Pentacam pachometry: comparison with non-contact specular microscopy on the central cornea and inter-session repeatability on the peripheral cornea.

Andrew K C Lam1, Davie Chen.   

Abstract

PURPOSE: Our aim was to compare specular microscopy (SM) and a new corneal topographer with rotating Scheimpflug camera (Pentacam) for non-contact pachometry of the central cornea. The repeatability of Pentacam for topographic pachometry was also studied.
METHODS: Thirty-nine subjects were recruited and one eye was selected randomly for non-contact pachometry by SM and by Pentacam, in random order. The corneal thickness (CT) was monitored at 30-second intervals for 10 minutes. Baseline CT was defined as the average of all measurements taken over the 10 minutes. Subjects were required to return within one to two weeks for a second corneal evaluation by Pentacam. Comparisons were made on the central CT between the two devices and on the topographic CT from Pentacam between the two visits.
RESULTS: The spontaneous variation of central CT was similar for the two instruments. Central CT varied within +/-2.3 microm during the 10 minutes. The 95% limits of agreement between SM and Pentacam were within +/-15 microm for the central CT. The spontaneous variation of peripheral CT was within +/-3.1 microm. There was no significant difference (p > 0.05) between the two visits on CT at different regions. Further analysis found that with Pentacam three measurements should be taken to obtain topographic CT measurement of one per cent precision.
CONCLUSION: Non-contact specular microscopy and Pentacam have good agreement for central CT measurement. Topographic pachometry from Pentacam requires three repeated measurements for one per cent precision.

Mesh:

Year:  2007        PMID: 17311572     DOI: 10.1111/j.1444-0938.2006.00103.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  6 in total

1.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

2.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

3.  Corneal elevation topography: best fit sphere, elevation distance, asphericity, toricity, and clinical implications.

Authors:  Damien Gatinel; Jacques Malet; Thanh Hoang-Xuan; Dimitri T Azar
Journal:  Cornea       Date:  2011-05       Impact factor: 2.651

4.  Analysis of continuous 24-hour intraocular pressure patterns in glaucoma.

Authors:  Kaweh Mansouri; John H K Liu; Robert N Weinreb; Ali Tafreshi; Felipe A Medeiros
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-12-13       Impact factor: 4.799

5.  Measurement of corneal thickness using Pentacam HR versus Nidek CEM-530 specular microscopy.

Authors:  Maddalena De Bernardo; Palmiro Cornetta; Giuseppe Marotta; Giulio Salerno; Ilaria De Pascale; Nicola Rosa
Journal:  J Int Med Res       Date:  2019-12-26       Impact factor: 1.671

6.  Limbal Metabolic Support Reduces Peripheral Corneal Edema with Contact-Lens Wear.

Authors:  Young Hyun Kim; Meng C Lin; Clayton J Radke
Journal:  Transl Vis Sci Technol       Date:  2020-06-30       Impact factor: 3.283

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.