This study compared the axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), mean anterior corneal radius of curvature (Rm), and postoperative refractive outcomes obtained from two different swept-source optical coherence biometers, the ARGOS (Movu, Nagoya, Japan), which uses the segmental refractive index for each segment, and the IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), which uses an equivalent refractive index for the entire eye. One hundred and six eyes of 106 patients with cataracts were included. The refractive outcomes using the Barrett Universal II, Haigis, Hoffer Q, and SRK/T formulas were evaluated. The mean AL, CCT, ACD, and Rm differed significantly (P < 0.001) with the IOLMaster 700 (25.22 mm, 559 µm, 3.23 mm, and 7.69 mm) compared with the ARGOS (25.14 mm, 533 µm, 3.33 mm, and 7.66 mm). The mean LTs did not differ significantly. The percentages of eyes within ±0.50 and ±1.00 diopter of the predicted refraction did not differ significantly (P > 0.05). The accuracy of the intraocular lens power calculations was clinically acceptable with both biometers, although the ocular biometry using these two biometers exhibited certain differences.Entities:
Mesh:
Year: 2019 PMID: 31024017 PMCID: PMC6483997 DOI: 10.1038/s41598-019-42968-3
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Comparison of biometric measurements using two optical biometers AL.
| Parameter | Optical Biometer | P value | |||
|---|---|---|---|---|---|
| ARGOS | IOLMaster700 | ||||
| Mean ± SD | Range | Mean ± SD | Range | ||
|
| |||||
| AL (mm) | 25.14 ± 1.90 | 22.24, 30.64 | 25.22 ± 1.95 | 22.15, 30.89 | <0.001 |
| CCT (μm) | 533 ± 32 | 458, 624 | 559 ± 32 | 476, 653 | <0.001 |
| ACD (mm) | 3.33 ± 0.42 | 2.18, 4.36 | 3.23 ± 0.42 | 2.06, 4.22 | <0.001 |
| LT (mm) | 4.47 ± 0.44 | 3.47, 5.66 | 4.46 ± 0.43 | 3.48, 5.64 | 0.515 |
| Rm (mm) | 7.66 ± 0.28 | 6.93, 8.14 | 7.69 ± 0.28 | 6.94, 8.17 | <0.001 |
|
| |||||
| AL (mm) | 24.16 ± 1.08 | 22.24, 25.86 | 24.20 ± 1.11 | 22.15, 25.97 | <0.001 |
| CCT (μm) | 532 ± 33 | 458, 624 | 557 ± 35 | 476, 653 | <0.001 |
| ACD (mm) | 3.20 ± 0.38 | 2.18, 4.03 | 3.10 ± 0.38 | 2.06, 3.93 | <0.05 |
| LT (mm) | 4.53 ± 0.43 | 3.47, 5.66 | 4.53 ± 0.42 | 3.48, 5.64 | 0.882 |
| Rm (mm) | 7.63 ± 0.27 | 6.93, 8.12 | 7.66 ± 0.26 | 6.94, 8.12 | <0.05 |
|
| |||||
| AL (mm) | 27.64 ± 1.01 | 25.90, 30.64 | 27.79 ± 1.04 | 26.01, 30.89 | <0.001 |
| CCT (μm) | 537 ± 29 | 470, 597 | 562 ± 26 | 522, 617 | <0.001 |
| ACD (mm) | 3.65 ± 0.33 | 3.11, 4.36 | 3.56 ± 0.32 | 3.00, 422 | <0.001 |
| LT (mm) | 4.30 ± 0.42 | 3.58, 5.24 | 4.29 ± 0.41 | 3.53, 5.24 | 0.246 |
| Rm (mm) | 7.75 ± 0.29 | 7.13, 8.14 | 7.77 ± 0.30 | 7.15, 8.17 | <0.001 |
SD = standard deviation.
Figure 1Bland-Altman plots show the agreements between the two biometers in the AL, CCT, ACD, LT, and Rm (n = 106). The solid lines indicate the mean difference. The dashed lines indicate the 95% LoAs.
Comparison of prediction errors for each formula using two devices.
| Arithmetic prediction error within (%) | Formula | Optical Biometer | P value | |
|---|---|---|---|---|
| ARGOS | IOLMaster 700 | |||
| ±0.50 D | Barrett Universal II | 66.0 | 62.3 | 0.481 |
| Haigis | 68.9 | 68.9 | 1 | |
| Hoffer Q | 58.5 | 56.6 | 0.815 | |
| SRK/T | 63.2 | 62.3 | 1 | |
| ±1.00 D | Barrett Universal II | 92.5 | 91.5 | 1 |
| Haigis | 93.4 | 93.4 | 1 | |
| Hoffer Q | 92.5 | 88.7 | 0.289 | |
| SRK/T | 92.5 | 91.5 | 1 | |
| ±0.50 D | Barrett Universal II | 64.5 | 61.8 | 0.791 |
| Haigis | 67.1 | 71.1 | 0.607 | |
| Hoffer Q | 67.1 | 67.1 | 1 | |
| SRK/T | 61.8 | 68.4 | 0.277 | |
| ±1.00 D | Barrett Universal II | 92.1 | 90.8 | 1 |
| Haigis | 93.4 | 93.4 | 1 | |
| Hoffer Q | 93.4 | 93.4 | 1 | |
| SRK/T | 92.1 | 89.5 | 0.625 | |
| ±0.50 D | Barrett Universal II | 70.0 | 63.3 | 0.625 |
| Haigis | 73.3 | 63.3 | 0.250 | |
| Hoffer Q | 36.7 | 30.0 | 0.625 | |
| SRK/T | 66.7 | 46.7 | 0.070 | |
| ±1.00 D | Barrett Universal II | 93.3 | 93.3 | 1 |
| Haigis | 93.3 | 93.3 | 1 | |
| Hoffer Q | 90.0 | 77.7 | 0.125 | |
| SRK/T | 93.3 | 96.7 | 1 | |
Figure 2Distribution of the arithmetic prediction error in refraction with the four IOL power calculation formulas and the two optical biometers for the entire AL range, medium and long AL subgroups. *P < 0.05.
Figure 3Distribution of the absolute prediction error in refraction with the four IOL power calculation formulas and the two optical biometers for the entire AL range, medium and long AL subgroups. *P < 0.05.