María Iglesias1, Bachar Kudsieh2, Andrea Laiseca3, Cristina Santos4, Jeroni Nadal3, Rafael Barraquer3,5, Ricardo P Casaroli-Marano6.
Abstract
BACKGROUND: The purpose of this study is to describe measurements using a newly developed modified Goldmann convex tonometer (CT) 1 year after myopic laser refractive surgery. Intraocular pressure (IOP) measurements were compared with IOP values obtained by Goldmann applanation tonometer (GAT), and Ocular Response Analyzer (ORA).Entities:
Keywords: Corneal biomechanics; Glaucoma; Intraocular pressure; LASIK; Myopia; PRK; Tonometry
Mesh:
Substances:
Year: 2022 PMID: 35168601 PMCID: PMC8849021 DOI: 10.1186/s12886-022-02309-x
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1Modified CT applanation tonometer. Observe the convex reflection in the tip (A) and in the side image (B, *) CT tonometer inserted in a handling tonometer device (B)
Descriptive statistics of variables in the evaluations before and 3 and 12 months after surgery
| PRE-SURGERY | POST 3 M | POST 12 M | |
|---|---|---|---|
| GAT (mmHg)a | |||
| • ALL | 15.80 ± 2.72 | 12.00 ± 2.32 | 13.04 ± 2.55 |
| • LASIK | 15.60 ± 3.27 | ||
| • PRK | 16.01 ± 1.45 | ||
| CT (mmHg)a | |||
| • ALL | 21.07 ± 4.35 | 15.80 ± 3.07 | 16.30 ± 3.34 |
| • LASIK | 21.05 ± 4.59 | ||
| • PRK | 22.00 ± 4.01 | ||
| IOPcc (mmHg)a | |||
| • ALL | 15.89 ± 3.07 | 13.07 ± 2.52 | 13.06 ± 2.40 |
| • LASIK | 16.40 ± 2.93 | ||
| • PRK | 15.02 ± 3.26 | 14.08 ± 2.83 | 13.38 ± 1.65 |
| IOPg (mmHg)a | |||
| • ALL | 15.49 ± 2.98 | 8.95 ± 2.79 | 9.62 ± 2.73 |
| • LASIK | 15.81 ± 2.77 | ||
| • PRK | 14.93 ± 3.37 | ||
| SER (dpts) | |||
| • ALL | − 4.98 ± 2.28 | 0.00 ± 0.30 | 0.02 ± 0.22 |
| • LASIK | −5.83 ± 2.21 | 0.05 ± 0.33 | 0.03 ± 0.25 |
| • PRK | −3.52 ± 1.33 | 0.05 ± 0.23 | 0.03 ± 0.15 |
| simK | |||
| • ALL | 43.70 ± 1.13 | 39.80 ± 1.96 | 39.90 ± 2.00 |
| • LASIK | 43.70 ± 1.16 | 39.30 ± 2.06 | 39.40 ± 2.04 |
| • PRK | 43.60 ± 1.11 | 40.60 ± 1.53 | 40.08 ± 1.62 |
| CCT (μ) | |||
| • ALL | 543 ± 27.08 | 457 ± 35.90 | 469 ± 41.30 |
| • LASIK | 554 ± 22.30 | 460 ± 40.70 | 466 ± 42.40 |
| • PRK | 524 ± 26.40 | 452 ± 26.60 | 473 ± 41.10 |
| CH | |||
| • ALL | 10.53 ± 1.28 | 7.88 ± 1.16 | 8.07 ± 1.03 |
| • LASIK | 10.44 ± 1.12 | 7.85 ± 1.10 | 8.04 ± 1.05 |
| • PRK | 10.67 ± 1.56 | 7.92 ± 1.32 | 8.11 ± 1.05 |
| CRF | |||
| • ALL | 9.88 ± 1.12 | 6.29 ± 1.40 | 6.59 ± 1.25 |
| • LASIK | 9.97 ± 1.09 | 6.05 ± 1.37 | 6.54 ± 1.20 |
| • PRK | 9.72 ± 1.21 | 6.70 ± 1.43 | 6.68 ± 1.39 |
| PTA | |||
| • ALL | 12.90 ± 4.73 | ||
| • LASIK | 14.40 ± 4.94 | ||
| • PRK | 10.30 ± 2.99 | ||
| Max.Abl | |||
| • ALL | 70.50 ± 26.80 | ||
| • LASIK | 79.90 ± 27.50 | ||
| • PRK | 54.10 ± 16.40 | ||
Significant differences between LASIK and PRK IOPm are marked in bold
Comparisons between pre- and post-surgery: a, Wilcoxon test, p < 0.001. All values are in mean ± standard deviation (MD ± SD). GAT Goldmann applanation tonometer, CT Convex tonometer, IOPcc Corneal-compensated IOP, IOPg Goldmann-correlated IOP, SER Spherical equivalent refraction, simK simulated keratometry, CCT Central corneal thickness, PTA Percent tissue altered, Max.Abl Maximum corneal ablation, CH Corneal hysteresis, CRF Corneal resistance factor, M Months
Fig. 2Mean IOP changes over time obtained using different tonometer devices in pre- and post-refractive surgery evaluations after 3 and 12 months. CX: L, LASIK (A); CX:P, PRK (B); GAT, Goldmann applanation tonometer; CT, convex tonometer; IOPg, Goldmann-correlated intraocular pressure and IOPcc, corneal-compensated IOP. M, months
Intraclass correlation coefficient (ICC) results
| Tonometer pair | N | ICC (95% CI) | ||
|---|---|---|---|---|
| All patients | GAT3M - GATpre | 30 | < 0.001 | |
| GAT12M - GATpre | 30 | < 0.006 | ||
| CT 3 M - GATpre | 30 | < 0.002 | ||
| CT 12 M - GATpre | 30 | < 0.001 | ||
IOPcc3M- CT3M | 30 | < 0.010 | ||
IOPcc12M - CT12M | 30 | < 0.003 | ||
| LASIK | GAT3M - GATpre | 19 | < 0.003 | |
| GAT12M - GATpre | 19 | < 0.015 | ||
| CT 3 M - GATpre | 19 | < 0.000 | ||
| CT 12 M - GATpre | 19 | < 0.000 | ||
IOPcc3M- CT3M | 19 | 0.405 (−0.252–0.748) | < 0.076 | |
IOPcc12M- CT12M | 19 | < 0.004 | ||
| PRK | GAT3M - GATpre | 11 | 0.134 (−0.234–0.590) | < 0.261 |
| GAT12M - GATpre | 11 | 0.262 (0.581–0.752) | < 0.251 | |
| CT 3 M - GATpre | 11 | −0.196 (−1.817–0.619) | < 0.638 | |
| CT 12 M - GATpre | 11 | −0.266 (−3.896–0.663) | < 0.642 | |
IOPcc3M- CT3M | 11 | 0.272 (− 0.328–0.729) | < 0.116 | |
IOPcc12M- CT12M | 11 | 0.256 (− 0.332–0.719) | < 0.182 |
ICC is based on absolute agreement. Significant differences are marked in bold. CI, confidence Interval; p, value for correlation
GAT Goldmann applanation tonometer, CT Convex tonometer, IOPg Goldmann-correlated IOP, IOPcc Corneal-compensated IOP, M Months
Mean differences for IOP measurements in pre-surgery with GAT and at 3 and 12 months after surgery with GAT, CT and IOPcc
| IOP difference | GAT3M - GATpre | GAT12M - | CT3M - GATpre | CT12M - GATpre | CT3M -IOPcc3M | CT12M -IOPcc12M | |
|---|---|---|---|---|---|---|---|
| All | md (bias) | −0.0333 | 0.533 | ||||
| Wilcoxon test | Z = -4.747; | Z = -3904; | Z = -0.361; | Z = -0.888; | Z = -3.754; | Z = -4.618; | |
LASIK GV | md (bias) | −1.11 | 0.158 | ||||
| Wilcoxon test | Z = -3.836; | Z = -3.209; | Z = -2.017; | Z = -0.776; | Z = -2.636; | Z = -3.582; | |
| PRK | md (bias) | 1.82 | 1.18 | ||||
| Wilcoxon test | Z = -2.185; | Z = -2.203; | Z = -1.616; | Z = -6.55; | Z = -2.667; | Z = -2.936; |
Mean differences between GATpre, GAT, IOPcc and CT after surgery (at 3 and 12 months) were calculated and tested for the absence of differences using the Wilcoxon test. Significant differences are marked in bold. md, mean difference; Z, Wilcoxon test; p, probability value
GAT Goldmann applanation tonometer, CT Convex tonometer, IOPg Goldmann-correlated IOP, IOPcc Corneal-compensated IOP, M Months