Literature DB >> 24833745

Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer.

Noor Q Ali1, Dipika V Patel1, Charles N J McGhee1.   

Abstract

PURPOSE: To examine the repeatability of measurements obtained using a noncontact Scheimpflug-based tonometer (Corvis ST) and investigate potential differences in these parameters between healthy and keratoconic (KC) corneas.
METHODS: Forty-five keratoconic eyes and 103 healthy eyes were examined using biomicroscopy, corneal tomography and the Corvis ST (CST).
RESULTS: Intraocular pressure and central corneal thickness (CCT) were highly repeatable (intraclass correlation [ICC] > 0.70, coefficient of variation [CV] < 0.20). Deformation amplitude (DA) and applanation-1 time (A1T) were fairly repeatable (ICC > 0.60, CV < 0.20). There was no association between DA and age, sex, or ethnicity in healthy eyes. There was a greater mean DA in the KC group compared with 46 age-matched healthy eyes (KC 1.37 ± 0.21 mm, healthy 1.05 ± 0.11 mm, P < 0.001). Multivariate analysis showed DA in KC was predicted by IOP, CCT, and the steepest simulated keratometry value (P = 0.03, P = 0.03, P = 0.001 respectively, R(2) = 0.75). A subgroup analysis of healthy and KC eyes with comparable mean CCT and IOP highlighted a statistically significant difference in mean DA (KC 1.25 ± 0.08 mm, thin healthy 1.13 ± 0.09 mm, P = 0.006). Receiver-operating characteristic analysis showed an area under the curve of 0.77 (95% CI 0.61-0.93, P = 0.006) but no ideal cutoff value for DA.
CONCLUSIONS: Key parameters assessed by the CST are repeatable. Keratoconus is associated with greater DA than in healthy eyes, even when controlled for CCT and IOP. Deformation amplitude may be a useful adjunct in keratoconus assessment and monitoring, but cannot solely discriminate between healthy and keratoconic corneas. Copyright 2014 The Association for Research in Vision and Ophthalmology, Inc.

Entities:  

Keywords:  Corvis ST; corneal deformation amplitude; diagnostic instruments; non-contact tonometry; repeatability

Mesh:

Year:  2014        PMID: 24833745     DOI: 10.1167/iovs.13-13715

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  48 in total

1.  Imaging Corneal Biomechanical Responses to Ocular Pulse Using High-Frequency Ultrasound.

Authors:  Elias Pavlatos; Hong Chen; Keyton Clayson; Xueliang Pan; Jun Liu
Journal:  IEEE Trans Med Imaging       Date:  2018-02       Impact factor: 10.048

Review 2.  Mechanical interactions and crosstalk between corneal keratocytes and the extracellular matrix.

Authors:  W Matthew Petroll; Miguel Miron-Mendoza
Journal:  Exp Eye Res       Date:  2015-04       Impact factor: 3.467

3.  Comparison of corneal biological parameters between transepithelial and epithelium-off corneal cross-linking in keratoconus.

Authors:  Bo-Wen Ouyang; Hui Ding; Han Wang; Zhen-Duo Yang; Tan Zhong; Hong-Ming Fan; Xing-Wu Zhong
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

Review 4.  [Dynamic Scheimpflug Analyzer (Corvis ST) for measurement of corneal biomechanical parameters : A praxis-related overview].

Authors:  R Herber; N Terai; K R Pillunat; F Raiskup; L E Pillunat; E Spörl
Journal:  Ophthalmologe       Date:  2018-08       Impact factor: 1.059

5.  Effect of topical prostaglandins on the biomechanics and shape of the cornea.

Authors:  Shiro Amano; Ryohei Nejima; Kenji Inoue; Kazunori Miyata
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-08-10       Impact factor: 3.117

6.  A new method for detecting the outer corneal contour in images from an ultra-fast Scheimpflug camera.

Authors:  Magdalena Jędzierowska; Robert Koprowski; Sławomir Wilczyński; Katarzyna Krysik
Journal:  Biomed Eng Online       Date:  2019-12-03       Impact factor: 2.819

7.  Measurement repeatability of the dynamic Scheimpflug analyzer.

Authors:  Atsuya Miki; Naoyuki Maeda; Tomoko Asai; Yasushi Ikuno; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2017-10-05       Impact factor: 2.447

8.  [Early diagnosis of keratoconus].

Authors:  Stefan J Lang; P Maier; T Böhringer; T Reinhard
Journal:  Ophthalmologe       Date:  2021-07-23       Impact factor: 1.059

Review 9.  Corneal biomechanics: Measurement and structural correlations.

Authors:  Jillian Chong; William J Dupps
Journal:  Exp Eye Res       Date:  2021-02-18       Impact factor: 3.467

10.  One-Year Follow-Up of Corneal Biomechanical Changes After Accelerated Transepithelial Corneal Cross-Linking in Pediatric Patients With Progressive Keratoconus.

Authors:  Weijun Jian; Mi Tian; Xiaoyu Zhang; Ling Sun; Yang Shen; Meiyan Li; Xingtao Zhou
Journal:  Front Med (Lausanne)       Date:  2021-07-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.