Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry.

Omür Ozlenen Uçakhan1, Muhip Ozkan, Ayfer Kanpolat.   

Abstract

PURPOSE: To compare central corneal thickness (CCT) measurements taken with the Pentacam comprehensive eye scanner (CES), noncontact specular microscopy (SM), and ultrasound pachymetry (UP) in normal and keratoconic corneas.
SETTING: Department of Ophthalmology, Ankara University School of Medicine, Ankara, Turkey.
METHODS: In a prospective study, 3 CCT measurements were taken with the Pentacam CES, SM, and UP in that sequence from 1 eye of 45 consecutive patients with myopia (group A) and 62 consecutive patients with keratoconus (group B). Eyes with keratoconus were further divided into 3 subgroups, mild, moderate, and severe, according to the mean keratometry readings.
RESULTS: Pentacam CES (r = 0.994) and UP (r = 0.993) demonstrated very high and comparable reproducibility in group A. In group B, Pentacam CES displayed better reproducibility (r = 0.988) than UP (r = 0.969) and SM (r = 0.901). The mean CCT measurements of Pentacam CES and UP were not significantly different in group A (P = .37) and in eyes with mild keratoconus (P = .29), whereas significant differences between all instrument pairs were evident in group B and in moderate and severely keratoconic eyes (P<.05). Noncontact SM measurements were significantly smaller than those of Pentacam CES and UP in all groups (P<.05) when a measurement could be obtained. There were significant linear correlations between CCT measurements of Pentacam CES, UP, and noncontact SM in groups A, B, and mildly keratoconic eyes (P<.001).
CONCLUSIONS: Results suggest that whereas Pentacam CES and UP may be used interchangeably in normal eyes in the clinical setting for the measurement of CCT, one should be cautious interpreting corneal thickness data from Pentacam CES, UP, and particularly SM in eyes with keratoconus. Whereas, in normal and mildly keratoconic eyes, Pentacam CES and UP demonstrated very high and comparable reproducibility, in moderately keratoconic eyes, Pentacam CES readings showed better reproducibility than UP.

Entities:  

Mesh:

Year:  2006        PMID: 16814055     DOI: 10.1016/j.jcrs.2006.02.037

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  40 in total

1.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

3.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

4.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

5.  [Influence of corneal pachymetric changes on functional results after cataract surgery].

Authors:  F N Auerbach; M P Holzer; G U Auffarth; R Khoramnia; B C Thomas; D Saure; T M Rabsilber
Journal:  Ophthalmologe       Date:  2015-10       Impact factor: 1.059

6.  Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

Authors:  Mohammad Naderan; Saeed Shoar; Morteza Naderan; Mohammad Amin Kamaleddin; Mohammad Taher Rajabi
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

7.  Corneal thickness and volume in subclinical and clinical keratoconus.

Authors:  Seyed Mahdi Ahmadi Hosseini; Norhani Mohidin; Fereshteh Abolbashari; Bariah Mohd-Ali; Chandramalar T Santhirathelagan
Journal:  Int Ophthalmol       Date:  2012-11-09       Impact factor: 2.031

8.  Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Authors:  Zaina N Al-Mohtaseb; Li Wang; Mitchell P Weikert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-17       Impact factor: 3.117

9.  Short-term corneal changes with gas-permeable contact lens wear in keratoconus subjects: a comparison of two fitting approaches.

Authors:  Miguel Romero-Jiménez; Jacinto Santodomingo-Rubido; Patricia Flores-Rodríguez; Jose-Manuel González-Méijome
Journal:  J Optom       Date:  2014-09-04

10.  Day to Day Clinically Relevant Corneal Elevation, Thickness, and Curvature Parameters Using the Orbscan II Scanning Slit Topographer and the Pentacam Scheimpflug Imaging Device.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.