J Jorge1, Jl Rosado, Ja Díaz-Rey, Jm González-Méijome.
Abstract
BACKGROUND: The purpose of this study was to compare the accuracy of the new Sirius(®) Scheimpflug anterior segment examination device for measurement of central corneal thickness (CCT) and anterior chamber depth (ACD) with that of CCT measurements obtained by ultrasound pachymetry and ACD measurements obtained by ultrasound biometry, respectively.Entities:
Keywords: Scheimpflug corneal tomography; limits of agreement; ultrasound biometry; ultrasound pachymetry
Year: 2013 PMID: 23467857 PMCID: PMC3587393 DOI: 10.2147/OPTH.S35121
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Image of anterior segment obtained with the Sirius® Scheimpflug device.
Abbreviations: ACD, anterior chamber depth; CCT, central corneal thickness.
Average values and standard deviation for central corneal thickness and anterior chamber depth measurements obtained with the ultrasound devices and Sirius® for three runs of measurements
| n = 50 | Central corneal thickness | Anterior chamber depth | ||
|---|---|---|---|---|
|
|
|
| ||
| Measure | Sirius (mean ± SD) (μm) | SP100 (mean ± SD) (μm) | Sirius (mean ± SD) (mm) | US800 (mean ± SD) (mm) |
| 1 | 546.5 ± 39.43 | 540.6 ± 34.46 | 2.95 ± 0.33 | 3.37 ± 0.30 |
| 2 | 545.4 ± 38.46 | 541.4 ± 36.07 | 2.97 ± 0.31 | 3.37 ± 0.32 |
| 3 | 545.5 ± 38.63 | 542.0 ± 35.01 | 2.96 ± 0.30 | 3.34 ± 0.30 |
| 0.987 | 0.980 | 0.852 | 0.964 | |
Notes:P, statistical significance for differences among the three test runs as obtained using
analysis of variance test;
Kruskal-Wallis test.
Abbreviation: SD, standard deviation.
Average values, standard deviation, and range for central corneal thickness and anterior chamber depth measurements obtained with ultrasound and the Sirius®
| n = 50 | Mean | SD | Maximum | Minimum | |
|---|---|---|---|---|---|
| Central corneal thickness (μm) | SP100 | 541.3 | 35.04 | 596.7 | 459.3 |
| Sirius | 546.0 | 38.53 | 614.0 | 456.7 | |
| Anterior chamber depth (mm) | US800 | 3.36 | 0.29 | 4.16 | 2.89 |
| Sirius | 2.96 | 0.30 | 3.59 | 2.26 |
Abbreviation: SD, standard deviation.
Figure 2Bland-Altman analysis of CCT values measured by ultrasound and Sirius®. Abbreviation: CCT, central corneal thickness.
Figure 3Bland-Altman analysis of ACD values measured with ultrasound biometry and Sirius® device.
Abbreviation: ACD, anterior chamber depth.
Figure 4Bland-Altman analysis of ACD values measured with ultrasound biometry and Sirius® device by adding the value of CCT to the Sirius ACD value.
Abbreviations: ACD, anterior chamber depth; CCT, central corneal thickness.
Figure 5Bland-Altman analysis of ACD values measured with ultrasound biometry subtracting the value of CCT from the ultrasound biometric measure and Sirius® device.
Abbreviations: ACD, anterior chamber depth; CCT, central corneal thickness.
Mean difference, standard deviation, level of statistical significance, and limits of agreement between each set of measurement techniques to be compared at the 95% confidence interval
| Mean | SD | Limits of agreement | |||
|---|---|---|---|---|---|
|
| |||||
| Mean difference −1.96 | Mean difference +1.96 | ||||
| CCT (μm) | |||||
| Sirius versus SP100 | 4.68 | 10.47 | 0.003 | −15.84 | 25.20 |
| ACD (mm) | |||||
| Sirius versus US800 | −0.40 | 0.16 | <0.001 | −0.72 | −0.07 |
| Sirius + CCT | 0.15 | 0.16 | <0.001 | −0.16 | 0.45 |
| Sirius versus US800 – CCT | 0.15 | 0.16 | <0.001 | −0.16 | 0.45 |
Notes:
Sirius + CCT represents the sum of ACD and CCT obtained with Sirius;
US800 – CCT represents the value of ACD obtained with US800 ultrasound biometry minus the CCT obtained with ultrasound pachymetry.
Abbreviations: ACD, anterior chamber depth; CCT, central corneal thickness; SD, standard deviation.