Amita Panda1, Ashok Nanda2, Kalyani Sahoo3.
Abstract
Purpose: To assess the comparability of ocular biometry measurements and refractive outcomes between ANTERION and IOL Master 700.Entities:
Keywords: ANTERION; IOLMaster 700; biometry
Mesh:
Year: 2022 PMID: 35502033 PMCID: PMC9332939 DOI: 10.4103/ijo.IJO_2433_21
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 2.969
Age and gender distribution of cases
| No. | % | |
|---|---|---|
| Age in years | ||
| ≤50 | 20 | 9.9 |
| 51-60 | 63 | 31 |
| 61-70 | 92 | 45.3 |
| ≥71 | 28 | 13.8 |
| Gender | ||
| Male | 114 | 56.2 |
| Female | 89 | 43.8 |
| Total | 203 | 100 |
Pair-wise comparison for agreement between biometric values of ANTERION and IOLMaster 700 (n=203)
| Mean±SD | Mean diff.±SD | ICC | SE |
| ||
|---|---|---|---|---|---|---|
|
| ||||||
| ANTERION | IOLMaster 700 | |||||
| Steep K | 44.3±1.6 | 44.5±1.7 | -0.18±0.26 | 0.994 | 0.02 | 0.000 |
| Axis | 89.2±55.1 | 91.2±63.9 | -2.03±49.29 | 0.794 | 3.46 | 0.558 |
| Flat K | 43.5±1.6 | 43.6±1.7 | -0.13±0.28 | 0.992 | 0.02 | 0.000 |
| Sim K | 43.9±1.6 | 44.0±1.6 | -0.15±0.23 | 0.995 | 0.02 | 0.000 |
| CCT | 525.0±31.6 | 526.5±33.0 | -1.48±5.73 | 0.992 | 0.40 | 0.000 |
| AQD | 2.8±0.4 | 2.7±0.4 | 0.05±0.12 | 0.970 | 0.01 | 0.000 |
| LT | 4.4±0.4 | 4.4±0.4 | 0.08±0.12 | 0.977 | 0.01 | 0.000 |
| AL | 23.4±1.2 | 23.4±1.1 | -0.02±0.09 | 0.999 | 0.01 | 0.001 |
| WTW | 11.8±0.4 | 12.0±0.4 | -0.19±0.21 | 0.925 | 0.01 | 0.000 |
Difference of measured parameters in ANTERION and IOLMaster 700 beyond pre-defined clinically acceptable range
| Difference | Classification | No. | % |
|---|---|---|---|
| ANTERION Steep K - IOLMaster 700 Steep K | >0.25 D | 88 | 43.3 |
| ≤0.25 D | 115 | 56.7 | |
| ANTERION Axis - IOLMaster 700 Axis | >10° | 95 | 46.8 |
| ≤10° | 108 | 53.2 | |
| ANTERION Flat K - IOLMaster 700 FLAT | >0.25 D | 82 | 40.4 |
| ≤0.25 D | 121 | 59.6 | |
| ANTERION SIM K - IOLMaster 700 SIM K | >0.25 D | 65 | 32 |
| ≤0.25 D | 138 | 68 | |
| ANTERION CCT - IOLMaster 700 CCT | >10 microns | 12 | 5.9 |
| ≤10 microns | 191 | 94.1 | |
| ANTERION AQD - IOLMaster 700 AQD | >0.1 mm | 21 | 10.3 |
| ≤0.1 mm | 182 | 89.7 | |
| ANTERION LT - IOLMaster 700 LT | >0.1 mm | 48 | 23.6 |
| ≤0.1 mm | 155 | 76.4 | |
| ANTERION AL - IOL Master 700 AL | >0.1 mm | 9 | 4.4 |
| ≤0.1 mm | 194 | 95.6 | |
| ANTERION WTW - IOLMaster 700 WTW | >0.1 mm | 156 | 76.8 |
Difference in actual postoperative refraction from the predicted residual refraction in both machines
| Difference of actual postoperative refraction from predicted residual refraction in ANTERION | Difference of actual postoperative refraction from predicted residual refraction in IOLMaster 700 | Total | Kappa | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
| |||||||||||
| -1.9D to -0.5D | -0.5 D to 0 | 0 to +0.5D | >+0.5 to +1.9D | ||||||||
|
|
|
|
|
| |||||||
| No. | % | No. | % | No. | % | No. | % | No. | % | ||
| -1.9D to -0.5D | 22 | 10.8 | 1 | 0.49 | 0 | 0 | 0 | 0 | 23 | 11.3 | Kappa=0.437 |
| -0.5D to 0 | 13 | 6.4 | 34 | 16.7 | 5 | 2.5 | 0 | 0 | 52 | 25.6 | |
| 0 to +0.5D | 0 | 0 | 20 | 10.1 | 26 | 12.8 | 1 | 0.49 | 47 | 23.2 | |
| >+0.5D to +1.9D | 0 | 0 | 3 | 1.5 | 26 | 12.8 | 52 | 25.6 | 81 | 39.9 | |
| Total | 35 | 17.2 | 58 | 28.6 | 57 | 28.1 | 53 | 26.1 | 203 | 100 | |