Goldmann applanation tonometry versus ocular response analyzer for intraocular pressure measurements in keratoconic eyes.

Yakov Goldich1, Yaniv Barkana, Isaac Avni, David Zadok.   

Abstract

PURPOSE: To compare intraocular pressure (IOP) measurements obtained with the Goldmann applanation tonometer (GAT) and the ocular response analyzer (ORA) in patients with keratoconus (KC) and analyze their dependence on ocular anatomic parameters.
METHODS: Patients with KC were recruited prospectively. IOP was measured using GAT and ORA. The ORA provided a Goldmann correlated IOP (IOPg) and a corneal correlated IOP (IOPcc). Assessment of refractive status, visual acuity, axial length, corneal topography, and pachymetry was done.
RESULTS: Fifty-nine eyes of 59 patients with KC (39 men, 20 women; mean age: 27.8 +/- 6.8 years) were included. The differences in mean IOP values between GAT (10.9 +/- 2.0 mm Hg) and IOPg (9.5 +/- 2.8 mm Hg) and between GAT and IOPcc (13.3 +/- 2.5 mm Hg) were statistically significant (all P < 0.001). Both pressure measurements provided by the ORA showed significant correlation with corneal curvature. No significant effect of corneal thickness on any of the pressures was observed.
CONCLUSIONS: IOP measurements taken with GAT and ORA in keratoconic eyes were significantly different. Although IOPcc was significantly higher, IOPg was significantly lower than GAT IOP. Unlike GAT measurements, ORA readings seemed to be affected mainly by corneal curvature. As a result of described differences, we suggest these devices should not be used interchangeably but rather in a complementary fashion to assess IOP in keratoconic eyes.

Entities:  

Mesh:

Year:  2010        PMID: 20539214     DOI: 10.1097/ICO.0b013e3181cda034

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  5 in total

1.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

2.  Which tonometry in eyes with keratoconus?

Authors:  H Altinkaynak; C Kocasarac; H Dundar; N Sayin; N Kara; E Bozkurt; N Duru
Journal:  Eye (Lond)       Date:  2015-12-04       Impact factor: 3.775

3.  Twenty-four hours intraocular pressure in keratoconic eyes assessed by applanation tonometry and Tono-Pen AVIA.

Authors:  Rafael Vidal Merola; Sebastião Cronemberger; Artur William Veloso; Alberto Diniz-Filho
Journal:  Int J Ophthalmol       Date:  2022-01-18       Impact factor: 1.779

4.  Collagen cross-linking: current status and future directions.

Authors:  Marine Hovakimyan; Rudolf F Guthoff; Oliver Stachs
Journal:  J Ophthalmol       Date:  2012-01-12       Impact factor: 1.909

5.  Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST.

Authors:  Lisa Ramm; Robert Herber; Eberhard Spoerl; Frederik Raiskup; Lutz E Pillunat; Naim Terai
Journal:  J Ophthalmol       Date:  2019-10-16       Impact factor: 1.909