Central corneal thickness measured with three optical devices and ultrasound pachometry.

Javier González-Pérez1, José Manuel González-Méijome, María Teresa Rodríguez Ares, Manuel Angel Parafita.   

Abstract

PURPOSE: To evaluate the reliability of three noninvasive pachometry methods against the ultrasound pachometer considered the gold standard.
METHODS: Central corneal thickness (CCT) was measured using a Paxis ultrasound (US) pachometer, Orbscan II, Pentacam, and Topcon SP-3000 specular microscope in 22 right eyes of 22 healthy adults (7 men, 15 women). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between pachometers.
RESULTS: Pentacam is the optical system that renders values of CCT closer to those obtained with US pachometry. Conversely, a specular microscope shows a poorer agreement with US measures, and differences depend on the thickness being measured. The Orbscan system shows a consistent trend toward underestimation of CCT compared with US and Pentacam irrespective of the value measured.
CONCLUSION: Different optical methods used to measure CCT showed significantly different results compared with US pachometry except for the Pentacam system. Clinicians should take into account the fact that specular microscopy might either underestimate or overestimate values of CCT, whereas Orbscan systematically overestimates CCT compared with US and Pentacam.

Entities:  

Mesh:

Year:  2011        PMID: 21301351     DOI: 10.1097/ICL.0b013e31820c6ffc

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  18 in total

1.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

2.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

3.  Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

Authors:  Mohammad Naderan; Saeed Shoar; Morteza Naderan; Mohammad Amin Kamaleddin; Mohammad Taher Rajabi
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

4.  Lamina cribrosa thickness is not correlated with central corneal thickness or axial length in healthy eyes: central corneal thickness, axial length, and lamina cribrosa thickness.

Authors:  Eun Ji Lee; Tae-Woo Kim; Robert N Weinreb; Min Hee Suh; Hyunjoong Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-09-20       Impact factor: 3.117

5.  A multimodal ophthalmic analysis in patients with systemic sclerosis using ocular response analyzer, corneal topography and specular microscopy.

Authors:  Huseyin Mayali; Muhammed Altinisik; Secil Sencan; Timur Pirildar; Emin Kurt
Journal:  Int Ophthalmol       Date:  2019-09-28       Impact factor: 2.031

6.  Measurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices.

Authors:  Ozgur Cakici; Remzi Karadag; Huseyin Bayramlar; Efe Koyun
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

7.  Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness.

Authors:  Kelechi C Ogbuehi; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2012-05-21

8.  Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound.

Authors:  J Jorge; Jl Rosado; Ja Díaz-Rey; Jm González-Méijome
Journal:  Clin Ophthalmol       Date:  2013-02-26

9.  Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes.

Authors:  Mohammad Mehdi Sadoughi; Bahram Einollahi; Neda Einollahi; Javad Rezaei; Danial Roshandel; Sepehr Feizi
Journal:  J Ophthalmic Vis Res       Date:  2015 Jan-Mar

10.  Customized Finite Element Modelling of the Human Cornea.

Authors:  Irene Simonini; Anna Pandolfi
Journal:  PLoS One       Date:  2015-06-22       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.