Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices.

Fangjun Bao1, Qinmei Wang, Shiming Cheng, Giacomo Savini, Weicong Lu, Yifan Feng, Ye Yu, Jinhai Huang.   

Abstract

PURPOSE: The aim of this study was to evaluate the repeatability of central corneal thickness (CCT) measurements in normal eyes using 2 new noncontact specular microscopes (NCSMs) EM-3000 (Tomey, Japan) and SP-02 (CSO, Italy) and to compare the results with those obtained from an SP-3000P NCSM (Topcon, Japan) and ultrasound pachymetry (USP).
METHODS: Seventy subjects were enrolled in a prospective study. A single experienced ophthalmologist performed tests with each of the 4 instruments. Measurements were obtained in the right eye during the same session. The testing sequence of the NCSM was randomly selected. After performing noncontact examinations, the USP was performed to derive the CCT measurements. Intraoperator repeatability was analyzed using within-subject coefficient of variation and intraclass correlation coefficients. The agreement between NCSMs or NCSM and USP was assessed with Bland-Altman plots and 95% limits of agreement (LoA).
RESULTS: The mean CCT values measured by SP-3000P, EM-3000, SP-02, and USP were 513.66 ± 33.14 μm, 529.12 ± 33.22 μm, 549.06 ± 40.27 μm, and 539.01 ± 35.73 μm, respectively. All coefficients of variation were <1.3%, and the intraclass correlation coefficients were >0.95. There were statistically significant differences between any 2 devices as determined by CCT measurements. The mean difference between paired comparisons was >9 μm. The 95% LoA ranges were broad, and the greatest 95% LoA was found to exist between SP-3000P and SP-02.
CONCLUSIONS: The new NCSMs and USP all show a high intraoperator repeatability for CCT measurements in normal eyes. However, interdevice agreement was poor and prevented the comparison of CCT measurements taken with different instruments.

Entities:  

Mesh:

Year:  2014        PMID: 24699563     DOI: 10.1097/ICO.0000000000000113

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  15 in total

1.  Letter to the editor. "Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery".

Authors:  Alberto López-Miguel; María Sanchidrián; Itziar Fernández; Alfredo Holgueras; Miguel J Maldonado
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-02-11       Impact factor: 3.117

2.  Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens.

Authors:  Manuel Garza-Leon; Hugo A Fuentes-de la Fuente; Ana V García-Treviño
Journal:  Int Ophthalmol       Date:  2016-10-21       Impact factor: 2.031

3.  Measurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices.

Authors:  Ozgur Cakici; Remzi Karadag; Huseyin Bayramlar; Efe Koyun
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

4.  Evaluation of corneal endothelium after UVA/riboflavin cross-linking in thin keratoconic corneas.

Authors:  Wei Chen; Zhi-Wei Li; Xiao-Min Zhao; Wen-Wen Xu; Guo-Ying Mu
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

5.  Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes.

Authors:  Irmak Karaca; Suzan Guven Yilmaz; Melis Palamar; Halil Ates
Journal:  Int Ophthalmol       Date:  2017-07-03       Impact factor: 2.031

6.  Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.

Authors:  Alberto López-Miguel; María Sanchidrián; Itziar Fernández; Alfredo Holgueras; Miguel J Maldonado
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-11-22       Impact factor: 3.117

Review 7.  The thick and thin of the central corneal thickness in glaucoma.

Authors:  Graham W Belovay; Ivan Goldberg
Journal:  Eye (Lond)       Date:  2018-02-15       Impact factor: 3.775

8.  Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

Authors:  Ayong Yu; Weiqi Zhao; Giacomo Savini; Zixu Huang; Fangjun Bao; Weicong Lu; Qinmei Wang; Jinhai Huang
Journal:  J Ophthalmol       Date:  2015-12-01       Impact factor: 1.909

9.  Age related changes in corneal morphological characteristics of healthy Pakistani eyes.

Authors:  Qamar Ul Islam; Muhammad Kamran Saeed; Mohammad Asim Mehboob
Journal:  Saudi J Ophthalmol       Date:  2017-03-09

10.  Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry.

Authors:  Rongrong Gao; Hao Chen; Giacomo Savini; Yaxin Miao; Xiaorui Wang; Jing Yang; Weiqi Zhao; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-05-30       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.