Radiation protection of operators and patients in a hybrid Angio-MR suite.

M Dehairs1, N W Marshall2, H Bosmans3, Martino Leghissa4.   

Abstract

This work investigates the patient eye lens dose and x-ray scatter to the operator expected for a proposed hybrid Angio-MR concept. Two geometries were simulated for comparative assessment: a standard C-arm device for neuro-angiography applications and an innovative hybrid Angio-MR system concept, proposed by Siemens Healthineers. The latter concept is based on an over-couch x-ray tube and a detector inside an MRI system, with the aim of allowing combined, simultaneous MRI and x-ray imaging for procedures such as neurovascular interventions (including x-ray fluoroscopy and angiography imaging, 3D imaging, diffusion, and perfusion). To calculate the scattered radiation dose to the physician, Monte Carlo simulations were performed. Dose estimates of simplified models of the brain and eyes of both the patient and the physician and of the physician's torso and legs have been calculated. A number of parameters were varied in the simulation including x-ray spectrum, field of view (FOV), x-ray tube angulation, presence of shielding material and position of the physician. Additionally, 3D dose distributions were calculated in the vertical and horizontal planes in both setups. The patient eye lens dose was also calculated using a detailed voxel phantom and measured by means of thermoluminescent dosimeters (TLDs) to obtain a more accurate estimate. Assuming the same number of x-rays and the same size of the irradiated area on the patient's head, the results show a significant decrease in the scattered radiation to the physician for the Angio-MR system, while large increases, depending on setup, are expected to patient eye lens dose.
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Comparative study; Dose estimates; Hybrid Angio-MR; Radioprotection; Simulation

Mesh:

Year:  2020        PMID: 32473413     DOI: 10.1016/j.ejmp.2020.04.028

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Radiation Eye Dose for Physicians in CT Fluoroscopy-Guided Biopsy.

Authors:  Yohei Inaba; Shin Hitachi; Munenori Watanuki; Koichi Chida
Journal:  Tomography       Date:  2022-02-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.