Lisa Ramm1, Robert Herber1, Eberhard Spoerl1, Frederik Raiskup1, Lutz E Pillunat1, Naim Terai1.
Abstract
PURPOSE: To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST (CST) in healthy subjects.Entities:
Year: 2019 PMID: 31737355 PMCID: PMC6815996 DOI: 10.1155/2019/3879651
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Descriptive data of participants and results of intraocular pressure (IOP) measurements (mean value ± standard deviation).
| Descriptive data | |
|---|---|
| Number of subjects | 94 |
| Eye (right/left) | 47 (50%)/47 (50%) |
| Gender (female/male) | 39 (41.5%)/55 (58.5%) |
| Mean age (years) | 45.6 ± 17.2 |
| CCT ( | 553 ± 33 |
| Intraocular pressure | |
| GAT-IOPc (mmHg) | 12.9 ± 2.4 |
| DCT-IOP (mmHg) | 16.1 ± 2.6 |
| IOPcc (mmHg) | 15.6 ± 3.3 |
| bIOP (mmHg) | 13.5 ± 2.4 |
GAT-IOPc: corrected Goldmann applanation tonometry-IOP; DCT-IOP: dynamic contour tonometry-IOP; IOPcc: corneal compensated IOP by using ocular response analyzer; bIOP: biomechanical corrected IOP by using Corvis ST; CCT: central corneal thickness.
Amount of mean difference (±standard deviation, in mmHg) of the results of intraocular pressure (IOP) measurement using different tonometers.
| DCT-IOP | IOPcc | bIOP | |
|---|---|---|---|
| GAT-IOPc | 3.2 ± 2.3, | 2.7 ± 3, | 0.6 ± 2.7, |
| DCT-IOP | — | 0.5 ± 2.4, | 2.6 ± 2.5, |
| IOPcc | — | — | 2.1 ± 2.9, |
GAT-IOPc: corrected Goldmann applanation tonometry-IOP, DCT-IOP: dynamic contour tonometry-IOP, IOPcc: corneal compensated IOP by using ocular response analyzer, bIOP: biomechanical corrected IOP by using Corvis ST.
Figure 1Correlation analysis between central corneal thickness (CCT) and intraocular pressure (IOP) measurements using (a) Goldmann applanation tonometry (GAT-IOPc), (b) dynamic contour tonometry (DCT-IOP), (c) ocular response analyzer (IOPcc) and (d) Corvis ST (bIOP); r = Pearson correlation coefficient.
Figure 2Bland–Altman analyses between mean intraocular pressure (IOP in mmHg) measured by (a) Goldmann applanation tonometry (GAT-IOPc) and dynamic contour tonometry (DCT-IOP), (b) ocular response analyzer (IOPcc) and (c) Corvis ST (bIOP).
Figure 3Correlation analysis between age and intraocular pressure (IOP) measurements using (a) Goldmann applanation tonometry (GAT-IOPc), (b) dynamic contour tonometry (DCT-IOP), (c) ocular response analyzer (IOPcc), and (d) Corvis ST (bIOP); r = Pearson correlation coefficient.