| Literature DB >> 28464806 |
Jinho Jeong1, Han Song2, Jimmy K Lee3, Roy S Chuck3, Ji-Won Kwon4,5.
Abstract
BACKGROUND: To evaluate how differences in ocular biometry affects the Hoffer Q, Holladay 1, SRK/T, and Haigis intraocular lens power calculation formulae predictions.Entities:
Keywords: Anterior chamber depth; IOL formula; Ocular biometry
Mesh:
Year: 2017 PMID: 28464806 PMCID: PMC5414130 DOI: 10.1186/s12886-017-0454-y
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Characteristics of preoperative biometric data and the power of IOL distribution
| Minimum | Maximum | Mean | |
|---|---|---|---|
| Age (year) | 47 | 81 | 70.71 ± 8.99 |
| Axial length (mm) | 21.61 | 27.91 | 23.73 ± 1.55 |
| Cylinder of refractive error (diopter) | 0 | 3.00 | 1.17 ± 0.73 |
| Corneal curvature (diopter) | 40.75 | 47.25 | 43.75 ± 1.60 |
| Pre-op Anterior chamber depth (mm) | 2.15 | 4.05 | 3.13 ± 0.49 |
| Post-op Anterior chamber depth (mm) | 2.75 | 5.50 | 4.49 ± 0.50 |
Fig. 1The scatterplot of preoperative biometric values with expected difference (ED) of the 3rd generation formulas from the Haigis formula
The correlation analysis of biometric values for the expected difference (ED) of each 3rd generation formulas from the Haigis formula
| Pearson Correlation | ED of Hoffer Q | ED of Holladay | ED of SRK/T |
|---|---|---|---|
| Corneal curvature (D) | 0.206 ( | 0.665 ( | 0.827 ( |
| Axial length (mm) | −0.681 ( | −0.653 ( | −0.382 ( |
| Anterior chamber depth (mm) | −0.756 ( | −0.548 ( | −0.252 ( |
| Delta ACD (mm)a | 0.424 ( | 0.201 ( | 0.117 ( |
aDelta ACD means post-operative 1 month change of anterior chamber depth (ACD)
The results of multiple regression analyses of preoperative biometric values for the expected difference (ED) of each 3rd generation formulas from the Haigis formula
| ED of Hoffer Q | ED of Holladay | ED of SRK/T | |
|---|---|---|---|
| Adjusted R2 | 0.659 | 0.715 | 0.747 |
| Constant sig.a | 0.000 | 0.000 | 0.000 |
| Corneal curvature (D) | 0.809 | 0.000 | 0.000 |
| Axial length (mm) | 0.000 | 0.072 | 0.005 |
| Anterior chamber depth (mm) | 0.000 | 0.000 | 0.000 |
a Constant sig. = constant significance, p value of the multiple regression analysis
Fig. 2The scatterplot of preoperative biometric values with post-op mean errors of the formulas
The correlation analysis of preoperative biometric values for the difference of each formula expected value from the post-operative refraction
| Hoffer Q error | Holladay error | SRK/T error | Haigis error | |
|---|---|---|---|---|
| Corneal curvature (D) | 0.127 ( | −0.019 ( | −0.141 ( | 0.187 ( |
| Axial length (mm) | 0.277 ( | 0.301 ( | 0.259 ( | 0.109 ( |
| Pre-op ACD (mm) | 0.265 ( | 0.238 ( | 0.176 ( | 0.077 ( |
| Post-op change of ACD (mm) | 0.266 ( | 0.219 (=0.037)a | 0.209 ( | 0.164 ( |
aStatistically significant
The correlation analysis of preoperative biometric values for mean absolute errors (MAE) of the formulas
| Hoffer Q MAE | Holladay MAE | SRK/T MAE | Haigis MAE | |
|---|---|---|---|---|
| Corneal curvature (D) | 0.019 ( | 0.067 ( | 0.155 ( | 0.043 ( |
| Axial length (mm) | −0.016 ( | 0.003 ( | −0.075 ( | −0.121 ( |
| Pre-op ACD (mm) | 0.032 ( | 0.061 ( | 0.054 ( | 0.002 ( |
| Post-op change of ACD (mm) | 0.198 ( | 0.191 (=0.069) | 0.206 ( | 0.180 ( |