Tungsten antimony shielding adjacent to the scan plane resulted in a maximum dose reduction of 92.3% to the patient. Maximum 85.6%, 93.3%, and 85.1% dose reductions were observed for the operator's torso, gonads, and hands, respectively. The use of double-layer lead-free gloves resulted in a maximum radiation dose reduction of 97%.
CONCLUSIONS: Methods to reduce exposure during CT fluoroscopy are effective and should be searched for. Significant reduction in radiation doses to the patient and operator can be accomplished with tungsten antimony shielding.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17185699      PMCID: PMC2408953          DOI: 10.1097/01.RVI.0000244847.63204.5f

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  18 in total

1.  Radiation doses in computed tomography. The increasing doses of radiation need to be controlled.

Authors:  M M Rehani; M Berry
Journal:  BMJ       Date:  2000-03-04

2.  Comparison of sonographic and CT guidance techniques: does CT fluoroscopy decrease procedure time?

Authors:  D H Sheafor; E K Paulson; M A Kliewer; D M DeLong; R C Nelson
Journal:  AJR Am J Roentgenol       Date:  2000-04       Impact factor: 3.959

3.  Value of CT fluoroscopy for percutaneous biopsy procedures.

Authors:  D Gianfelice; L Lepanto; P Perreault; C Chartrand-Lefebvre; P C Milette
Journal:  J Vasc Interv Radiol       Date:  2000 Jul-Aug       Impact factor: 3.464

4.  Robotically driven interventions: a method of using CT fluoroscopy without radiation exposure to the physician.

Authors:  Stephen B Solomon; Alexandru Patriciu; Mark E Bohlman; Louis R Kavoussi; Dan Stoianovici
Journal:  Radiology       Date:  2002-10       Impact factor: 11.105

5.  Radiation dose to the radiologist's hand during continuous CT fluoroscopy-guided interventions.

Authors:  Beate M Stoeckelhuber; Thorsten Leibecke; Edda Schulz; Uwe H Melchert; Claudia U Bergmann-Koester; Thomas Helmberger; Joerg Gellissen
Journal:  Cardiovasc Intervent Radiol       Date:  2005 Sep-Oct       Impact factor: 2.740

6.  Precise biopsy localization by computer tomography.

Authors:  J R Haaga; R J Alfidi
Journal:  Radiology       Date:  1976-03       Impact factor: 11.105

7.  Benefits and safety of CT fluoroscopy in interventional radiologic procedures.

Authors:  S K Carlson; C E Bender; K L Classic; F E Zink; J P Quam; E M Ward; A L Oberg
Journal:  Radiology       Date:  2001-05       Impact factor: 11.105

8.  A radiation protection device for use during percutaneous transhepatic examinations.

Authors:  D Miotto; G Feltrin; M Calamosca
Journal:  Radiology       Date:  1984-06       Impact factor: 11.105

9.  Patient and personnel exposure during CT fluoroscopy-guided interventional procedures.

Authors:  R D Nawfel; P F Judy; S G Silverman; S Hooton; K Tuncali; D F Adams
Journal:  Radiology       Date:  2000-07       Impact factor: 11.105

10.  In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and clinical data.

Authors:  Bradley L Fricke; Lane F Donnelly; Donald P Frush; Terry Yoshizumi; Vladimir Varchena; Stacy A Poe; Javier Lucaya
Journal:  AJR Am J Roentgenol       Date:  2003-02       Impact factor: 3.959

View more
  14 in total

Review 1.  Personal dosimetry for interventional operators: when and how should monitoring be done?

Authors:  C J Martin
Journal:  Br J Radiol       Date:  2010-12-15       Impact factor: 3.039

2.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

3.  Dose to the interventional radiologist in CTF-guided procedures.

Authors:  J G Alves; S Sarmento; J S Pereira; M F Pereira; M J Sousa; L Cunha; A Dias; A D Oliveira; J V Cardoso; L M Santos; J Lencart; M Gouvêa; J A M Santos
Journal:  Radiat Environ Biophys       Date:  2019-04-16       Impact factor: 1.925

4.  Reduction of radiation exposure by lead curtain shielding in dedicated extremity cone beam CT.

Authors:  C-H Lee; J H Ryu; Y-H Lee; K-H Yoon
Journal:  Br J Radiol       Date:  2015-03-26       Impact factor: 3.039

5.  Real-time X-ray-based 4D image guidance of minimally invasive interventions.

Authors:  Jan Kuntz; Rajiv Gupta; Stefan O Schönberg; Wolfhard Semmler; Marc Kachelrieß; Sönke Bartling
Journal:  Eur Radiol       Date:  2013-01-12       Impact factor: 5.315

6.  Interventional Angiography: Radiation Protection for the Examiner by using Lead-free Gloves.

Authors:  Peter Kamusella; Fabian Scheer; Christopher Wilhelm Lüdtke; Philipp Wiggermann; Christian Wissgott; Reimer Andresen
Journal:  J Clin Diagn Res       Date:  2017-07-01

7.  Improvement of image quality and dose management in CT fluoroscopy by iterative 3D image reconstruction.

Authors:  Oliver S Grosser; Christian Wybranski; Dennis Kupitz; Maciej Powerski; Konrad Mohnike; Maciej Pech; Holger Amthauer; Jens Ricke
Journal:  Eur Radiol       Date:  2017-02-06       Impact factor: 5.315

8.  Dual guidance (CT and fluoroscopy) vertebroplasty: radiation dose to radiologists. How much and where?

Authors:  Nicolas Amoretti; Virginie Lesbats; Pierre-Yves Marcy; Olivier Hauger; Marie-eve Amoretti; Yvonne Maratos; Antoine Ianessi; Pascal Boileau
Journal:  Skeletal Radiol       Date:  2010-05-23       Impact factor: 2.199

9.  First experience of efficacy and radiation exposure in 320-detector row CT fluoroscopy-guided interventions.

Authors:  Shota Yamamoto; Tomohiro Matsumoto; Satoshi Suda; Kosuke Tomita; Shunsuke Kamei; Kazunobu Hashida; Yutaka Imai; Kazuyuki Endo; Katsuki Murakami; Terumitsu Hasebe
Journal:  Br J Radiol       Date:  2021-02-05       Impact factor: 3.039

10.  Fluoroscopic Cranial Radiation Exposure in Spine Surgery: A Prospective Single-Center Evaluation in Operating Room Personnel.

Authors:  Laila Malani Mohammad; James Messegee; M Omar Chohan; Christopher L Taylor
Journal:  Int J Spine Surg       Date:  2019-02-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.