Comparison of Axial Length, Corneal Curvature, and Anterior Chamber Depth Measurements of 2 Recently Introduced Devices to a Known Biometer.

Mehdi Shajari1, Carla Cremonese1, Kerstin Petermann1, Pankaj Singh1, Michael Müller1, Thomas Kohnen2.   

Abstract

PURPOSE: Evaluate agreement and repeatability of axial length (AL), corneal curvature, and anterior chamber depth measurements of 2 new devices, 1 using Scheimpflug technology combined with partial coherence interferometry (Pentacam AXL) and 1 using swept-source optical coherence tomography (IOL Master 700) to a widely used optical biometer (IOL Master 500).
DESIGN: Prospective, instrument reliability analysis.
METHODS: This study was conducted in the Department of Ophthalmology at Goethe-University, Frankfurt, Germany. Seventy-nine eyes in 79 patients were measured with the IOL Master 500, the IOL Master 700, and with the Pentacam AXL. All eyes were measured twice with each device. The results were statistically compared using Bland-Altman plots. Outcome measures included AL, corneal curvature, and the measurement of anterior chamber depth.
RESULTS: The Bland-Altman analysis of axial length measurements with the IOL Master 500 and IOL Master 700 showed a mean difference of 0.008 mm between the 2 devices, with a coefficient of repeatability (COR) of 0.044 (P = .05). For measurements with the IOL Master 500 and Pentacam AXL, a mean difference of 0.026 mm and COR of 0.055 (P = .75) was found. For measurements with Pentacam AXL and IOL Master 700, a mean difference of -0.019 mm and COR of 0.052 (P = .06) was found. In terms of repeatability, there was no significant difference in measurement of any variable with any device. Vector analysis showed no significant difference for corneal astigmatism measurements (P = .84).
CONCLUSION: There was no statistically significant difference in compared biometric parameters performed with all 3 devices.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28263734     DOI: 10.1016/j.ajo.2017.02.027

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  24 in total

1.  Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness.

Authors:  Xuepei Li; Yijing Zhou; Charlotte Aimee Young; Aiming Chen; Guangming Jin; Danying Zheng
Journal:  Ann Transl Med       Date:  2020-07

2.  Accuracy of segmented measurement of axial length in ultra-high myopia filled with silicone oil using immersion B-scan ultrasonography.

Authors:  Qing-Hua Yang; Hong-Tao Zhang; Xiao-Qi Li; Bing Chen; Zhao-Hui Li; Yi-Fei Huang; Xin Jin; Ying Zhang; Li-Qiang Wang
Journal:  Int J Ophthalmol       Date:  2022-05-18       Impact factor: 1.645

3.  Effectiveness, Sensitivity, and Specificity of Intraocular Lens Power Calculation Formulas for Short Eyes.

Authors:  Wiktor Stopyra
Journal:  Turk J Ophthalmol       Date:  2022-06-29

4.  Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging.

Authors:  Giacomo Savini; Leonardo Taroni; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  Eye (Lond)       Date:  2020-11-02       Impact factor: 4.456

Review 5.  Optimizing outcomes with toric intraocular lenses.

Authors:  Manpreet Kaur; Farin Shaikh; Ruchita Falera; Jeewan S Titiyal
Journal:  Indian J Ophthalmol       Date:  2017-12       Impact factor: 1.848

6.  Internal Astigmatism and Its Role in the Growth of Axial Length in School-Age Children.

Authors:  Liangcheng Wu; Chenghai Weng; Fei Xia; Xiaoying Wang; Xingtao Zhou
Journal:  J Ophthalmol       Date:  2018-04-18       Impact factor: 1.909

7.  Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry.

Authors:  Tomoaki Higashiyama; Hazuki Mori; Fumi Nakajima; Masahito Ohji
Journal:  PLoS One       Date:  2018-04-24       Impact factor: 3.240

8.  Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation.

Authors:  Samuel F Passi; Atalie C Thompson; Preeya K Gupta
Journal:  Clin Ophthalmol       Date:  2018-11-01

9.  Swivel haptics anterior capsule-fixated intraocular lens.

Authors:  Sri Ganesh; Sheetal Brar
Journal:  Indian J Ophthalmol       Date:  2018-09       Impact factor: 1.848

10.  Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices.

Authors:  Miki Kamikawatoko Omoto; Hidemasa Torii; Sachiko Masui; Masahiko Ayaki; Kazuo Tsubota; Kazuno Negishi
Journal:  Sci Rep       Date:  2019-04-25       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.