| Literature DB >> 31878803 |
Maddalena De Bernardo1, Palmiro Cornetta1, Giuseppe Marotta1, Giulio Salerno1, Ilaria De Pascale1, Nicola Rosa1.
Abstract
OBJECTIVE: To compare corneal thickness (CT) measurements using the CEM-530 (Nidal, Gamagori, Japan) and Pentacam HR (Oculus, Wetzlar, Germany).Entities:
Keywords: CEM-530; Cornea; Pentacam HR; corneal thickness; regression formula; specular microscopy
Mesh:
Year: 2019 PMID: 31878803 PMCID: PMC7783267 DOI: 10.1177/0300060519892385
Source DB: PubMed Journal: J Int Med Res ISSN: 0300-0605 Impact factor: 1.671
Corneal thickness measurements obtained with Pentacam HR and Nidek CEM 530.
|
|
| |||
|---|---|---|---|---|
| Apex (µm) | Pupil center (µm) | Thinnest point (µm) | Corneal center (µm) | |
| Mean | 552.49 | 551.03 | 546.05 | 558.49 |
| Standard deviation | 23.44 | 23.58 | 23.37 | 25.02 |
| Minimum | 505 | 504 | 503 | 503 |
| Maximum | 614 | 612 | 600 | 622 |
| 95% limits of agreement | −9.83 to +22.01 | −9.34 to +24.26 | −7.25 to +32.12 | |
Figure 1.Bland–Altman correlation between corneal thickness measurements in microns obtained with the Nidek CEM-530 and Pentacam HR at the corneal apex.
Figure 2.Bland–Altman correlation between corneal thickness measurements in microns obtained with the Nidek CEM-530 and Pentacam HR at the pupil center.
Figure 3.Bland–Altman correlation between corneal thickness measurements in microns obtained with the Nidek CEM-530 and Pentacam HR at the thinnest point.