Corneal deformation measurement using Scheimpflug noncontact tonometry.

Ying Hon1, Andrew K C Lam.   

Abstract

PURPOSE: To determine the intraexaminer repeatability and intersession reproducibility of corneal deformation measurement using Scheimpflug noncontact tonometry (Corvis ST) on normal subjects.
METHODS: Thirty-seven adults aged 20 to 48 years were invited to have their corneal deformation and curvature measurements taken using Corvis ST and Pentacam, respectively. Three consecutive measurements were taken for each instrument between 9:00 and 11:00 AM for intraexaminer repeatability analysis. Participants returned between 3:00 and 5:00 PM the same day for intersession reproducibility analysis.
RESULTS: The most repeatable corneal parameter measured by Corvis ST was central corneal thickness ([CCT] ICC, 0.96; precision, 10.85 μm; repeatability, 15.34 μm; CV, 1.01%), followed by deformation amplitude ([DA] ICC, 0.80; precision, 0.08 mm; repeatability, 0.13 mm; CV, 4.33%), first applanation time ([1st A-time] ICC, 0.77; precision, 0.22 milliseconds; repeatability, 0.31 milliseconds; CV, 1.42), and intraocular pressure ([IOP] ICC, 0.75; precision, 1.39 mm Hg; repeatability, 1.97 mm Hg; CV, 4.98). Other parameters showed poor repeatability. The DA and 1st A-time showed good intersession reproducibility. The 95% limits of agreement were +0.13 to -0.13 mm for DA and +0.27 to -0.33 milliseconds for 1st A-time. The DA was negatively correlated with central corneal thickness (r = -0.53, p < 0.001) but not with corneal curvatures (flattest curvature, r = 0.13, p = 0.46; steepest curvature, r = 0.05, p = 0.75).
CONCLUSIONS: Corneal deformation parameters DA and 1st A-time were repeatable and reproducible. A thinner cornea was associated with a higher corneal deformation. Measurement of DA serves as an indicator of corneal biomechanical properties.

Entities:  

Mesh:

Year:  2013        PMID: 23238261     DOI: 10.1097/OPX.0b013e318279eb87

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  70 in total

1.  [New aspects on biomechanics of the cornea in keratoconus].

Authors:  Z Gatzioufas; B Seitz
Journal:  Ophthalmologe       Date:  2013-09       Impact factor: 1.059

2.  Imaging Corneal Biomechanical Responses to Ocular Pulse Using High-Frequency Ultrasound.

Authors:  Elias Pavlatos; Hong Chen; Keyton Clayson; Xueliang Pan; Jun Liu
Journal:  IEEE Trans Med Imaging       Date:  2018-02       Impact factor: 10.048

3.  Corneal biomechanical properties in 3 corneal transplantation techniques with a dynamic Scheimpflug analyzer.

Authors:  Naoyuki Maeda; Ryotaro Ueki; Mutsumi Fuchihata; Hisataka Fujimoto; Shizuka Koh; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2014-09-05       Impact factor: 2.447

4.  Spatial characterization of corneal biomechanical properties with optical coherence elastography after UV cross-linking.

Authors:  Michael D Twa; Jiasong Li; Srilatha Vantipalli; Manmohan Singh; Salavat Aglyamov; Stanislav Emelianov; Kirill V Larin
Journal:  Biomed Opt Express       Date:  2014-04-04       Impact factor: 3.732

5.  Changes in ocular biomechanics after treatment for active Graves' orbitopathy.

Authors:  H X Li; X H Zhao; Y Song; B K Mu; Y Pan; H Zhao; Y Wang
Journal:  J Endocrinol Invest       Date:  2020-06-07       Impact factor: 4.256

6.  Comparison of corneal biomechanics in Sjögren's syndrome and non-Sjögren's syndrome dry eyes by Scheimpflug based device.

Authors:  Qin Long; Jing-Yi Wang; Dong Xu; Ying Li
Journal:  Int J Ophthalmol       Date:  2017-05-18       Impact factor: 1.779

7.  Evaluation of Corvis ST tonometer with the updated software in glaucoma practice.

Authors:  Ioannis Halkiadakis; Vasilios Tzimis; Alexandros Gryparis; Ioannis Markopoulos; Vasiliki Konstadinidou; Elias Zintzaras; Michalis Tzakos
Journal:  Int J Ophthalmol       Date:  2022-03-18       Impact factor: 1.779

8.  Measurement repeatability of the dynamic Scheimpflug analyzer.

Authors:  Atsuya Miki; Naoyuki Maeda; Tomoko Asai; Yasushi Ikuno; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2017-10-05       Impact factor: 2.447

9.  Evaluation of Intraocular Pressure and Other Biomechanical Parameters to Distinguish between Subclinical Keratoconus and Healthy Corneas.

Authors:  Cristina Peris-Martínez; María Amparo Díez-Ajenjo; María Carmen García-Domene; María Dolores Pinazo-Durán; María José Luque-Cobija; María Ángeles Del Buey-Sayas; Susana Ortí-Navarro
Journal:  J Clin Med       Date:  2021-04-28       Impact factor: 4.241

10.  Acoustic Radiation Force Optical Coherence Elastography of Corneal Tissue.

Authors:  Yueqiao Qu; Teng Ma; Youmin He; Jiang Zhu; Cuixia Dai; Mingyue Yu; Shenghai Huang; Fan Lu; K Kirk Shung; Qifa Zhou; Zhongping Chen
Journal:  IEEE J Sel Top Quantum Electron       Date:  2016-02-08       Impact factor: 4.544

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.