Dose reduction during CT fluoroscopy: phantom study of angular beam modulation.

Christian Hohl1, Christoph Suess, Joachim E Wildberger, Dagmar Honnef, Marco Das, Georg Mühlenbruch, Andreas Schaller, Rolf W Günther, Andreas H Mahnken.   

Abstract

PURPOSE: To prospectively evaluate, in a phantom, the dose reductions achievable by using angular beam modulation (ABM) during computed tomographic (CT) fluoroscopy-guided thoracic interventions.
MATERIALS AND METHODS: To enable measurement of organ doses and effective patient dose, a female Alderson-Rando phantom was equipped with thermoluminescent dosimeters (TLDs) in 41 positions, with three TLDs in each position. Additionally, the local dose was assessed in 22 locations above the phantom to estimate the radiation exposure to the radiologist's hand and the patient's skin dose during thoracic interventions. Radiation exposure was performed with a 64-section multidetector CT scanner in the CT fluoroscopy mode, simulating a CT fluoroscopy-guided chest intervention. Effective dose, breast dose, and the dose to the radiologist's hand during the simulated chest intervention were measured with and without ABM. Image noise as an indicator for image quality was compared for both settings. Statistical significance of the measured dose reductions and the image noise was tested by using the paired-samples t test, with P < .05 indicating a significant difference.
RESULTS: ABM significantly reduced the effective patient dose by 35%, the skin dose by 75%, the breast dose by 47% (P < .001 for all), and the physician's hand dose by between 27% (scattered radiation, P = .007) and 72% (direct radiation, P < .001). No significant difference was found in a comparison of the image noise with and that without ABM.
CONCLUSION: ABM leads to significant dose reductions for both patients and personnel during CT fluoroscopy-guided thoracic interventions, without impairing image quality. (c) RSNA, 2008.

Entities:  

Mesh:

Year:  2008        PMID: 18227544     DOI: 10.1148/radiol.2462061968

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  18 in total

1.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

2.  [Computed tomography guidance: fluoroscopy and more].

Authors:  P M Paprottka; T Helmberger; M F Reiser; C G Trumm
Journal:  Radiologe       Date:  2013-11       Impact factor: 0.635

3.  CT fluoroscopy-guided vertebral augmentation with a radiofrequency-induced, high-viscosity bone cement (StabiliT(®)): technical results and polymethylmethacrylate leakages in 25 patients.

Authors:  Christoph Gregor Trumm; Tobias F Jakobs; Robert Stahl; Torleif A Sandner; Philipp M Paprottka; Maximilian F Reiser; Christoph J Zech; Ralf-Thorsten Hoffmann
Journal:  Skeletal Radiol       Date:  2012-03-16       Impact factor: 2.199

4.  Radiation dose and image quality of CT fluoroscopy with partial exposure mode.

Authors:  Keisuke Takiguchi; Atsushi Urikura; Tsukasa Yoshida; Yoshihiro Nakaya; Masahiro Endo; Takeshi Aramaki
Journal:  Diagn Interv Radiol       Date:  2020-07       Impact factor: 2.630

5.  CT fluoroscopy guided percutaneous gastrostomy or jejunostomy without (CT-PG/PJ) or with simultaneous endoscopy (CT-PEG/PEJ) in otherwise untreatable patients.

Authors:  Fritz W Spelsberg; Ralf-Thorsten Hoffmann; Reinhold A Lang; Hauke Winter; Rolf Weidenhagen; Maximilian Reiser; Karl-Walter Jauch; Christoph Trumm
Journal:  Surg Endosc       Date:  2012-12-12       Impact factor: 4.584

6.  Evaluation of surface dose and image quality using the half-scan mode in chest computed tomography-guided interventional radiology: a phantom study.

Authors:  Hiroaki Hasegawa; Jiro Sato; Ikuo Kobayashi
Journal:  Radiol Phys Technol       Date:  2018-02-27

7.  Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology.

Authors:  Donald L Miller; Eliseo Vañó; Gabriel Bartal; Stephen Balter; Robert Dixon; Renato Padovani; Beth Schueler; John F Cardella; Thierry de Baère
Journal:  Cardiovasc Intervent Radiol       Date:  2009-12-18       Impact factor: 2.740

8.  Improvement of image quality and dose management in CT fluoroscopy by iterative 3D image reconstruction.

Authors:  Oliver S Grosser; Christian Wybranski; Dennis Kupitz; Maciej Powerski; Konrad Mohnike; Maciej Pech; Holger Amthauer; Jens Ricke
Journal:  Eur Radiol       Date:  2017-02-06       Impact factor: 5.315

9.  Technical and clinical outcome of percutaneous CT fluoroscopy-guided screw placement in unstable injuries of the posterior pelvic ring.

Authors:  Frederik F Strobl; Sophia M Haeussler; Philipp M Paprottka; Ralf-Thorsten Hoffmann; Oliver Pieske; Maximilian F Reiser; Christoph G Trumm
Journal:  Skeletal Radiol       Date:  2014-05-10       Impact factor: 2.199

10.  CT fluoroscopy-guided percutaneous osteoplasty with or without radiofrequency ablation in the treatment of painful extraspinal and spinal bone metastases: technical outcome and complications in 29 patients.

Authors:  Caroline A Burgard; Julien Dinkel; Frederik Strobl; Philipp M Paprottka; Nicolai Schramm; Maximilian Reiser; Christoph G Trumm
Journal:  Diagn Interv Radiol       Date:  2018 May-Jun       Impact factor: 2.630

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.