Validity and repeatability of the Aladdin ocular biometer.

Priyanka Mandal1, Emma J Berrow, Shehzad A Naroo, James S Wolffsohn, Detlef Uthoff, Detlef Holland, Sunil Shah.   

Abstract

AIM: To assess the accuracy and reproducibility of biometry undertaken with the Aladdin (Topcon, Tokyo, Japan) in comparison with the current gold standard device, the IOLMaster 500 (Zeiss, Jena, Germany).
SETTING: University Eye Clinic, Birmingham, UK and Refractive Surgery Centre, Kiel, Germany.
METHODS: The right eye of 75 patients with cataracts and 22 healthy participants were assessed using the two devices. Measurements of axial length (AL), anterior chamber depth (ACD) and keratometry (K) were undertaken with the Aladdin and IOLMaster 500 in random order by an experienced practitioner. A second practitioner then obtained measurements for each participant using the Aladdin biometer in order to assess interobserver variability.
RESULTS: No statistically significant differences (p>0.05) between the two biometers were found for average difference (AL)±95% CI=0.01±0.06 mm), ACD (0.00±0.11 mm) or mean K values (0.08±0.51 D). Furthermore, interobserver variability was very good for each parameter (weighted κ≥0.85). One patient's IOL powers could not be calculated with either biometer measurements, whereas a further three could not be analysed by the IOLMaster 500. The IOL power calculated from the valid measurements was not statistically significantly different between the biometers (p=0.842), with 91% of predictions within±0.25 D.
CONCLUSIONS: The Aladdin is a quick, easy-to-use biometer that produces valid and reproducible results that are comparable with those obtained with the IOLMaster 500.

Entities:  

Keywords:  Eye (Globe); Imaging

Mesh:

Year:  2013        PMID: 24227803     DOI: 10.1136/bjophthalmol-2013-304002

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  14 in total

1.  Evaluation of biometry and corneal astigmatism in cataract surgery patients in Northern United Arab Emirates.

Authors:  Eui Seok Han; Moonjung Kim
Journal:  Int Ophthalmol       Date:  2019-05-27       Impact factor: 2.031

2.  Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.

Authors:  Shi-Ming Cheng; Jia-Sheng Zhang; Xu Shao; Ze-Tong Wu; Tian-Tian Li; Po Wang; Jun-Hai Lin; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-16       Impact factor: 3.117

3.  The long-term observation in Chinese children with monocular myelinated retinal nerve fibers, myopia and amblyopia.

Authors:  Yang Shen; Jing Zhao; Ling Sun; Li Zeng; Zhi Chen; Mi Tian; Xingtao Zhou
Journal:  Transl Pediatr       Date:  2021-04

4.  Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents.

Authors:  Mehmet Kola; Hikmet Duran; Adem Turk; Suleyman Mollamehmetoglu; Ahmet Kalkisim; Hidayet Erdol
Journal:  J Ophthalmol       Date:  2014-07-22       Impact factor: 1.909

5.  Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography.

Authors:  Onur Polat; Zeki Baysal; Serkan Özcan; Sibel İnan; Ümit Übeyt İnan
Journal:  Turk J Ophthalmol       Date:  2016-12-01

6.  Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculation.

Authors:  Giacomo Savini; Kenneth J Hoffer; Piero Barboni; Nicole Balducci; Domenico Schiano-Lomoriello; Pietro Ducoli
Journal:  PLoS One       Date:  2017-02-23       Impact factor: 3.240

7.  Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices.

Authors:  Miki Kamikawatoko Omoto; Hidemasa Torii; Sachiko Masui; Masahiko Ayaki; Kazuo Tsubota; Kazuno Negishi
Journal:  Sci Rep       Date:  2019-04-25       Impact factor: 4.379

8.  OCT Biometry (B-OCT): A New Method for Measuring Ocular Axial Dimensions.

Authors:  Bartosz L Sikorski; Pawel Suchon
Journal:  J Ophthalmol       Date:  2019-08-14       Impact factor: 1.909

9.  Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes.

Authors:  Yanjun Hua; Wei Qiu; Qiuyi Xiao; Qiang Wu
Journal:  PLoS One       Date:  2018-02-27       Impact factor: 3.240

10.  Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device.

Authors:  Adam Wylęgała; Bartłomiej Bolek; Robert Mazur; Edward Wylęgała
Journal:  Sci Rep       Date:  2020-09-02       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.