| Literature DB >> 27863469 |
Yang Shen1,2, Xiangjian Su3, Xiu Liu3, Huamao Miao1,2, Xuejun Fang4, Xingtao Zhou5,6.
Abstract
BACKGROUND: Corneal biomechanical properties are always compromised after corneal refractive surgeries thus leading to underestimated intraocular pressure (IOP) that complicates the management of IOP. We investigated the changes in postoperative baseline of IOP values measured with noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology (CST) in the early phase after small incision lenticule extraction (SMILE).Entities:
Keywords: Corneal biomechanical property; Corneal deformation parameter; Corvis scheimpflug technology tonometer; Intraocular pressure; Ocular response analyzer; Small incision lenticule extraction
Mesh:
Year: 2016 PMID: 27863469 PMCID: PMC5116167 DOI: 10.1186/s12886-016-0381-3
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
The main demographic and topographic data (n = 22)
| Variables | Pre-operative | Postoperative 24 h | F valuea |
| ||
|---|---|---|---|---|---|---|
| Mean ± SD | Range | Mean ± SD | Range | |||
| Age (year) | 28.4 ± 7.1 | 18 to 42 | 28.4 ± 7.1 | 18 to 42 | – | – |
| MRSE (D) | −6.76 ± 1.39 | −10.00 to −4.13 | −0.22 ± 0.48 | −1.38 to 0.75 | 508.580 | <0.001b |
| K1 (D) | 43.23 ± 1.40 | 40.1 to 45.2 | 38.35 ± 2.12 | 32.3 to 40.9 | 510.718 | <0.001b |
| K2 (D) | 44.56 ± 1.84 | 41.0 to 47.6 | 39.03 ± 2.23 | 32.7 to 42.3 | 557.309 | <0.001b |
| Km (D) | 43.88 ± 1.54 | 40.6 to 46.4 | 38.69 ± 2.15 | 32.5 to 41.6 | 622.613 | <0.001b |
| Pentacam-CCT (μm) | 546.6 ± 23.4 | 515 to 592 | 431.9 ± 28.0 | 399 to 501 | 521.421 | <0.001b |
MRSE manifest refraction spherical equivalent, D diopter, K1 flat curvature power, K2 flat curvature power, Km mean curvature power, CCTcentral corneal thickness
aRepeated measures analysis of variance (RM-ANOVA)
bSignificant difference was detected
Changes in IOP measurements (n = 22)
| Variables | Pre-operative | Postoperative 20 min | Postoperative 24 h | F valuea |
| |||
|---|---|---|---|---|---|---|---|---|
| Mean ± SD | Range | Mean ± SD | Range | Mean ± SD | Range | |||
| NCT-IOP (mmHg) | 15.04 ± 2.91 | 11.1 to 20.9 | 9.19 ± 2.52 | 4.6 to 15.2 | 9.70 ± 2.95 | 5.8 to 17.2 | 81.173 | <0.001b |
| IOPcc (mmHg) | 14.99 ± 2.46 | 10.2 to 19.0 | 16.65 ± 4.29 | 12.1 to 29.4 | 14.01 ± 3.10 | 6.8 to 20.3 | 4.975 | 0.030b |
| IOPg (mmHg) | 15.10 ± 3.21 | 9.3 to 21.9 | 11.20 ± 3.38 | 4.7 to 18.2 | 9.52 ± 2.85 | 4.5 to 15.9 | 40.128 | <0.001b |
| CST-IOP (mmHg) | 17.74 ± 2.16 | 13.0 to 23.0 | 14.29 ± 2.09 | 10.0 to 17.5 | 14.57 ± 1.58 | 10.5 to 18.0 | 31.886 | <0.001b |
IOPcc corneal-compensated IOP, IOPg Goldmann-correlated IOP
aRepeated measures analysis of variance (RM-ANOVA)
bSignificant difference was detected over time
Fig. 1Changes in intraocular pressure following SMILE obtained by, NCT (NCT-IOP), ORA(ORA-IOPcc and ORA-IOPg) and CST (CST-IOP). Pre-op = postoperative; Post-op = postoperative; 20 M = 20 min; 24H = 24 h. “*” refers to Fisher’s least significant difference (LSD) post hoc P value <0.05; “**” refers to LSD post hoc P value < 0.01
Abbreviations for corneal deformation parameters
| Abbreviation | Definition |
|---|---|
| AT1 | Applination Time (time 1): the duration from the initiation to the moment that a cornea just deformed for the first time |
| ATh | Applination Time (hightest concavity): the duration from the initiation to the moment that a cornea is depressed to the highest concavity |
| AT2 | Applination Time (time 2): the duration from the initiation to the moment that a corneal deformation just completely resumed |
| AL1 | Applination length (time 1): the cord length recorded at the moment that a cornea just deformed for the first time |
| AL2 | Applination length (time 2): the cord length recorded at the moment that a corneal deformation just completely resumed |
| AV1 | Applination velocity (time 1): the instantaneous velocity recorded when a cornea just deformed |
| AV2 | Applination velocity (time 2): the instantaneous velocity recorded when a cornea just completely resumed |
| Radius | Radius: the corneal radius obtained when a cornea is depressed to the highest concavity |
| PD | Peak Distance: the distance between the two corneal peaks recorded when a cornea is depressed to the highest concavity |
| DA | Deformation Amplitude: the maximum amplitude of corneal deformation recorded when a cornea is depressed to the highest concavity |
Changes in corneal biomechanical and deformation parameters (n = 22)
| Variables | Pre-operative | Postoperative 20 min | Postoperative 24 h | F valuea |
| |||
|---|---|---|---|---|---|---|---|---|
| Mean ± SD | Range | Mean ± SD | Range | Mean ± SD | Range | |||
| CH (mmHg) | 11.05 ± 1.81 | 8.7 to 16.7 | 6.82 ± 0.98 | 5.3 to 8.4 | 7.49 ± 1.32 | 5.6 to 11.0 | 90.283 | <0.001b |
| CRF (mmHg) | 10.84 ± 2.20 | 7.1 to 17.3 | 6.07 ± 1.26 | 2.9 to 9.3 | 6.09 ± 1.27 | 4.1 to 9.1 | 131.239 | <0.001b |
| CST-CCT (μm) | 545.8 ± 24.8 | 515.0 to 618.0 | 469.7 ± 46.4 | 389 to 580 | 440.9 ± 29.5 | 407 to 522 | 136.093 | <0.001b |
| AT1 (ms) | 7.28 ± 0.26 | 6.71 to 7.97 | 6.91 ± 0.23 | 6.52 to 7.32 | 6.90 ± 0.18 | 6.54 to 7.37 | 36.484 | <0.001b |
| ATh (ms) | 16.94 ± 0.53 | 16.17 to 17.79 | 16.77 ± 0.86 | 14.32 to 18.02 | 16.94 ± 0.63 | 15.71 to 18.02 | 0.465 | 0.631 |
| AT2 (ms) | 21.64 ± 1.29 | 16.89 to 22.82 | 22.13 ± 0.41 | 21.36 to 22.86 | 22.17 ± 0.29 | 21.61 to 22.87 | 3.042 | 0.089 |
| AL1 (mm) | 1.78 ± 0.04 | 1.71 to 1.87 | 1.72 ± 0.17 | 1.26 to 1.92 | 1.74 ± 0.20 | 1.19 to 1.94 | 0.798 | 0.429 |
| AL2 (mm) | 1.49 ± 0.39 | 0.44 to 2.00 | 1.35 ± 0.44 | 0.73 to 2.00 | 1.22 ± 0.41 | 0.73 to 1.89 | 2.281 | 0.131 |
| AV1 (m/s) | 0.16 ± 0.02 | 0.09 to 0.22 | 0.14 ± 0.02 | 0.08 to 0.19 | 0.15 ± 0.02 | 0.11 to 0.20 | 5.528 | 0.007b |
| AV2 (m/s) | −0.44 ± 0.14 | −0.92 to −0.20 | −0.56 ± 0.13 | −0.75 to −0.33 | −0.61 ± 0.16 | −1.11 to −0.33 | 8.910 | 0.001b |
| PD (mm) | 4.27 ± 1.19 | 2.36 to 5.43 | 4.51 ± 1.25 | 2.62 to 6.19 | 4.53 ± 1.27 | 2.57 to 6.23 | 0.355 | 0.704 |
| Radius (mm) | 6.97 ± 1.46 | 1.66 to 9.45 | 5.51 ± 1.00 | 1.78 to 6.88 | 5.69 ± 0.54 | 4.89 to 7.05 | 13.208 | <0.001b |
| DA (mm) | 1.09 ± 0.09 | 0.89 to 1.37 | 1.17 ± 0.10 | 1.03 to 1.43 | 1.14 ± 0.07 | 0.99 to 1.34 | 10.100 | <0.001b |
MRSE manifest refraction spherical equivalent, D diopter, CH corneal hysteresis, CRF corneal resistance factor, CCT central corneal thickness, AT1 applanation time 1, ATh applanation time at the highest concavity; AT2 applanation time 2, AL1 applanation length 1; AL2 applanation length 2, AV1 applanation velocity 1, AV2 applanation velocity 2, PD peak distance, DA deformation amplitude
aRepeated measures analysis of variance (RM-ANOVA)
bSignificant difference was detected over time
The stepwise multiple linear regression models for predicting theoretical post-op IOP values (n = 22)
| Dependent Variables | Main Predictors | Ba | SEb | tc | Sig. | βd | Regression Equation | Adjusted R2 | Fe | Sig. |
|---|---|---|---|---|---|---|---|---|---|---|
| Pre-op NCT-IOP | Practical postop | 0.646 | 0.167 | 3.873 | <0.001 | 0.655 | Theoretical post-op NCT-IOP (mmHg) = 0.646 × Practical post-op NCT-IOP(mmHg) + 8.774(mmHg) | 0.400 | 15.000 | <0.001 |
| Constant | 8.774 | 1.688 | 5.198 | 0.001 | - | |||||
| Pre-op ORAIOPcc | Practical postop | 0.558 | 0.126 | 4.445 | <0.001 | 0.705 | Theoretical post-op IOPcc (mmHg) = 0.558 × practical postoperative IOPcc (mmHg) + 7.166(mmHg) | 0.472 | 19.755 | <0.001 |
| Constant | 7.166 | 1.800 | 3.981 | 0.001 | - | |||||
| Pre-op ORAIOPg | Practical postopIOPg | 0.848 | 0.167 | 5.084 | <0.001 | 0.751 | Theoretical post-op IOPg (mmHg) = 0.848 × practical postoperative IOPg (mmHg) + 7.024 (mmHg) | 0.542 | 25.850 | <0.001 |
| Constant | 7.024 | 1.654 | 4.247 | <0.001 | - | |||||
| Pre-op CST-IOP | Post-op AT1 | 8.079 | 1.660 | 4.866 | <0.001 | 0.680 | Theoretical post-op CST-IOP | 0.596 | 11.336 | <0.001 |
| Pre-op CCT | 0.035 | 0.013 | 2.732 | 0.014 | 0.379 | |||||
| Post-op PD | 0.515 | 0.237 | 2.176 | 0.043 | 0.304 | |||||
| Constant | −59.47 | 13.524 | −4.397 | <0.001 | - |
Pre-op pre-operative, Post-op post-operative, AT1 applanation time 1, CCT central corneal thickness, PDpeak distance, Sig. Significance
aUnstandardized Coefficients
bStandard Error of Unstandardized Coefficients
cUnstandardized Coefficients/Standard Error
dStandardized Coefficients (Beta)
eMultiple Linear Regression Model (Stepwise)