Reliability of corneal dynamic scheimpflug analyser measurements in virgin and post-PRK eyes.

Xiangjun Chen1, Aleksandar Stojanovic2, Yanjun Hua3, Jon Roger Eidet4, Di Hu5, Jingting Wang5, Tor Paaske Utheim6.   

Abstract

PURPOSE: To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups.
METHODS: Forty virgin eyes and 42 post-PRK eyes underwent CorVis ST measurements performed by two technicians. Repeatability was evaluated by comparing three consecutive measurements by technician A. Reproducibility was determined by comparing the first measurement by technician A with one performed by technician B. Intraobserver and interobserver intraclass correlation coefficients (ICCs) were calculated. Univariate analysis of covariance (ANCOVA) was used to compare measured parameters between virgin and post-PRK eyes.
RESULTS: The intraocular pressure (IOP), central corneal thickness (CCT) and 1st applanation time demonstrated good intraobserver repeatability and interobserver reproducibility (ICC ≧ 0.90) in virgin and post-PRK eyes. The deformation amplitude showed a good or close to good repeatability and reproducibility in both groups (ICC ≧ 0.88). The CCT correlated positively with 1st applanation time (r = 0.437 and 0.483, respectively, p<0.05) and negatively with deformation amplitude (r = -0.384 and -0.375, respectively, p<0.05) in both groups. Compared to post-PRK eyes, virgin eyes showed longer 1st applanation time (7.29 ± 0.21 vs. 6.96 ± 0.17 ms, p<0.05) and lower deformation amplitude (1.06 ± 0.07 vs. 1.17 ± 0.08 mm, p < 0.05).
CONCLUSIONS: CorVis ST demonstrated reliable measurements for CCT, IOP, and 1st applanation time, as well as relatively reliable measurement for deformation amplitude in both virgin and post-PRK eyes. There were differences in 1st applanation time and deformation amplitude between virgin and post-PRK eyes, which may reflect corneal biomechanical changes occurring after the surgery in the latter.

Entities:  

Mesh:

Year:  2014        PMID: 25302580      PMCID: PMC4193795          DOI: 10.1371/journal.pone.0109577

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

The cornea is a viscoelastic structure with quantifiable biomechanical properties [1]. These properties are related to corneal thickness, age, intraocular pressure (IOP), hydration, and various pathologies [2]–[5]. The cornea's biomechanical behaviour is mostly dictated by the stroma, which encompasses 90% of the total corneal thickness and has a greater mechanical stiffness than the other corneal layers [6]. Corneal biomechanical failure is the basis of keratectatic diseases [7] such as keratoconus and pellucid marginal degeneration. The ability to quantify corneal biomechanical failure represents an important step towards better understanding and treatment of keratectatic diseases. In addition, corneal refractive laser ablation in virgin eyes weakens the cornea mechanically due to tissue removal, leading to deterioration in corneal biomechanical strength [8]. Biomechanical changes may also affect the refractive outcome [9]. Moreover, biomechanical weakening after corneal refractive laser treatment may potentially induce iatrogenic keratectasia [10]. Therefore, knowledge of corneal biomechanical properties is important in predicting clinical outcomes [11] and in identifying cases with high risk for postoperative keratectasia after corneal refractive surgery. Most of the earlier studies concerning corneal biomechanical properties were performed in vitro [12]–[14]. The Ocular Response Analyser (ORA, Reichert, Inc., Depew, NY) was the first device available to evaluate in vivo corneal biomechanical response to an air-puff [1]. It employs a quantitative electro-optical system to monitor the pressures at which the cornea flattens inward and outward by registering the corneal reflex of infrared light. The recently introduced ultra-high-speed Scheimpflug video-imaging device (CorVis ST; Oculus, Wetzlar, Germany) is the first instrument allowing visualization and measurement of corneal deformation in response to a standardized air-puff pressure. Data evaluating the intraobserver repeatability and interobserver reproducibility of measurements with this relatively new device are scarce [15], [16]. Furthermore, such studies as are available concern only healthy virgin eyes. The main goal of the present study was to test the hypothesis that the CorVis ST performs reliable measurements in both virgin and post-refractive surgery eyes. To our knowledge, this is the first study to evaluate the repeatability and reproducibility of CorVis ST measurements in post-refractive surgery eyes. The secondary purpose was to test the hypothesis that the measurements can reveal differences in biomechanical properties between these two groups.

CorVis ST

The CorVis ST utilizes an ultraviolet free blue (455 nm wavelength) light emitting diode (LED) and an ultra-high-speed (4330 frames per second) Scheimpflug camera to record the corneal deformation response to a high intensity air impulse. The air impulse originates from a metered, symmetrical, and fixed maximal internal pump generating a pressure of 25 kilopascal [16]. When the eye is aligned and the Scheimpflug image is in focus, the air puff gets released automatically and the cornea is imaged during the deformation event. The air pulse (lasting approximately 20 ms) forces the cornea inwards through applanation until it achieves its highest concavity (concavity phase). On its way back, the cornea undergoes a second applanation before achieving its natural shape. A total of approximately 140 images of the cornea's two-dimensional cross-section are collected. By software tracing of the anterior and posterior corneal boundaries in individual image frames, parameters describing the corneal deformation response are automatically generated by the instrument. The CorVis ST software version 1.00r30 rev. 771 was used in the current study. With the Corvis ST the biomechanical response of the cornea is characterized by three phases: 1st applanation, highest concavity, and 2nd applanation. In addition to intraocular pressure (IOP) and central corneal thickness (CCT) values, time (time to reach applanation), length (the length of the flattened central cornea), and velocity (the velocity of the corneal apex movement during applanation) at the moment of both the 1st and 2nd applanation events are recorded. The following characteristics at the point of highest concavity are also presented: the highest concavity time, the deformation amplitude, the distance between bending points of the cornea (peak distance), and the concave radius of curvature. (Figure 1.)
Figure 1

The CorVis ST utilizes the Scheimpflug camera to record the dynamic procedure of the corneal response to an air puff.

A) The 1st applanation is achieved. B) The cornea reaches its highest concavity. C) The 2nd applanation is achieved when the cornea rebounds to its original position from the highest concavity.

The CorVis ST utilizes the Scheimpflug camera to record the dynamic procedure of the corneal response to an air puff.

A) The 1st applanation is achieved. B) The cornea reaches its highest concavity. C) The 2nd applanation is achieved when the cornea rebounds to its original position from the highest concavity.

Patients and Methods

Forty candidates for laser refractive surgery (virgin-eye group: 28 males and 12 females) and 42 subjects treated for myopia and astigmatism with photorefractive keratectomy (PRK) earlier (post-PRK group: 23 males and 19 females) were recruited. The PRK treatments were performed using topography-guided transepithelial surface ablation with the iRES system (iRES, iVIS Technology, Taranto, Italy) at SynsLaser Clinic in Tromsø, Norway, 12.69±10.08 months (range: 2 to 48) prior to the current examination. All participants received an extensive ophthalmic examination including Placido-based topography (Nidek OPD Scan II, Nidek Co. Ltd., Aichi, Japan), Scheimpflug topo/tomography (Precisio, iVIS Technology, Taranto, Italy), slit-lamp biomicroscopy and tonometry (Icare tonometer, Revenio Group Corporation, Helsinki, Finland) to exclude corneal and other ocular pathologies. The Regional Committee for Medical and Health Research Ethics in Norway approved the study entitled "2013/762 - Biomechanical cornea measurements by means of CorVis ST". The research complied with the tenets of the Declaration of Helsinki and written informed consent was obtained from each participant before examination. Only the data from the right eye of each participant was used for the present study. The CorVis ST measurements were performed three times by technician A and one time by technician B. The measurement sequence between the technicians was randomized using a randomization table. A one-minute pause was given between each measurement. Repeatability was evaluated by comparing the three consecutive measurements performed by technician A. Reproducibility was determined by comparing the first measurement by technician A with the one performed by technician B. Mean CorVis ST measured values obtained from the three measurements by technician A were used to compare the differences between the virgin and post-PRK eyes groups, as well as for the correlation analysis.

Statistical Analysis

MedCalc software 11.4.2 (MedCalc Software, Ostend, Belgium) and SPSS for Mac software (version 19. SPSS, Inc) were used for statistical analysis. A p-value of less than 0.05 was considered statistically significant. Descriptive statistical results were expressed as mean ± standard deviation (SD). The within-subject standard deviation (Sw), within-subject coefficient of variation (COV), and intraclass correlation coefficient (ICC) were determined to assess the intraobserver repeatability. Interobserver Sw, COV, and ICC were calculated to assess interobserver reproducibility. Independent sample t-test was used to compare the CorVis measured parameters in virgin and post-PRK eyes groups. For the parameters that showed significant differences, univariate analysis of covariance (ANCOVA) was then applied to adjust for selected covariates (age, CCT measured by the CorVis ST, and mean simulated keratometry (simK) value measured by OPD Scan II) to control for potentially confounding factors. Pearson or Spearman correlations were applied to examine the relationship between CCT, manifest refraction spherical equivalent (MRSE) and the deformation parameters.

Results

Patient Demographics

The mean age of the participants at the time of the examination was 27.6±9.0 (range, 18 to 48) and 31.8±6.7 years (range, 20 to 48) for the virgin-eye and post-PRK groups, respectively. In the virgin-eye group, the values of central corneal thickness (CCT) measured by Precisio, IOP measured by Icare rebound tonometer, and mean simK measured by OPD Scan II were not significantly different from the preoperative values of the post-PRK group (Table 1). The mean manifest refraction spherical equivalent (MRSE) in the virgin-eye and its preoperative value in the post-PRK groups were −2.15±2.28 D and −3.52±1.93 D, respectively. In the post-PRK group, the mean maximum ablation depth was 66.71±27.84 µm (range: 18 to 129).
Table 1

Demographic Data of Participants.

Virgin eyes (n = 40)Post-PRK eyes (n = 42) p
Age, years 27.9±9.0 (18, 48)31.8±6.9 (20, 48)0.03
CCT (Precisio), µm 547.82±26.78preop 542.02±30.68 postop 485.00±40.100.30* 0.000*
IOP (Icare), mmHg 15.20±2.57preop 15.81±3.29 postop 12.71±2.770.46* 0.000*
MRSE, D −2.15±2.28preop −3.52±1.93 postop 0.01±0.480.03* 0.000*
Mean simK (OPD Scan II), D 43.47±1.38preop 43.81±1.58 postop 40.87±1.630.18* 0.000*

CCT =  central corneal thickness; IOP =  intraocular pressure; MRSE =  manifest refraction spherical equivalent; simK =  simulated karatometry.

* p values were adjusted for age-difference.

CCT =  central corneal thickness; IOP =  intraocular pressure; MRSE =  manifest refraction spherical equivalent; simK =  simulated karatometry. * p values were adjusted for age-difference.

Intraobserver Repeatability and Interobserver Reproducibility

Tables 2 and 3 present the intraobserver repeatability of the CorVis ST measurements. In the virgin-eye group, the IOP, CCT, 1st applanation time, and 2nd applanation time demonstrated good repeatability (ICC≧0.92), followed by deformation amplitude (ICC: 0.88), Radius of Curvature (ICC: 0.70), 2nd applanation velocity (ICC: 0.65), and highest concavity time (ICC: 0.64). The other parameters showed poor repeatability with large COVs and low ICCs. In the post-PRK group, the IOP, CCT, 1st applanation time, and deformation amplitude demonstrated good repeatability (ICC≧0.90), followed by 2nd applanation time (ICC: 0.89), 2nd applanation velocity (ICC: 0.79), highest concavity time (ICC: 0.66), and radius of curvature (ICC: 0.63). The other parameters showed poor repeatability with large COVs and low ICCs.
Table 2

Intraobserver Repeatability of Parameters Obtained by Corvis in Virgin-Eye Group (n = 40).

ParametersMean ± SDSw2.77SwCOV (%)ICC (95%CI)
IOP (mmHg)14.46±1.330.591.623.590.93 (0.89∼0.96)
CCT (µm)543.32±25.085.3412.560.690.99 (0.98∼0.99)
1st appl. time (ms)7.29±0.210.090.241.090.94 (0.90∼0.97)
1st appl. length (mm)1.83±0.180.290.8113.940.10 (−0.52∼0.49)
1st appl. velocity (m/s)0.14±0.020.030.0818.820.25 (−0.26∼0.57)
2nd appl. time (ms)21.65±0.340.170.480.710.92 (0.87∼0.95)
2nd appl. length (mm)1.89±0.290.451.2421.530.17 (−0.39∼0.53)
2nd appl. velocity (m/s)−0.34±0.040.040.10−9.770.65 (0.42∼0.80)
Highest concavity time (ms)16.40±0.370.371.022.020.64 (0.40∼0.80)
Peak distance (mm)4.36±0.661.173.2321.80−0.04 (−0.74∼0.41)
Radius of curvature (mm)7.49±0.600.551.516.310.70 (0.49∼0.83)
Deformation amplitude (mm)1.06±0.070.040.113.340.88 (0.81∼0.93)

SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness.

Table 3

Intraobserver Repeatability of Parameters Obtained by Corvis in Post-PRK Group (n = 42).

ParametersMean ± SDSw2.77SwCOV (%)ICC (95%CI)
IOP (mmHg)12.38±1.020.551.523.850.90 (0.84∼0.94)
CCT (µm)481.18±42.458.1622.612.290.99 (0.98∼0.99)
1st appl. time (ms)6.96±0.170.090.251.110.91 (0.85∼0.95)
1st appl. length (mm)1.82±0.220.360.9917.650.08 (−0.54∼0.47)
1st appl. velocity (m/s)0.13±0.020.030.0923.30−0.27 (−1.14∼0.28)
2nd appl. time (ms)21.96±0.310.180.490.690.89 (0.81∼0.94)
2nd appl. length (mm)1.70±0.390.491.3726.010.48 (0.12∼0.70)
2nd appl. velocity (m/s)−0.40±0.060.040.12−10.350.79 (0.65∼0.88)
Highest concavity time (ms)16.48±0.350.411.152.120.66 (0.43∼0.81)
Peak distance (mm)4.56±0.771.123.1018.340.30 (−0.18∼0.60)
Radius of curvature (mm)6.43±0.660.701.946.760.63 (0.38∼0.79)
Deformation amplitude (mm)1.17±0.080.040.123.160.92 (0.86∼0.95)

SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness.

SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness. SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness. When comparing the interobserver reproducibility of the CorVis ST parameters, the IOP, CCT, 1st applanation time, and 2nd applanation time demonstrated good reproducibility (ICC≧0.91), followed by deformation amplitude (ICC: 0.88), radius of curvature (ICC: 0.64) and 2nd applanation velocity (ICC: 0.59) in the virgin-eye group. In the post-PRK group, the IOP, CCT, and 1st applanation time demonstrated good reproducibility (ICC≥0.90), followed by deformation amplitude (ICC: 0.88), radius of curvature (ICC: 0.83), 2nd applanation time (ICC: 0.79), highest concavity time (ICC: 0.63), 2nd applanation velocity (ICC: 0.60), and 2nd applanation length (ICC: 0.52), (Table 4 and 5). The other parameters showed poor reproducibility.
Table 4

Interobserver Reproducibility of Parameters Obtained by Corvis in Virgin-Eye Group (n = 40).

ParametersMean Difference ± SDSw2.77SwCOV (%)ICC (95%CI)
IOP (mmHg)−0.01±0.820.581.603.250.92 (0.86∼0.96)
CCT (µm)−1.49±6.764.7813.240.720.98 (0.97∼0.99)
1st appl. time (ms)0.01±0.120.080.230.960.93 (0.88∼0.96)
1st appl. length (mm)−0.02±0.380.270.7311.670.29 (−0.35(0.63)
1st appl. velocity (m/s)(0.005±0.050.030.0919.100.08 ((0.75(0.51)
2nd appl. time (ms)(0.01±0.240.170.470.610.91 (0.82(0.95)
2nd appl. length (mm)(0.10±0.580.441.1417.370.12 ((0.65(0.53)
2nd appl. velocity (m/s)(0.01±0.060.040.11(9.630.59 (0.25(0.78)
Highest concavity time (ms)(0.02±0.540.381.061.810.47 (0.00(0.72)
Peak distance (mm)0.11±1.471.042.8913.080.06 ((1.45(0.63)
Radius of curvature (mm)0.01±0.810.571.585.010.64 (0.31∼0.81)
Deformation amplitude (mm)−0.002±0.060.040.111.950.88 (0.78∼0.94)

SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness.

Table 5

Interobserver Reproducibility of Parameters Obtained by Corvis in Post-PRK Group (n = 42).

ParametersMean Difference ± SDSw2.77SwCOV (%)ICC (95%CI)
IOP (mmHg)−0.17±0.700.501.382.810.90 (0.81∼0.95)
CCT (µm)0.43±5.053.579.890.581.00 (0.99∼1.00)
1st appl. time (ms)−0.02±0.110.080.230.840.90 (0.82∼0.95)
1st appl. length (mm)0.08±0.460.320.9014.450.27 (−0.36∼0.60)
1st appl. velocity (m/s)−0.01±0.040.030.0917.540.45 (−0.03∼0.70)
2nd appl. time (ms)0.04±0.280.200.550.640.79 (0.61∼0.89)
2nd appl. length (mm)−0.11±0.640.451.2520.230.52 (0.12∼0.74)
2nd appl. velocity (m/s)0.01±0.070.050.15−11.030.60 (0.26∼0.78)
Highest concavity time (ms)0.05±0.490.350.971.540.63 (0.31∼0.80)
Peak distance (mm)0.001±1.571.113.0815.230.26 (−0.39∼0.61)
Radius of curvature (mm)−0.06±0.490.350.974.270.83 (0.68∼0.91)
Deformation amplitude (mm)−0.02±0.150.040.123.890.88 (0.78∼0.94)

SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness.

SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness. SD  =  standard deviation, ICC  =  intraclass correlation coefficient, CI  =  confidence interval, Sw  =  within-subject standard deviation, COV  =  within-subject coefficient of variation, IOP  =  intraocular pressure, CCT  =  central corneal thickness. The IOP, CCT, and 1st applanation time demonstrated good intraobserver repeatability and interobserver reproducibility in both groups. The 2nd applanation time had good repeatability and reproducibility in the virgin eyes, with close to good repeatability but not good reproducibility in post-PRK eyes. The deformation amplitude showed a good or close to good repeatability and reproducibility in both groups.

Comparison of the Measurements between Virgin-Eye and Post-PRK Groups

Differences in the CorVis ST measured parameters between the virgin and post-PRK eyes are listed in Table 6. After adjustment for age, CCT, and mean simK, the differences in the mean values of IOP, 1st applanation time, 2nd applanation time, radius of curvature, and deformation amplitude remained significant. Compared to the virgin-eye group, the post-PRK group demonstrated a shorter 1st applanation time, longer 2nd applanation time, smaller radius of curvature, and larger deformation amplitude. The CCT demonstrated a confounding effect in the above-mentioned parameters (p<0.05 in all analyses), while age and simK did not show statistically significant confounding effects (p>0.05 in all analyses).
Table 6

Comparison of The CorVis ST Measurments Between Virgin-Eye and Post-PRK Groups.

Virgin-eye (n = 40)post-PRK (n = 42)Difference (mean ± SE) p Adjusted p* Adjusted R2
IOP (mmHg) 14.46±1.3312.38±1.022.08±0.260.0000.0020.520
CCT (µm) 543.32 ±25.08485.41±39.0057.90±7.280.000
1st applanation
Time (ms) 7.29±0.216.96±0.170.33±0.040.0000.0030.519
Length (mm) 1.83±0.181.81±0.220.01±0.040.832
Velocity (m/s) 0.14±0.020.13±0.020.01±0.000.0100.6140.160
2nd applanation
Time (ms) 21.65±0.3421.96±0.31−0.31±0.070.0000.0320.221
Length (mm) 1.89±0.291.70±0.390.19±0.080.0130.9580.133
Velocity (m/s) −0.34±0.04−0.40±0.060.07±0.010.0000.0530.372
Highest concavity
Time (ms) 16.40±0.3716.48±0.41−0.07±0.090.374
Peak distance 4.36±0.664.56±0.77−0.20±0.160.216
Radius (mm) 7.49±0.606.43±0.661.06±0.140.0000.0010.551
Deformation amplitude (mm) 1.06±0.071.17±0.08−0.10±0.020.0000.0050.402

SE =  standard error, IOP  =  intraocular pressure, CCT  =  central corneal thickness.

p* values were adjusted for the effect of the age, CCT, mean simK difference between the virgin-eye and post-PRK groups.

SE =  standard error, IOP  =  intraocular pressure, CCT  =  central corneal thickness. p* values were adjusted for the effect of the age, CCT, mean simK difference between the virgin-eye and post-PRK groups. Central corneal thickness measured with the CorVis ST correlated to IOP, 1st applanation time, radius of curvature, and deformation amplitude (r = 0.439, 0.437, 0.357, and −0.384, respectively, p<0.05), without significant correlation to other parameters in the virgin-eye group. In the post-PRK group, it correlated to IOP, 1st applanation time, 1st applanation velocity, 2nd applanation length, 2nd applanation velocity, radius of curvature, and deformation amplitude (r = 0.482, 0.483, 0.401, 0.440, 0.395, 0.583, −0.375, respectively, p<0.05). The MRSE correlated to IOP, 1st applanation time, and 2nd applanation time, without significant correlation to other parameters in the virgin-eye group (r = −0.485, −0.492, and 0.420, respectively, p<0.05). The postoperative MRSE was found to correlate only to radius of curvature in the post-PRK group (r = 0.583, p<0.05).

Discussion

In vitro experiments [12], [13] as well as theoretical mathematical models [17], [18] have demonstrated that the cornea exhibits both elastic and viscoelastic properties. When loaded, the cornea shows instantaneous deformation (purely elastic behaviour) followed by a time-dependent deformation response (viscoelastic behaviour) [19]. The ideal device for measuring corneal biomechanical properties in vivo should be accurate, provide repeatable and reproducible results, and be minimally invasive. In the current study, the intraobserver repeatability and interobserver reproducibility of CorVis ST measurements in virgin eyes and post-PRK eyes were investigated. Similar to the studies performed by Nemeth et al. [16] and Hon et al. [15], we found that the following parameters had the best repeatability in both groups: CCT, IOP, 1st applanation time, and deformation amplitude. The current study also presented good repeatability for 2nd applanation time. In addition, the ICCs in the current study were generally higher than in the mentioned studies for most of the parameters measured. The differences between the studies may be attributed to different patient populations and software versions. For example, in the study by Nemeth et al., the mean age was 61.24±15.72 years (95% CI: 57.62 to 64.86 years), while the population in the current study was much younger. In the study by Hon et al., the software did not offer values for radius of curvature and peak distance. When comparing reproducibility, Hon et al. found a statistically significant difference in the CCT measurement between the two sessions. However, the intersession difference was calculated by comparing the examinations performed in the morning (9:00–10:99 am) and afternoon (3:99–5:99 pm) by the same observer. This time difference may have affected the reproducibility evaluation, as corneal thickness demonstrates diurnal variation [20]. The other parameters measured with the CorVis ST did not show satisfactory reliability. The ICCs varied between the virgin and post-PRK eyes. It is conceivable that the cornea would be more difficult to deform and would deform less in eyes with a greater CCT. In line with other studies [15], [21], we revealed a negative correlation between CCT and deformation amplitude in both groups. In addition, the CCT correlated positively with 1st applanation time and radius of curvature in both virgin and post-PRK eyes. However, correlations between CCT and 1st applanation velocity, 2nd applanation time, length, and velocity were only found in post-PRK eyes. This may imply that CCT in normal virgin eyes does not introduce much variation to some of the CorVis ST parameters, while affecting those measurements in biomechanically compromised corneas. The MRSE in our virgin-eyes group demonstrated correlation with some of the parameters measured by CorVis ST. This may need to be taken into consideration if a database of “healthy corneas” is built for the purpose of identifying biomechanically weaker corneas. The IOP measured with the CorVis ST was significantly lower in the post-PRK eye group compared to the virgin-eye group, while the historical preoperative data (IOP measured by Icare, CCT, and corneal curvature) of the post-PRK group showed no significant difference compared to the respective data in the virgin-eye group. The CorVis ST measurements in our post-PRK group were performed a minimum of two months postoperatively, by which time the patients had discontinued the use of local steroids for at least three weeks, to exclude a possible pharmacological effect on their IOP. Some studies have demonstrated that IOP measured with the CorVis ST is more reliable compared to Goldmann applanation tonometry (GAT) and Topcon noncontact tonometry in virgin eyes (Topcon CT-80A Computerized Tonometer; Topcon, Tokyo, Japan) [22]. Still, in the version of CorVis ST used in this study, IOP is calculated based on the timing of the 1st applanation event and is not adjusted for corneal biomechanical properties. Both CCT and corneal biomechanical properties can affect IOP measurements, with the latter suggested to be more influential [18]. The difference in the CorVis ST measured IOP between the groups was most likely caused by changes in corneal biomechanical properties and CCT after PRK. Interestingly, before being adjusted for age, CCT, and simK, the CorVis ST parameters that demonstrated differences between the virgin and post-PRK eyes (1st applanation time, 1st applanation velocity, 2nd applanation time, 2nd applanation velocity, deformation amplitude and radius of curvature) were the same parameters as those showing differences between normal eyes and keratoconus eyes in the study conducted by Ali et al. [23]. It seems that these parameters may be of value in evaluating corneal biomechanical properties. The earlier start of the apex indentation (shorter 1st applanation time) and greater deformation amplitude in post-PRK eyes indicates a lower resistance to deformation due to a decrease in corneal stiffness [24], [25]. Shen et al. [26] compared corneal deformation parameters after femtosecond laser small incision lenticule extraction (SMILE), laser-assisted sub-epithelial keratomileusis (LASEK), and femtosecond laser-assisted LASIK (FS-LASIK). They found greater deformation amplitude and shorter 1st applanation time in the FS-LASIK group compared to the LASEK group. However, those parameters did not differ significantly between the SMILE and LASEK groups, or between SMILE and FS-LASIK groups. This indicates that corneal refractive surgery alters the stiffness of the cornea to different degrees with respect to different surgical approaches. In the current study the CorVis ST measurements in virgin- and post-PRK eyes were taken from two groups of unrelated populations. Pre- and postoperative comparison of the same population would have been better suited to evaluate the changes in biomechanical properties caused by the surgery. We attempted to compensate for this by applying age, CCT, and simK as covariates to adjust for potential confounding factors. For the sake of this discussion we also introduced a separate group of 28 eyes of 16 patients who underwent PRK for myopic astigmatism (mean preoperative MRSE: −3.35±1.98 D, mean postoperative time 9.21±5.09 months) with both pre- and postoperative CorVis ST measurements. The pre- and postoperative CorVis ST measurements of CCT and IOP in that group [547.53±28.89 µm vs. 460.32±48.57 µm (p<0.05), and 15.00±1.48 mmHg vs. 13.48±1.24 mmHg (p<0.001), respectively] were similar to the differences found in the virgin and post-PRK eyes in the current study. Comparable similarity was also found for the 1st applanation time [7.37±0.23 vs. 7.14±0.20 ms (p<0.001)], 2nd applanation time (21.39±0.32 vs. 21.57±0.25 ms (p<0.05), radius of curvature [7.76±0.83 vs. 6.55±0.66 mm (p<0.001)] and deformation amplitude [1.03±0.08 vs. 1.10±0.08 mm, (p<0.05)]. Still, a separate study measuring pre- and post-PRK parameters with a larger population is warranted. The current study demonstrated that, in addition to measurements of CCT and IOP, the CorVis ST showed relatively good reliability in measurements of 1st applanation time and deformation amplitude in both virgin- and post-PRK eyes. The differences in 1st applanation time and deformation amplitude between virgin and post-PRK eyes may imply that the CorVis ST's direct view of the corneal deformation may offer information that promises to yield clinically relevant parameters correlated with corneal biomechanical properties.
  26 in total

1.  Repeatability of ocular biomechanical data measurements with a Scheimpflug-based noncontact device on normal corneas.

Authors:  Gabor Nemeth; Ziad Hassan; Adrienne Csutak; Eszter Szalai; Andras Berta; Laszlo Modis
Journal:  J Refract Surg       Date:  2013-08       Impact factor: 3.573

2.  Corneal deformation measurement using Scheimpflug noncontact tonometry.

Authors:  Ying Hon; Andrew K C Lam
Journal:  Optom Vis Sci       Date:  2013-01       Impact factor: 1.973

3.  The elasticity and rigidity of the outer coats of the eye.

Authors:  M Asejczyk-Widlicka; B K Pierscionek
Journal:  Br J Ophthalmol       Date:  2008-10       Impact factor: 4.638

4.  Effects of altered corneal stiffness on native and postoperative LASIK corneal biomechanical behavior: A whole-eye finite element analysis.

Authors:  Abhijit Sinha Roy; William J Dupps
Journal:  J Refract Surg       Date:  2009-10-12       Impact factor: 3.573

5.  An ultra-high-speed Scheimpflug camera for evaluation of corneal deformation response and its impact on IOP measurement.

Authors:  Christopher Kai-Shun Leung; Cong Ye; Robert N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-04-23       Impact factor: 4.799

6.  Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma.

Authors:  Lukas Reznicek; Daniel Muth; Anselm Kampik; Aljoscha S Neubauer; Christoph Hirneiss
Journal:  Br J Ophthalmol       Date:  2013-08-22       Impact factor: 4.638

7.  Effects of aging on corneal biomechanical properties and their impact on 24-hour measurement of intraocular pressure.

Authors:  Teruyo Kida; John H K Liu; Robert N Weinreb
Journal:  Am J Ophthalmol       Date:  2008-07-09       Impact factor: 5.258

Review 8.  Keratoconus: a biomechanical perspective on loss of corneal stiffness.

Authors:  Abhijit Sinha Roy; Rohit Shetty; Mathew Kurian Kummelil
Journal:  Indian J Ophthalmol       Date:  2013-08       Impact factor: 1.848

9.  Biomechanical properties of the cornea in Fuchs' corneal dystrophy.

Authors:  María A del Buey; José A Cristóbal; Francisco J Ascaso; Laura Lavilla; Elena Lanchares
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-02-28       Impact factor: 4.799

10.  Contributing factors to corneal deformation in air puff measurements.

Authors:  Sabine Kling; Susana Marcos
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-07-26       Impact factor: 4.799

View more
  11 in total

1.  Analysis of corneal biomechanical properties 25 years after myopic photorefractive keratectomy.

Authors:  Sibel Özdoğan; Gökhan Gürelik; Kamil Bilgihan
Journal:  Int Ophthalmol       Date:  2022-08-02       Impact factor: 2.029

2.  Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

Authors:  Ayong Yu; Weiqi Zhao; Giacomo Savini; Zixu Huang; Fangjun Bao; Weicong Lu; Qinmei Wang; Jinhai Huang
Journal:  J Ophthalmol       Date:  2015-12-01       Impact factor: 1.909

3.  Open source software for the analysis of corneal deformation parameters on the images from the Corvis tonometer.

Authors:  Robert Koprowski
Journal:  Biomed Eng Online       Date:  2015-04-11       Impact factor: 2.819

4.  Age-Related Changes in Corneal Deformation Dynamics Utilizing Scheimpflug Imaging.

Authors:  Marta E Rogowska; D Robert Iskander
Journal:  PLoS One       Date:  2015-10-13       Impact factor: 3.240

5.  Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST.

Authors:  Lisa Ramm; Robert Herber; Eberhard Spoerl; Frederik Raiskup; Lutz E Pillunat; Naim Terai
Journal:  J Ophthalmol       Date:  2019-10-16       Impact factor: 1.909

6.  Comparison of Two Cap Thickness in Small Incision Lenticule Extraction: 100μm versus 160μm.

Authors:  Miao He; Wei Wang; Hui Ding; Xingwu Zhong
Journal:  PLoS One       Date:  2016-09-21       Impact factor: 3.240

7.  Repeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis ST.

Authors:  Bernardo T Lopes; Cynthia J Roberts; Ahmed Elsheikh; Riccardo Vinciguerra; Paolo Vinciguerra; Sven Reisdorf; Stefanie Berger; Robert Koprowski; Renato Ambrósio
Journal:  J Ophthalmol       Date:  2017-06-06       Impact factor: 1.909

8.  Corneal biomechanical properties in healthy children measured by corneal visualization scheimpflug technology.

Authors:  Miao He; Hui Ding; Hong He; Chi Zhang; Liangping Liu; Xingwu Zhong
Journal:  BMC Ophthalmol       Date:  2017-05-17       Impact factor: 2.209

9.  Repeatability and comparison of new Corvis ST parameters in normal and keratoconus eyes.

Authors:  Kaili Yang; Liyan Xu; Qi Fan; Dongqing Zhao; Shengwei Ren
Journal:  Sci Rep       Date:  2019-10-25       Impact factor: 4.379

10.  Corneal biomechanical responses detected using corvis st in primary open angle glaucoma and normal tension glaucoma.

Authors:  Younhea Jung; Hae-Young L Park; Sieun Oh; Chan Kee Park
Journal:  Medicine (Baltimore)       Date:  2020-02       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.