Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound.

Birgit Lackner1, Gerald Schmidinger, Stefan Pieh, Martin A Funovics, Christian Skorpik.   

Abstract

PURPOSE: The purpose of this study was to compare central corneal thickness (CCT) measurements obtained with a novel rotating Scheimpflug camera (Pentacam; Oculus) with scanning slit topography (Orbscan; Bausch & Lomb), and with ultrasound pachymetry (SP-2000; Tomey).
METHODS: CCT in 30 healthy eyes was measured twice with each modality by 2 independent observers in random order. The results from scanning slit topography are given both with and without multiplication with the "acoustic correction factor" of 0.92. In addition, the displayed images from the rotating Scheimpflug camera and scanning slit topography were used to calculate the signal difference-to-noise ratios (SD/N) between cornea and background signal.
RESULTS: The mean CCT values as determined with the different modalities (+/-standard deviation) were: 542+/-29 microm, 576+/-37 microm, 530+/-34 microm, and 552+/-32 microm for rotating Scheimpflug imaging, for uncorrected and for corrected scanning slit pachymetry, and for ultrasound, respectively. The differences between modalities (+/-95% limits of agreement) were -9.8+/-31 microm between rotating Scheimpflug and ultrasound, 24+/-31.2 microm between scanning slit and ultrasound, and 33+/-27 microm between scanning slit and rotating Scheimpflug imaging. The limits of agreement for within and between observer effects were within 4.2% of the absolute CCT values for scanning slit and ultrasound and within 2.2% for the rotating Scheimpflug imaging. The rotating Scheimpflug camera showed similar SD/N ratios but steeper edges of the corneal surfaces in the intensity profile plots.
CONCLUSION: In the assessment of normal corneas, the Pentacam measured CCT values closer to ultrasound pachymetry and with less variability compared with Orbscan. The (interobserver) reproducibility with the Pentacam was highest of all 3 modalities.

Entities:  

Mesh:

Year:  2005        PMID: 16276321     DOI: 10.1097/01.opx.0000180817.46312.0a

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  73 in total

1.  Terahertz sensing in corneal tissues.

Authors:  David B Bennett; Zachary D Taylor; Pria Tewari; Rahul S Singh; Martin O Culjat; Warren S Grundfest; Daniel J Sassoon; R Duncan Johnson; Jean-Pierre Hubschman; Elliott R Brown
Journal:  J Biomed Opt       Date:  2011-05       Impact factor: 3.170

2.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

3.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

Review 4.  Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?

Authors:  Aristides Konstantopoulos; Parwez Hossain; David F Anderson
Journal:  Br J Ophthalmol       Date:  2007-04       Impact factor: 4.638

5.  Comparison of ultrasonic pachymetry with Orbscan in corneal haze.

Authors:  Rana Altan-Yaycioglu; Aysel Pelit; Yonca Aydin Akova
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-06-22       Impact factor: 3.117

6.  Photorefractive keratectomy in the cat eye: biological and optical outcomes.

Authors:  Lana J Nagy; Scott MacRae; Geunyoung Yoon; Matthew Wyble; Jianhua Wang; Ian Cox; Krystel R Huxlin
Journal:  J Cataract Refract Surg       Date:  2007-06       Impact factor: 3.351

7.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

Review 8.  [Corneal topography. Analysis of the corneal surface].

Authors:  C Herrmann; U Ludwig; G Duncker
Journal:  Ophthalmologe       Date:  2008-02       Impact factor: 1.059

9.  Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

Authors:  Ahmet Taylan Yazıcı; Gökhan Pekel; Ercüment Bozkurt; Yusuf Yıldırım; Evre Pekel; Ahmet Demirok; Omer Faruk Yılmaz
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

10.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.