Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Hani S Al-Mezaine1, Saleh A Al-Amro, Dustan Kangave, Abdulkareem Sadaawy, Taher A Wehaib, Saleh Al-Obeidan.   

Abstract

PURPOSE: To compare the accuracy of central corneal thickness (CCT) measurements by the oculus pentacam scheimpflug system, with those obtained with the DGH ultrasound pachymeter (UP) and to assess the agreement between the two devices.
METHODS: In a prospective study, measurement agreement was assessed in 984 eyes of 492 healthy subjects using both oculus pentacam and ultrasonic pachymetry at the Eye Consultants Center, Riyadh, Saudi Arabia. RESULT: In the measurement agreement experiment, the mean CCT was 552.4 +/- 37.0 microm with oculus pentacam and 544.1 +/- 35.4 microm with ultrasonic pachymetry. Regression analysis showed a high correlation between the values obtained with both devices (r = 0.912, P < 0.001). Compared with UP, pentacam overestimated the CCT by a mean of 8.2 microm as demonstrated in a Bland-Altman plot.
CONCLUSION: the CCT measurements by the pentacam and UP are highly correlated. The pentacam agrees well with UP and is a reliable alternative to UP in CCT measurements.

Entities:  

Mesh:

Year:  2007        PMID: 17898935     DOI: 10.1007/s10792-007-9143-9

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  30 in total

1.  Corneal thickness measurements with contact and noncontact specular microscopic and ultrasonic pachymetry.

Authors:  L Módis; A Langenbucher; B Seitz
Journal:  Am J Ophthalmol       Date:  2001-10       Impact factor: 5.258

2.  Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures.

Authors:  N G Iskander; E Anderson Penno; N T Peters; H V Gimbel; M Ferensowicz
Journal:  J Cataract Refract Surg       Date:  2001-05       Impact factor: 3.351

3.  Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry.

Authors:  Jose Manuel González-Méijome; Alejandro Cerviño; Eva Yebra-Pimentel; Manuel A Parafita
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

4.  Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system.

Authors:  Georg Rainer; Oliver Findl; Vanessa Petternel; Barbara Kiss; Wolfgang Drexler; Christian Skorpik; Michael Georgopoulos; Leopold Schmetterer
Journal:  Ophthalmology       Date:  2004-05       Impact factor: 12.079

5.  Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry.

Authors:  Omür Ozlenen Uçakhan; Muhip Ozkan; Ayfer Kanpolat
Journal:  J Cataract Refract Surg       Date:  2006-06       Impact factor: 3.351

6.  Pentacam pachometry: comparison with non-contact specular microscopy on the central cornea and inter-session repeatability on the peripheral cornea.

Authors:  Andrew K C Lam; Davie Chen
Journal:  Clin Exp Optom       Date:  2007-03       Impact factor: 2.742

7.  A clinical comparison of optical and ultrasonic pachometry.

Authors:  J Nissen; J O Hjortdal; N Ehlers; K Frost-Larsen; T Sørensen
Journal:  Acta Ophthalmol (Copenh)       Date:  1991-10

8.  Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry.

Authors:  V Yaylali; S C Kaufman; H W Thompson
Journal:  J Cataract Refract Surg       Date:  1997-11       Impact factor: 3.351

9.  Comparison of three methods of measuring corneal thickness and anterior chamber depth.

Authors:  Wolf Buehl; Danijela Stojanac; Stefan Sacu; Wolfgang Drexler; Oliver Findl
Journal:  Am J Ophthalmol       Date:  2006-01       Impact factor: 5.258

10.  Epithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF).

Authors:  H F Li; W M Petroll; T Møller-Pedersen; J K Maurer; H D Cavanagh; J V Jester
Journal:  Curr Eye Res       Date:  1997-03       Impact factor: 2.424

View more
  17 in total

1.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

3.  Measurement of central corneal thickness in health and disease.

Authors:  Imtiaz A Chaudhry
Journal:  Saudi J Ophthalmol       Date:  2009-10

4.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

5.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

6.  Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Authors:  Zaina N Al-Mohtaseb; Li Wang; Mitchell P Weikert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-17       Impact factor: 3.117

7.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

8.  Day to Day Clinically Relevant Corneal Elevation, Thickness, and Curvature Parameters Using the Orbscan II Scanning Slit Topographer and the Pentacam Scheimpflug Imaging Device.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01

9.  Measurement of central corneal thickness and pre-corneal tear film thickness of rabbits using the Scheimpflug system.

Authors:  Jing Dong; Qiang Wu; Xiao-Gang Wang
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

10.  Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:  Saleh Al-Ageel; Abdulrahman M Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-10-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.