Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound.

D Gherghel1, S L Hosking, S Mantry, S Banerjee, S A Naroo, S Shah.   

Abstract

PURPOSE: To compare corneal thickness measurements using Orbscan II (OII) and ultrasonic (US) pachymetry in normal and in keratoconic eyes.
SETTING: Eye Department, Heartlands and Solihull NHS Trust, Birmingham, United Kingdom.
METHODS: Central corneal thickness (CCT) was measured by means of OII and US pachymetry in 1 eye of 72 normal subjects and 36 keratoconus patients. The apical corneal thickness (ACT) in keratoconus patients was also evaluated using each method. The mean of the difference, standard deviation (SD), and 95% limits of agreement (LoA = mean +/- 2 SD), with and without applying the default linear correction factor (LCF), were determined for each sample. The Student t test was used to identify significant differences between methods, and the correlation between methods was determined using the Pearson bivariate correlation. Bland-Altman analysis was performed to confirm that the results of the 2 instruments were clinically comparable.
RESULTS: In normal eyes, the mean difference (+/- 95% LoA) in CCT was 1.04 microm +/- 68.52 (SD) (P>.05; r = 0.71) when the LCF was used and 46.73 +/- 75.40 microm (P =.0001; r = 0.71) without the LCF. In keratoconus patients, the mean difference (+/- 95% LoA) in CCT between methods was 42.46 +/- 66.56 microm (P<.0001: r = 0.85) with the LCF, and 2.51 +/- 73.00 microm (P>.05: r = 0.85) without the LCF. The mean difference (+/- 95% LoA) in ACT for this group was 49.24 +/- 60.88 microm (P<.0001: r = 0.89) with the LCF and 12.71 +/- 68.14 microm (P =.0077; r = 0.89) when the LCF was not used.
CONCLUSIONS: This study suggests that OII and US pachymetry provide similar readings for CCT in normal subjects when an LCF is used. In keratoconus patients, OII provides a valid clinical tool for the noninvasive assessment of CCT when the LCF is not applied.

Entities:  

Mesh:

Year:  2004        PMID: 15177603     DOI: 10.1016/j.jcrs.2003.11.049

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  19 in total

Review 1.  Biomechanics and wound healing in the cornea.

Authors:  William J Dupps; Steven E Wilson
Journal:  Exp Eye Res       Date:  2006-05-23       Impact factor: 3.467

2.  A novel insight into keratoconus: mechanical fatigue of the cornea.

Authors:  Amir Norouzpour; Alireza Mehdizadeh
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2012

3.  Epithelial, stromal, and total corneal thickness in keratoconus: three-dimensional display with artemis very-high frequency digital ultrasound.

Authors:  Dan Z Reinstein; Marine Gobbe; Timothy J Archer; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2010-04-07       Impact factor: 3.573

4.  Corneal pachymetry mapping with high-speed optical coherence tomography.

Authors:  Yan Li; Raj Shekhar; David Huang
Journal:  Ophthalmology       Date:  2006-05       Impact factor: 12.079

5.  Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.

Authors:  Roxana Ursea; Matthew Feng; Raksha Urs; Arindam RoyChoudhury; Ronald H Silverman
Journal:  J Refract Surg       Date:  2012-11-30       Impact factor: 3.573

6.  Keratoconus diagnosis with optical coherence tomography–based pachymetric scoring system.

Authors:  Bing Qin; Shihao Chen; Robert Brass; Yan Li; Maolong Tang; Xinbo Zhang; Xiaoyu Wang; Qinmei Wang; David Huang
Journal:  J Cataract Refract Surg       Date:  2013-12       Impact factor: 3.351

7.  Keratoconus diagnosis with optical coherence tomography pachymetry mapping.

Authors:  Yan Li; David M Meisler; Maolong Tang; Ake T H Lu; Vishakha Thakrar; Bibiana J Reiser; David Huang
Journal:  Ophthalmology       Date:  2008-11-05       Impact factor: 12.079

8.  Effect of central corneal thickness on Goldmann applanation tonometry measures - a different result with different pachymeters.

Authors:  Michael J Doughty; Sven Jonuscheit
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-05-24       Impact factor: 3.117

9.  Comparison of central corneal thickness with four different optical devices.

Authors:  Kuddusi Teberik; Mehmet Tahir Eski; Murat Kaya; Handan Ankaralı
Journal:  Int Ophthalmol       Date:  2017-10-11       Impact factor: 2.031

10.  Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system.

Authors:  Tania Paul; Mira Lim; Christopher E Starr; Harriet O Lloyd; D Jackson Coleman; Ronald H Silverman
Journal:  J Cataract Refract Surg       Date:  2008-11       Impact factor: 3.351

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.