Behrooz Kouchaki1, Hassan Hashemi2,1, Abbasali Yekta3, Mehdi Khabazkhoob4,1.
Abstract
PURPOSE: To compare four tonometry techniques: Goldmann applanation tonometer (GAT), Dynamic contour tonometer (DCT), Non-contact tonometer (NCT), and Ocular Response Analyzer (ORA) in the measurement of intraocular pressure (IOP) and the impact of some corneal biomechanical factors on their performance.Entities:
Keywords: Dynamic contour tonometer; Goldmann applanation tonometer; Intraocular pressure; Non-contact tonometer; Ocular response analyze; Tonometry
Year: 2016 PMID: 28626817 PMCID: PMC5463014 DOI: 10.1016/j.joco.2016.08.010
Source DB: PubMed Journal: J Curr Ophthalmol ISSN: 2452-2325
Characteristics of studied parameters.
| Mean ± SD | Range | |
|---|---|---|
| SE (diopter) | −0.4 ± 1.6 | −5.5–2.3 |
| IOP-GAT (mmHg) | 16.4 ± 3.5 | 10.0–26.0 |
| IOP-NCT (mmHg) | 16.2 ± 3.9 | 9.0–26.7 |
| IOP-DCT (mmHg) | 18.1 ± 3.4 | 11.7–28.3 |
| ORA-IOPCC (mmHg) | 17.3 ± 3.4 | 11–26.2 |
| ORA-IOPG (mmHg) | 15.6 ± 4.2 | 7.2–26.5 |
| CH (mmHg) | 9.2 ± 1.6 | 5.1–12.6 |
| CRF(mmHg) | 9.5 ± 2.2 | 3.6–14.7 |
| Axial length mm | 23.9 ± 1.3 | 21.2–27.1 |
| CCT apex (μm) | 527.9 ± 36.6 | 429–610 |
| Mean K (diopter) | 43.6 ± 2.2 | 39.2–49.7 |
| Corneal volume (mm3) | 58.3 ± 4.8 | 50.7–70.9 |
| Chamber volume (mm3) | 174.7 ± 40.5 | 97–267 |
| ACD (mm) | 3.1 ± 0.6 | 2.3–4.9 |
SE: Spherical equivalent.
IOP-GAT: IOP measured by Goldmann applanation tonometer.
IOP-NCT: IOP measured by Non-contact tonometer.
IOP-DCT: IOP measured by Dynamic contour tonometer.
ORA-IOPCC: Corneal compensated IOP measured by Ocular Response Analyzer.
ORA-IOPG: Godmann correlated IOP measured by Ocular Response Analyzer.
CH: Corneal hysteresis measured by Ocular Response Analyzer.
CRF: Corneal resistance factor measured by Ocular Response Analyzer.
CCT: Central corneal thickness.
ACD: Anterior chamber depth.
Simple regressions analysis for intraocular pressure (IOP) and the studied variables.
| Variable | GAT | NCT | DCT | IOPCC | IOPG | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Co | Co | Co | Co | Co | ||||||
| Age (year) | 0.02 | 0.550 | 0.01 | 0.688 | 0.02 | 0.552 | 0.04 | 0.097 | 0.04 | 0.249 |
| Gender | 0.67 | 0.520 | 1.39 | 0.230 | 1.08 | 0.293 | −0.68 | 0.510 | 0.81 | 0.518 |
| SE (Diopter) | 0.44 | 0.172 | 0.44 | 0.219 | 0.25 | 0.444 | 0.49 | 0.126 | 0.62 | 0.109 |
| CH | 0.47 | 0.137 | 0.96 | 0.005 | 0.04 | 0.908 | −0.29 | 0.369 | 0.89 | 0.018 |
| CRF | 0.99 | <0.001 | 1.25 | <0.001 | 0.71 | 0.002 | 0.67 | 0.004 | 1.55 | <0.001 |
| Axial length (mm) | −0.49 | 0.232 | −0.37 | 0.424 | −0.79 | 0.052 | −0.60 | 0.144 | −0.78 | 0.116 |
| CCT | 0.04 | 0.006 | 0.06 | <0.001 | 0.04 | 0.009 | 0.03 | 0.030 | 0.07 | <0.001 |
| mean-K | 0.27 | 0.252 | −0.08 | 0.757 | 0.22 | 0.357 | 0.29 | 0.213 | 0.18 | 0.530 |
| Corneal volume | 0.09 | 0.423 | 0.14 | 0.240 | 0.02 | 0.823 | −0.07 | 0.509 | 0.08 | 0.541 |
Co: Coefficient.
SE: Spherical equivalents.
CH: Corneal hysteresis.
CRF: Corneal resistance factor.
CCT: Central Corneal Thicknesses.
GAT: Goldmann applanation tonometer
NCT: Non-contact tonometer
DCT: Dynamic contour tonometer
IOPCC: Corneal compensated IOP
IOPG: Godmann correlated IOP
Fig. 1Correlation between central corneal thickness (CCT) with corneal resistance factor (CRF) (A) and corneal hysteresis (CH) (B).
Fig. 2Correlation between corneal volume (CV) with corneal resistance factor (CRF) (A) and corneal hysteresis (CH) (B).
Fig. 3Agreement between Goldmann applanation tonometer (GAT) and Non-contact tonometer measurements (NCT) of the intraocular pressure (IOP). The middle line indicates the mean difference, and the two dashed side lines show the 95% limits of agreement.
Fig. 4Agreement between Goldmann applanation tonometer (GAT) and Dynamic contour tonometer (DCT) measurements of the intraocular pressure (IOP). The middle line indicates the mean difference, and the two dashed side lines show the 95% limits of agreement.
Fig. 5Agreement between Goldmann applanation tonometer (GAT) and Ocular Response Analyzer (IOPCC) measurements of the intraocular pressure (IOP). The middle line indicates the mean difference, and the two dashed side lines show the 95% limits of agreement.
Fig. 6Agreement between Goldmann applanation tonometer (GAT) and Ocular Response Analyzer (IOPG) measurements of the intraocular pressure (IOP). The middle line indicates the mean difference, and the two dashed side lines show the 95% limits of agreement.
Prediction of the measured intraocular pressure (IOP) by Non-contact tonometer (NCT), Dynamic contour tonometer (DCT), ORA IOPCC and IOPG based on GAT by linear regression analysis.
| IOP by GAT = 1.05 + IOP by NCT × 0.92 |
| IOP by GAT = 4.7 + IOP by DCT × 0.81 |
| IOP by GAT = 4.56 + IOP by IOPCC × 0.77 |
| IOP by GAT = −0.96 + IOP by IOPG × 1.01 |
GAT: Goldmann applanation tonometer
NCT: Non-contact tonometer
DCT: Dynamic contour tonometer
IOPCC: Corneal compensated IOP
IOPG: Godmann correlated IOP