Xiangjun Chen1, Aleksandar Stojanovic2, Yanjun Hua3, Jon Roger Eidet4, Di Hu5, Jingting Wang5, Tor Paaske Utheim6.
Abstract
PURPOSE: To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups.Entities:
Mesh:
Year: 2014 PMID: 25302580 PMCID: PMC4193795 DOI: 10.1371/journal.pone.0109577
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1The CorVis ST utilizes the Scheimpflug camera to record the dynamic procedure of the corneal response to an air puff.
A) The 1st applanation is achieved. B) The cornea reaches its highest concavity. C) The 2nd applanation is achieved when the cornea rebounds to its original position from the highest concavity.
Demographic Data of Participants.
| Virgin eyes (n = 40) | Post-PRK eyes (n = 42) |
| |
|
| 27.9±9.0 (18, 48) | 31.8±6.9 (20, 48) | 0.03 |
|
| 547.82±26.78 | preop 542.02±30.68 postop 485.00±40.10 | 0.30 |
|
| 15.20±2.57 | preop 15.81±3.29 postop 12.71±2.77 | 0.46 |
|
| −2.15±2.28 | preop −3.52±1.93 postop 0.01±0.48 | 0.03 |
|
| 43.47±1.38 | preop 43.81±1.58 postop 40.87±1.63 | 0.18 |
CCT = central corneal thickness; IOP = intraocular pressure; MRSE = manifest refraction spherical equivalent; simK = simulated karatometry.
* p values were adjusted for age-difference.
Intraobserver Repeatability of Parameters Obtained by Corvis in Virgin-Eye Group (n = 40).
| Parameters | Mean ± SD | Sw | 2.77Sw | COV (%) | ICC (95%CI) |
| IOP (mmHg) | 14.46±1.33 | 0.59 | 1.62 | 3.59 | 0.93 (0.89∼0.96) |
| CCT ( | 543.32±25.08 | 5.34 | 12.56 | 0.69 | 0.99 (0.98∼0.99) |
| 1st appl. time (ms) | 7.29±0.21 | 0.09 | 0.24 | 1.09 | 0.94 (0.90∼0.97) |
| 1st appl. length (mm) | 1.83±0.18 | 0.29 | 0.81 | 13.94 | 0.10 (−0.52∼0.49) |
| 1st appl. velocity (m/s) | 0.14±0.02 | 0.03 | 0.08 | 18.82 | 0.25 (−0.26∼0.57) |
| 2nd appl. time (ms) | 21.65±0.34 | 0.17 | 0.48 | 0.71 | 0.92 (0.87∼0.95) |
| 2nd appl. length (mm) | 1.89±0.29 | 0.45 | 1.24 | 21.53 | 0.17 (−0.39∼0.53) |
| 2nd appl. velocity (m/s) | −0.34±0.04 | 0.04 | 0.10 | −9.77 | 0.65 (0.42∼0.80) |
| Highest concavity time (ms) | 16.40±0.37 | 0.37 | 1.02 | 2.02 | 0.64 (0.40∼0.80) |
| Peak distance (mm) | 4.36±0.66 | 1.17 | 3.23 | 21.80 | −0.04 (−0.74∼0.41) |
| Radius of curvature (mm) | 7.49±0.60 | 0.55 | 1.51 | 6.31 | 0.70 (0.49∼0.83) |
| Deformation amplitude (mm) | 1.06±0.07 | 0.04 | 0.11 | 3.34 | 0.88 (0.81∼0.93) |
SD = standard deviation, ICC = intraclass correlation coefficient, CI = confidence interval, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, IOP = intraocular pressure, CCT = central corneal thickness.
Intraobserver Repeatability of Parameters Obtained by Corvis in Post-PRK Group (n = 42).
| Parameters | Mean ± SD | Sw | 2.77Sw | COV (%) | ICC (95%CI) |
| IOP (mmHg) | 12.38±1.02 | 0.55 | 1.52 | 3.85 | 0.90 (0.84∼0.94) |
| CCT (µm) | 481.18±42.45 | 8.16 | 22.61 | 2.29 | 0.99 (0.98∼0.99) |
| 1st appl. time (ms) | 6.96±0.17 | 0.09 | 0.25 | 1.11 | 0.91 (0.85∼0.95) |
| 1st appl. length (mm) | 1.82±0.22 | 0.36 | 0.99 | 17.65 | 0.08 (−0.54∼0.47) |
| 1st appl. velocity (m/s) | 0.13±0.02 | 0.03 | 0.09 | 23.30 | −0.27 (−1.14∼0.28) |
| 2nd appl. time (ms) | 21.96±0.31 | 0.18 | 0.49 | 0.69 | 0.89 (0.81∼0.94) |
| 2nd appl. length (mm) | 1.70±0.39 | 0.49 | 1.37 | 26.01 | 0.48 (0.12∼0.70) |
| 2nd appl. velocity (m/s) | −0.40±0.06 | 0.04 | 0.12 | −10.35 | 0.79 (0.65∼0.88) |
| Highest concavity time (ms) | 16.48±0.35 | 0.41 | 1.15 | 2.12 | 0.66 (0.43∼0.81) |
| Peak distance (mm) | 4.56±0.77 | 1.12 | 3.10 | 18.34 | 0.30 (−0.18∼0.60) |
| Radius of curvature (mm) | 6.43±0.66 | 0.70 | 1.94 | 6.76 | 0.63 (0.38∼0.79) |
| Deformation amplitude (mm) | 1.17±0.08 | 0.04 | 0.12 | 3.16 | 0.92 (0.86∼0.95) |
SD = standard deviation, ICC = intraclass correlation coefficient, CI = confidence interval, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, IOP = intraocular pressure, CCT = central corneal thickness.
Interobserver Reproducibility of Parameters Obtained by Corvis in Virgin-Eye Group (n = 40).
| Parameters | Mean Difference ± SD | Sw | 2.77Sw | COV (%) | ICC (95%CI) |
| IOP (mmHg) | −0.01±0.82 | 0.58 | 1.60 | 3.25 | 0.92 (0.86∼0.96) |
| CCT (µm) | −1.49±6.76 | 4.78 | 13.24 | 0.72 | 0.98 (0.97∼0.99) |
| 1st appl. time (ms) | 0.01±0.12 | 0.08 | 0.23 | 0.96 | 0.93 (0.88∼0.96) |
| 1st appl. length (mm) | −0.02±0.38 | 0.27 | 0.73 | 11.67 | 0.29 (−0.35(0.63) |
| 1st appl. velocity (m/s) | (0.005±0.05 | 0.03 | 0.09 | 19.10 | 0.08 ((0.75(0.51) |
| 2nd appl. time (ms) | (0.01±0.24 | 0.17 | 0.47 | 0.61 | 0.91 (0.82(0.95) |
| 2nd appl. length (mm) | (0.10±0.58 | 0.44 | 1.14 | 17.37 | 0.12 ((0.65(0.53) |
| 2nd appl. velocity (m/s) | (0.01±0.06 | 0.04 | 0.11 | (9.63 | 0.59 (0.25(0.78) |
| Highest concavity time (ms) | (0.02±0.54 | 0.38 | 1.06 | 1.81 | 0.47 (0.00(0.72) |
| Peak distance (mm) | 0.11±1.47 | 1.04 | 2.89 | 13.08 | 0.06 ((1.45(0.63) |
| Radius of curvature (mm) | 0.01±0.81 | 0.57 | 1.58 | 5.01 | 0.64 (0.31∼0.81) |
| Deformation amplitude (mm) | −0.002±0.06 | 0.04 | 0.11 | 1.95 | 0.88 (0.78∼0.94) |
SD = standard deviation, ICC = intraclass correlation coefficient, CI = confidence interval, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, IOP = intraocular pressure, CCT = central corneal thickness.
Interobserver Reproducibility of Parameters Obtained by Corvis in Post-PRK Group (n = 42).
| Parameters | Mean Difference ± SD | Sw | 2.77Sw | COV (%) | ICC (95%CI) |
| IOP (mmHg) | −0.17±0.70 | 0.50 | 1.38 | 2.81 | 0.90 (0.81∼0.95) |
| CCT (µm) | 0.43±5.05 | 3.57 | 9.89 | 0.58 | 1.00 (0.99∼1.00) |
| 1st appl. time (ms) | −0.02±0.11 | 0.08 | 0.23 | 0.84 | 0.90 (0.82∼0.95) |
| 1st appl. length (mm) | 0.08±0.46 | 0.32 | 0.90 | 14.45 | 0.27 (−0.36∼0.60) |
| 1st appl. velocity (m/s) | −0.01±0.04 | 0.03 | 0.09 | 17.54 | 0.45 (−0.03∼0.70) |
| 2nd appl. time (ms) | 0.04±0.28 | 0.20 | 0.55 | 0.64 | 0.79 (0.61∼0.89) |
| 2nd appl. length (mm) | −0.11±0.64 | 0.45 | 1.25 | 20.23 | 0.52 (0.12∼0.74) |
| 2nd appl. velocity (m/s) | 0.01±0.07 | 0.05 | 0.15 | −11.03 | 0.60 (0.26∼0.78) |
| Highest concavity time (ms) | 0.05±0.49 | 0.35 | 0.97 | 1.54 | 0.63 (0.31∼0.80) |
| Peak distance (mm) | 0.001±1.57 | 1.11 | 3.08 | 15.23 | 0.26 (−0.39∼0.61) |
| Radius of curvature (mm) | −0.06±0.49 | 0.35 | 0.97 | 4.27 | 0.83 (0.68∼0.91) |
| Deformation amplitude (mm) | −0.02±0.15 | 0.04 | 0.12 | 3.89 | 0.88 (0.78∼0.94) |
SD = standard deviation, ICC = intraclass correlation coefficient, CI = confidence interval, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, IOP = intraocular pressure, CCT = central corneal thickness.
Comparison of The CorVis ST Measurments Between Virgin-Eye and Post-PRK Groups.
| Virgin-eye (n = 40) | post-PRK (n = 42) | Difference (mean ± SE) |
| Adjusted | Adjusted R2 | |
|
| 14.46±1.33 | 12.38±1.02 | 2.08±0.26 | 0.000 | 0.002 | 0.520 |
|
| 543.32 ±25.08 | 485.41±39.00 | 57.90±7.28 | 0.000 | ||
|
| ||||||
|
| 7.29±0.21 | 6.96±0.17 | 0.33±0.04 | 0.000 | 0.003 | 0.519 |
|
| 1.83±0.18 | 1.81±0.22 | 0.01±0.04 | 0.832 | ||
|
| 0.14±0.02 | 0.13±0.02 | 0.01±0.00 | 0.010 | 0.614 | 0.160 |
|
| ||||||
|
| 21.65±0.34 | 21.96±0.31 | −0.31±0.07 | 0.000 | 0.032 | 0.221 |
|
| 1.89±0.29 | 1.70±0.39 | 0.19±0.08 | 0.013 | 0.958 | 0.133 |
|
| −0.34±0.04 | −0.40±0.06 | 0.07±0.01 | 0.000 | 0.053 | 0.372 |
|
| ||||||
|
| 16.40±0.37 | 16.48±0.41 | −0.07±0.09 | 0.374 | ||
|
| 4.36±0.66 | 4.56±0.77 | −0.20±0.16 | 0.216 | ||
|
| 7.49±0.60 | 6.43±0.66 | 1.06±0.14 | 0.000 | 0.001 | 0.551 |
|
| 1.06±0.07 | 1.17±0.08 | −0.10±0.02 | 0.000 | 0.005 | 0.402 |
SE = standard error, IOP = intraocular pressure, CCT = central corneal thickness.
p* values were adjusted for the effect of the age, CCT, mean simK difference between the virgin-eye and post-PRK groups.