Pentacam vs SP3000P specular microscopy in measuring corneal thickness.

Maddalena De Bernardo1, Maria Borrelli2, Marco Mariniello1, Michele Lanza2, Nicola Rosa3.   

Abstract

PURPOSE: To compare corneal thickness (CT) measurements obtained with SP3000P specular microscope and Pentacam in eyes screened to undergo refractive surgery (RS).
METHODS: In this prospective study, a non-randomized consecutive series of 73 healthy eyes (age range: 14-78 years; mean=38.4±14.9) underwent CT measurement with both Oculus Pentacam and Topcon SP3000P specular microscope to assess their suitability to undergo corneal RS.
RESULTS: CT measurements with SP3000P ranged from 451 to 609 μs (mean 523.1±34.4 μs). The measurements obtained with Pentacam ranged at the corneal apex from 477 to 672 μs (mean 558.9±38.9 μs), at the pupil center from 477 to 672 μs (mean 557.9±38.9 μs) and at the thinnest point from 474 to 669 μs (mean 551.1±39.4 μs). CT measurements showed a good correlation but a statistically significant difference at the pupil center (mean -34.9±14.6 μs, R(2)=0.860, p<0.001), at the apex (mean -35.9±14.9 μs, R(2)=0.856, p<0.001) and at the thinnest point (mean -32±14.8 μs, R(2)=0.862, p<0.001). The calculated regression formulas were: y=0.816x+66.94 for the apex, y=0.819x+66.07 for the pupil center and y=0.810x+73.13 for the thinnest point; where y is the CT measured with the SP3000P and x is the measurement obtained with the Oculus Pentacam.
CONCLUSIONS: Our results suggest that SP3000P measures thinner corneas compared to the Pentacam and that the correcting factor we identified should be applied to make comparisons between these two devices.
Copyright © 2014 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Corneal thickness; Scheimpflug camera; Specular microscopy

Mesh:

Year:  2014        PMID: 25240777     DOI: 10.1016/j.clae.2014.08.006

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  7 in total

1.  Central corneal thickness determination in corneal edema.

Authors:  Nicola Rosa; Maddalena De Bernardo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-03-15       Impact factor: 3.117

2.  Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.

Authors:  Maddalena De Bernardo; Nicola Rosa
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-02-03       Impact factor: 3.117

3.  Central corneal thickness measurement with Cirrus HD-OCT and Topcon SP-3000P.

Authors:  Maddalena De Bernardo; Francesco Antonio Salzano; Nicola Rosa
Journal:  J Optom       Date:  2017-12-14

4.  Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer.

Authors:  Maddalena De Bernardo; Maria Borrelli; Giovanni Cembalo; Nicola Rosa
Journal:  Medicina (Kaunas)       Date:  2019-10-18       Impact factor: 2.430

5.  Corneal thickness evaluation in healthy eyes: Comparison between two different Scheimpflug devices.

Authors:  Nicola Rosa; Maddalena De Bernardo; Angela Pepe; Livio Vitiello; Giuseppe Marotta; Roberto Imparato; Luigi Capasso
Journal:  PLoS One       Date:  2020-12-03       Impact factor: 3.240

6.  Measurement of corneal thickness using Pentacam HR versus Nidek CEM-530 specular microscopy.

Authors:  Maddalena De Bernardo; Palmiro Cornetta; Giuseppe Marotta; Giulio Salerno; Ilaria De Pascale; Nicola Rosa
Journal:  J Int Med Res       Date:  2019-12-26       Impact factor: 1.671

7.  Calculation of Unknown Preoperative K Readings in Postrefractive Surgery Patients.

Authors:  Nicola Rosa; Maddalena De Bernardo; Maria Borrelli
Journal:  J Ophthalmol       Date:  2018-02-11       Impact factor: 1.909

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.