Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Zaina N Al-Mohtaseb1, Li Wang, Mitchell P Weikert.   

Abstract

BACKGROUND: To assess the repeatability and comparability of central corneal thickness (CCT) measurements obtained from the Galilei Dual Scheimpflug Analyzer (Galilei-DSA) and from ultrasonic pachymetry (USP) in normal eyes and in eyes following laser refractive surgery.
METHODS: In this prospective observational study, 77 eyes of 39 patients with normal corneas and 39 eyes of 20 patients with prior laser refractive surgery were enrolled. All 116 eyes had three repeated measurements of central CCT with the Galilei-DSA first and then with USP by the same examiner. Reliability of both devices was assessed statistically, and CCT measurements by each were compared.
RESULTS: The mean CCT measurement of normal eyes was 549.2 ± 30.5 μm and 548.5 ± 33.1 μm by the Galilei-DSA and USP respectively; the difference between the two modalities was -0.7  ± 7.1 μm (P = 0.365). In post-refractive surgery eyes, the mean CCT measured by the Galilei-DSA and USP was 500.4 ± 44.8 and 494.3 ± 50.3 respectively; the difference between the two devices was -6.2 ± 9.9 μm (P < 0.001). For the three repeated measurements, the mean standard deviation was 2 μm with the Galilei-DSA and 1.8 μm with USP in normal eyes, and 2.4 μm with the Galilei-DSA and 1.9 μm with USP in post-refractive surgery eyes.
CONCLUSION: The Galilei-DSA has high repeatability and excellent to moderate agreement with ultrasonic pachymetry when measuring CCT in normal and post-refractive surgery eyes.

Entities:  

Mesh:

Year:  2013        PMID: 23417340     DOI: 10.1007/s00417-013-2280-0

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  23 in total

1.  Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound.

Authors:  Birgit Lackner; Gerald Schmidinger; Stefan Pieh; Martin A Funovics; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-10       Impact factor: 1.973

2.  Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry.

Authors:  Omür Ozlenen Uçakhan; Muhip Ozkan; Ayfer Kanpolat
Journal:  J Cataract Refract Surg       Date:  2006-06       Impact factor: 3.351

3.  Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia.

Authors:  Thomas Ho; Arthur C K Cheng; Srinivas K Rao; Silvania Lau; Chris K S Leung; Dennis S C Lam
Journal:  J Cataract Refract Surg       Date:  2007-07       Impact factor: 3.351

4.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

5.  Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  J Cataract Refract Surg       Date:  2007-10       Impact factor: 3.351

6.  Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes.

Authors:  Claudia Maria Prospero Ponce; Karolinne Maia Rocha; Scott D Smith; Ronald R Krueger
Journal:  J Cataract Refract Surg       Date:  2009-06       Impact factor: 3.351

Review 7.  Measuring the cornea: the latest developments in corneal topography.

Authors:  Tracy Swartz; Lisa Marten; Ming Wang
Journal:  Curr Opin Ophthalmol       Date:  2007-07       Impact factor: 3.761

8.  Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry.

Authors:  V Yaylali; S C Kaufman; H W Thompson
Journal:  J Cataract Refract Surg       Date:  1997-11       Impact factor: 3.351

9.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

10.  Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry.

Authors:  Jeevan S Ladi; Nitant A Shah
Journal:  Indian J Ophthalmol       Date:  2010 Sep-Oct       Impact factor: 1.848

View more
  6 in total

1.  Measurement of central corneal thickness and pre-corneal tear film thickness of rabbits using the Scheimpflug system.

Authors:  Jing Dong; Qiang Wu; Xiao-Gang Wang
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

2.  Precision of corneal thickness measurements obtained using the scheimpflug-placido imaging and agreement with ultrasound pachymetry.

Authors:  Jinhai Huang; Giacomo Savini; Chengfang Wang; Weicong Lu; Rongrong Gao; Yuanguang Li; Qinmei Wang; Yune Zhao
Journal:  J Ophthalmol       Date:  2015-02-24       Impact factor: 1.909

3.  Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

Authors:  Ayong Yu; Weiqi Zhao; Giacomo Savini; Zixu Huang; Fangjun Bao; Weicong Lu; Qinmei Wang; Jinhai Huang
Journal:  J Ophthalmol       Date:  2015-12-01       Impact factor: 1.909

4.  Correlation analysis between central corneal thickness and intraocular pressure in juveniles in Northern China: the Jinan city eye study.

Authors:  Wen Wei; Zhaoshan Fan; Lihua Wang; Zhiwei Li; Wanzhen Jiao; Yun Li
Journal:  PLoS One       Date:  2014-08-22       Impact factor: 3.240

5.  Sirius Scheimpflug-Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis.

Authors:  Yili Jin; Colm McAlinden; Yong Sun; Daizong Wen; Yiran Wang; Jinjin Yu; Ke Feng; Benhao Song; Qinmei Wang; Shihao Chen; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2021-02-18

6.  Evaluation of anterior segment parameters and possible influencing factors in normal subjects using a dual Scheimpflug analyzer.

Authors:  Xiaogang Wang; Jing Dong; Qiang Wu
Journal:  PLoS One       Date:  2014-05-16       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.