Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomography.

Dexter Y L Leung1, Douglas K T Lam, Barry Y M Yeung, Dennis S C Lam.   

Abstract

PURPOSE: Measurement of central corneal thickness (CCT) plays an important role in both diagnostic and therapeutic assessment of ocular diseases. Although ultrasound pachymetry (U-PACH) is regarded as the golden standard for measurement of CCT, optical coherence tomography (OCT) may offer advantages as it can locate the central cornea with precision with no corneal touch. Nevertheless, the agreement of OCT with U-PACH has not yet been gauged by Bland-Altman analysis. This study compares CCT measurement by OCT with that by U-PACH.
METHODS: Healthy subjects without ocular abnormality (except refractive errors less than or equal to -6.0 D), contact lens wear or ocular surgery were recruited. CCT was measured in one eye of normal subjects using OCT and U-PACH. Results were compared using correlation and Bland-Altman plots.
RESULTS: Fifty subjects were recruited. Mean +/- SD CCT measured by OCT was 565 +/- 33 microm. This was highly correlated (Pearson's coefficient = 0.934) with the mean thickness measured by U-PACH (543 +/- 33 microm). The coefficients of variation were good and comparable at 7.9% for U-PACH and 3.5% for OCT. Compared with U-PACH, OCT consistently overestimated the CCT by a mean of 23 microm as shown on Bland-Altman plot.
CONCLUSION: CCT measured by OCT and U-PACH is highly correlated. With appropriate adjustment factor, OCT agrees well with U-PACH and is a reliable alternative for CCT measurement.

Entities:  

Mesh:

Year:  2006        PMID: 17073897     DOI: 10.1111/j.1442-9071.2006.01343.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  8 in total

1.  Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry.

Authors:  Hanna Y Kim; Donald L Budenz; Pak S Lee; William J Feuer; Keith Barton
Journal:  Am J Ophthalmol       Date:  2007-12-11       Impact factor: 5.258

2.  Central corneal thickness in adult Chinese. Association with ocular and general parameters. The Beijing Eye Study.

Authors:  Haitao Zhang; Liang Xu; Changxi Chen; Jost B Jonas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

3.  Multimodal Assessment of Corneal Thinning Using Optical Coherence Tomography, Scheimpflug Imaging, Pachymetry, and Slit-Lamp Examination.

Authors:  Julius T Oatts; Jeremy D Keenan; Tova Mannis; Tom M Lietman; Jennifer Rose-Nussbaumer
Journal:  Cornea       Date:  2017-04       Impact factor: 2.651

4.  Anterior Segment Optical Coherence Tomography and its Clinical Applications in Glaucoma.

Authors:  Haitao Li; Vishal Jhanji; Syril Dorairaj; Andrea Liu; Dennis Sc Lam; Christopher K Leung
Journal:  J Curr Glaucoma Pract       Date:  2012-08-16

5.  Measurement of corneal thickness using Pentacam HR versus Nidek CEM-530 specular microscopy.

Authors:  Maddalena De Bernardo; Palmiro Cornetta; Giuseppe Marotta; Giulio Salerno; Ilaria De Pascale; Nicola Rosa
Journal:  J Int Med Res       Date:  2019-12-26       Impact factor: 1.671

6.  Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry.

Authors:  Ilona Piotrowiak; Beata Soldanska; Mateusz Burduk; Bartlomiej J Kaluzny; Jozef Kaluzny
Journal:  ISRN Ophthalmol       Date:  2012-09-02

7.  Scaling and maintenance of corneal thickness during aging.

Authors:  Takenori Inomata; Alireza Mashaghi; Jiaxu Hong; Takeshi Nakao; Reza Dana
Journal:  PLoS One       Date:  2017-10-06       Impact factor: 3.240

8.  Repeatability and reproducibility of post-mortem central corneal thickness measurements using a portable optical coherence tomography system in humans: a prospective multicenter study.

Authors:  Pietro Emanuele Napoli; Matteo Nioi; Letizia Gabiati; Michela Laurenzo; Fabio De-Giorgio; Vincenzo Scorcia; Simone Grassi; Ernesto d'Aloja; Maurizio Fossarello
Journal:  Sci Rep       Date:  2020-09-02       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.