Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam.

Clare O'Donnell1, Carole Maldonado-Codina.   

Abstract

PURPOSE: To compare the accuracy and repeatability of the OCULUS Pentacam (a new Scheimpflug-based imaging system) with ultrasound pachymetry in the measurement of central corneal thickness (CCT).
METHODS: CCT was measured in 21 subjects (21 normal corneas) on 2 separate occasions by the same examiner, using an Allergan-Humphrey 850 ultrasonic pachymeter and an OCULUS Pentacam instrument.
RESULTS: Mean values of CCT for both visits for each instrument were 534 +/- 47 microm and 528 +/- 45 microm using the ultrasonic pachymeter and the Pentacam, respectively. Plots of differences against means displayed relatively good agreement (limits of agreement were -13.0 to +26.6 microm). The repeatability (limits of agreement) of the ultrasound pachymeter was -18.3 to +17.7 microm, while for the OCULUS Pentacam it was -24.1 to +21.1 microm.
CONCLUSIONS: Our data showed that the Pentacam instrument provided measurements that were slightly but systematically lower than the measurements provided by ultrasonic pachymetry, which is currently the clinical gold standard method. The results, coupled with a unique ability to image and analyze the anterior chamber in vivo, make the OCULUS Pentacam a promising new instrument for anterior eye evaluation.

Mesh:

Year:  2005        PMID: 16227833     DOI: 10.1097/01.ico.0000157422.01146.e9

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  41 in total

1.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

Authors:  Domenico Schiano Lomoriello; Marco Lombardo; Laura Tranchina; Francesco Oddone; Sebastiano Serrao; Pietro Ducoli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

Review 3.  Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?

Authors:  Aristides Konstantopoulos; Parwez Hossain; David F Anderson
Journal:  Br J Ophthalmol       Date:  2007-04       Impact factor: 4.638

4.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

5.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

6.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

7.  Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery.

Authors:  Jumpei Matsuda; Osamu Hieda; Shigeru Kinoshita
Journal:  Jpn J Ophthalmol       Date:  2008-09-05       Impact factor: 2.447

8.  [Influence of corneal pachymetric changes on functional results after cataract surgery].

Authors:  F N Auerbach; M P Holzer; G U Auffarth; R Khoramnia; B C Thomas; D Saure; T M Rabsilber
Journal:  Ophthalmologe       Date:  2015-10       Impact factor: 1.059

9.  Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Authors:  Zaina N Al-Mohtaseb; Li Wang; Mitchell P Weikert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-17       Impact factor: 3.117

10.  Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis.

Authors:  Rajeev Jain; Grewal Dilraj; Satinder Pal Singh Grewal
Journal:  Indian J Ophthalmol       Date:  2007 Sep-Oct       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.