Central corneal thickness measurements with different imaging devices and ultrasound pachymetry.

Lai-Yong Tai1, Keat-Ween Khaw, Choung-Min Ng, Visvaraja Subrayan.   

Abstract

PURPOSE: The aim of this study was to compare 4 methods of central corneal thickness (CCT) measurements in terms of their agreement, repeatability, and measurement time.
METHODS: CCT was measured in 184 eyes of 92 healthy subjects by the same examiner. The methods used were as follows: noncontact specular microscopy (SM; Topcon SP-3000P; Topcon Corporation, Tokyo, Japan), Pentacam rotating Scheimpflug photography system (Oculus Inc, Wetzlar, Germany), optical low-coherence reflectometry (OLCR; LenStar LS900; Haag-Streit AG, Koeniz, Switzerland), and ultrasound pachymetry (UP; PachPen; Accutome Inc, Malvern, PA). The duration for each examination was measured by an independent observer.
RESULTS: The mean age (±SD) of the subjects was 54.3 (± 15.3) years. The mean CCTs (± SD) for SM, Pentacam, OLCR, and UP were 507.8 (± 30.2), 538.4 (± 31.7), 531.8 (± 31.4), and 528.3 (± 32.9) μm, respectively. The Bland-Altman plots showed closest agreement for OLCR-UP, followed by OLCR-Pentacam and Pentacam-UP. SM had the poorest agreement with the other methods. CCTs measured by SM were on average 20 to 30 μm thinner than those of the other methods. The coefficient of repeatability for SM, Pentacam, OLCR, and UP were 3.14%, 4.23%, 1.51%, and 3.46%, respectively. The mean measurement times (± SD) were 13.5 (± 5.7), 45.7 (± 12.3), 18.5 (± 7.1), and 5.6 (± 1.0) seconds, respectively.
CONCLUSIONS: CCT measurements between OLCR-UP and OLCR-Pentacam are comparable and can be used interchangeably in clinical practice. However, SM underestimates CCT compared with the other methods, whereas Pentacam was found to be the least repeatable and took the longest time.

Mesh:

Year:  2013        PMID: 23095499     DOI: 10.1097/ICO.0b013e318269938d

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  25 in total

1.  Letter to the editor. "Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery".

Authors:  Alberto López-Miguel; María Sanchidrián; Itziar Fernández; Alfredo Holgueras; Miguel J Maldonado
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-02-11       Impact factor: 3.117

2.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

3.  A multimodal ophthalmic analysis in patients with systemic sclerosis using ocular response analyzer, corneal topography and specular microscopy.

Authors:  Huseyin Mayali; Muhammed Altinisik; Secil Sencan; Timur Pirildar; Emin Kurt
Journal:  Int Ophthalmol       Date:  2019-09-28       Impact factor: 2.031

4.  Measurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices.

Authors:  Ozgur Cakici; Remzi Karadag; Huseyin Bayramlar; Efe Koyun
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

5.  Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.

Authors:  Alberto López-Miguel; María Sanchidrián; Itziar Fernández; Alfredo Holgueras; Miguel J Maldonado
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-11-22       Impact factor: 3.117

6.  Evaluation of Corvis ST tonometer with the updated software in glaucoma practice.

Authors:  Ioannis Halkiadakis; Vasilios Tzimis; Alexandros Gryparis; Ioannis Markopoulos; Vasiliki Konstadinidou; Elias Zintzaras; Michalis Tzakos
Journal:  Int J Ophthalmol       Date:  2022-03-18       Impact factor: 1.779

7.  Monitoring of central corneal thickness after phacoemulsification-comparison of statical and rotating Scheimpflug pachymetry, and spectral-domain OCT.

Authors:  Daniel M Handzel; Carsten H Meyer; Alfred Wegener
Journal:  Int J Ophthalmol       Date:  2022-08-18       Impact factor: 1.645

8.  Decreased central corneal thickness in ankylosing spondylitis.

Authors:  Huseyin Ortak; Ahmet Inanır; Selim Demir; Alper Uysal; Şafak Şahin; Mustafa Sağcan; Yalçın Önder; Sait Alim; Ayşe Kevser Demir
Journal:  Int Ophthalmol       Date:  2013-07-12       Impact factor: 2.031

9.  Comparison of central corneal thickness with four different optical devices.

Authors:  Kuddusi Teberik; Mehmet Tahir Eski; Murat Kaya; Handan Ankaralı
Journal:  Int Ophthalmol       Date:  2017-10-11       Impact factor: 2.031

Review 10.  The Future Is Now: Incorporating Telemedicine into Glaucoma Care.

Authors:  Monica K Ertel; Malik Y Kahook; Cara E Capitena Young
Journal:  Curr Ophthalmol Rep       Date:  2021-07-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.