Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices.

Remzi Karadag1, Murat Unluzeybek2, Ozgur Cakici2, Ayse Yagmur Kanra2, Huseyin Bayramlar3.   

Abstract

AIM: The aim of the study is to compare the measurements of central corneal thickness (CCT) performed by two examiners with four different methods at different times inter- and intra-individually.
METHODS: Thirty healthy people were included in the study. In these measurements, an optical low-coherence reflectometry (OLCR), an optic coherence tomography (OCT), a specular microscopy (SM), and a corneal topography (CT) were used. Two examiners performed the measurements in a consecutive manner. After 1-7 days of the first measurements, the second measurements were performed again consecutively. The mean of three measurements was taken in each session for all devices.
RESULTS: In OCT measurements, there was a significant difference between two examiners in both sessions (p < 0.001), while no significant differences were found between two examiners in first and second sessions in SM, CT, and OLCR measurements. When each examiner's measurements were compared to two sessions, there were no significant differences (p > 0.05, for all) except the SM measurements of the first examiner (p = 0.041). When the first measurements of two examiners were compared, the smallest values were of OCT. At the first session of two examiners, there was a significant difference between OCT and CT measurements, and between OCT and OLCR (p < 0.001, p = 0.002 for the first examiner and p < 0.001 for the second examiner, respectively).
CONCLUSION: Our study showed that CCT measurements made by CT and OLCR methods were almost same and highly correlated for both the examiners' measurements. CCTs measured by OCT were on average 30 μm thinner than CT and OLCR.

Entities:  

Keywords:  Central corneal thickness; Corneal topography; Optic coherence tomography; Optical low-coherence reflectometry; Pachymetry; Specular microscopy

Mesh:

Year:  2016        PMID: 27723007      PMCID: PMC5517583          DOI: 10.1007/s10792-016-0369-2

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


Introduction

Central corneal thickness (CCT) has a great deal of importance in the diagnosis and follow-up of glaucoma and refractive surgical interventions such as cross-linking [1]. Furthermore, measurement of CCT is one of the indirect methods to evaluate whether the cornea endothelium is healthy or not. Therefore, CCT measurement, which can simply be performed through various methods, is also used in the follow-up of some diseases such as endothelial dystrophies [2]. Ultrasonic Pachymeter is the gold standard. In this method, however, there are some limitations, such as direct contact of the equipment with the cornea, measurement of the estimated central point, limited definition of the thinnest point of the cornea, and variation according to the individual performing the measurement. Currently, devices such as optical, low-coherence reflectometry, optical coherence tomography, noncontact specular microscopy, and corneal topography have been developed to measure CCT with a noncontact optic method [3]. These noncontact methods are less dependent on the operator and have the advantage of demonstrating the thinnest parts of the cornea. Thus, the reliability of the CCT measurements is very important in the diagnosis and management of various corneal diseases [2]. The aim of this study is to compare the measurements and operating individuals in terms of measurement values obtained in four different noncontact methods of CCT measurement performed by two different individuals, to compare the measurements they perform in different time periods, and to evaluate whether these methods can be interchangeable or not.

Materials and methods

Thirty individuals who applied to our clinic were included in the study. All cases were informed about the study, and informed consent was obtained. This study was performed in accordance with the Helsinki Declaration criteria. Thirty right eyes of 14 male and 16 female patients without any systemic diseases and aged between 22 and 60 years were included in the study. Patients with prior ocular surgery, the presence of glaucoma, refractive errors greater than ±2 diopters, cataracts, corneal opacity, and keratoconus were excluded from the study. Optical low-coherence reflectometry (OLCR; LenStar LS900; Haag-Streit AG, Koeniz, Switzerland), optical coherence tomography (OCT; Topcon 3D OCT-2000, Topcon Medical System, Oakland, New Jersey, USA), noncontact Specular Microscopy (SM; Topcon SP-3000P; Topcon Corporation, Tokyo, Japan), and corneal topography (CT; Sirius; Costruzione Strumenti Oftalmici, Florence, Italy) equipments were used to measure CCT. Corneal thickness measurements were performed by two individuals consecutively. One to seven days after the first measurement, second measurements were performed in all patients by the same two examiners. Measurements were performed between 10 a.m. and 2 p.m. to avoid diurnal variations. Three measurements were performed on each piece of equipment, and the means were calculated. SPSS 16 program was used in the statistical analysis, and parametric tests that were used for the distribution of the data were normal. The level of significance was accepted as p < 0.05.

Results

Fourteen males and 16 females, with a mean age of 30.97 ± 7.75 years, were included in the study. The results of the first and second measurements performed by the first and second investigators are shown in Table 1. A statistically significant correlation was present between the first and second measurements conducted by both investigators in all equipments (in all: p < 0.001). A powerful correlation was found between the measurements of the two sessions in the SM, CT, and OLCR methods, while a correlation of an intermediate level was present between the measurements performed using OCT equipment (r: 0.747, r: 0.691, Table 1, respectively).
Table 1

Correlation between researchers of both the first and second measurements in the same device (r: Pearson correlation coefficient)

MethodsFirst session (μm)Second session (μm) r value p value
First examiner
 OCT514.18 ± 25.60512.60 ± 20.850.747<0.001
 Specular microscopy525.54 ± 34.64523.21 ± 34.510.985<0.001
 Corneal topography547.46 ± 32.96546.61 ± 31.440.989<0.001
 OLCR544.25 ± 31.16544.64 ± 30.900.987<0.001
Second examiner
 OCT502.06 ± 22.33499.24 ± 22.290.691<0.001
 Specular microscopy524.53 ± 34.60524.13 ± 33.810.988<0.001
 Corneal topography546.69 ± 32.14546.12 ± 31.220.989<0.001
 OLCR544.81 ± 30.96544.45 ± 32.490.986<0.001

OCT optic coherence tomography, OLCR optical low-coherence reflectometry

Correlation between researchers of both the first and second measurements in the same device (r: Pearson correlation coefficient) OCT optic coherence tomography, OLCR optical low-coherence reflectometry In terms of the measurements of the same individual in both sessions, no significant differences were found between any first and second session measurements of the same individual (p > 0.05 for all, Table 2), except the SM measurements of the first investigator (p = 0.041).
Table 2

Statistical differences between researchers of both the first and second measurement values of the same unit

DevicesFirst session mean (range)Second session mean (range) p value
First examiner (μm)
 OCT514.18 ± 25.60 (455.67–553.00)512.60 ± 20.85 (469.00–548.00)0.617
 Specular microscopy525.54 ± 34.64 (462.33–579.00)523.21 ± 34.51 (460.00–573.00) 0.041
 Corneal topography547.46 ± 32.96 (497.33–602.33)546.61 ± 31.44 (492.00–600.00)0.355
 OLCR544.25 ± 31.16 (494.33–594.00)544.64 ± 30.90 (491.67–597.33)0.672
Second examiner
 OCT502.06 ± 22.33 (459.33–546.67)499.24 ± 22.29 (460.67–551.33)0.387
 Specular microscopy524.53 ± 34.60 (466.00–581.00)524.13 ± 33.81 (456.67–573.67)0.689
 Corneal topography546.69 ± 32.14 (492.33–597.33)546.12 ± 31.22 (494.67–593.67)0.521
 OLCR544.81 ± 30.96 (494.00–598.00)544.45 ± 32.49 (486.33–592.00)0.716
Differences between two examiners
 OCT14.37 ± 9.52 (0.33–35.00)16.87 ± 14.16 (1.66–58.00)0.759
 Specular microscopy3.86 ± 4.30 (0.00–20.00)3.30 ± 2.52 (0.33–10.34)0.474
 Corneal topography2.81 ± 2.35 (0.00–9.34)2.60 ± 1.99 (0.00–6.66)0.725
 OLCR2.80 ± 2.29 (0.00–9.00)3.58 ± 3.59 (0.00–16.66)0.933

Bold values are statistically significant measurements

OCT optic coherence tomography, OLCR optical low-coherence reflectometry

Statistical differences between researchers of both the first and second measurement values of the same unit Bold values are statistically significant measurements OCT optic coherence tomography, OLCR optical low-coherence reflectometry When the two investigators were compared with each other based on the measurements obtained, the measurements by both investigators using SM, CT, and OLCR were similar (in all: p > 0.05); however, OCT measurements by the two investigators were different from each other, both in the first (p < 0.001) and second sessions (p < 0.001) (Table 3). When the first and second session measurements performed using different types of equipment were compared to each other, the difference between the two sessions was statistically significant in the OCT measurements, but no difference was found in the two measurements in other equipment (Table 3). SM, CT, and OLCR measurements demonstrated a powerful correlation between the two investigators in both sessions. OCT measurements by the two investigators were highly correlated in the first session (r: 0.876) and intermediately correlated in the second session (r: 0.668) (Table 4).
Table 3

Statistical differences between the two session’s measurement values of the examiners in the same method

MethodsFirst examiner mean (range)Second examiner mean (range) p value
First session (μm)
 OCT514.18 ± 25.60 (455.67–553.00)502.06 ± 22.33 (459.33–546.67) <0.001
 Specular microscopy525.54 ± 34.64 (462.33–579.00)524.53 ± 34.60 (466.00–581.00)0.341
 Corneal topography547.46 ± 32.96 (497.33–602.33)546.69 ± 32.14 (492.33–597.33)0.255
 OLCR544.25 ± 31.16 (494.33–594.00)544.81 ± 30.96 (494.00–598.00)0.405
Second session (μm)
 OCT512.60 ± 20.85 (469.00–548.00)499.24 ± 22.29 (460.67–551.33) <0.001
 Specular microscopy523.21 ± 34.51 (460.00–573.00)524.13 ± 33.81 (456.67–573.67)0.227
 Corneal topography546.61 ± 31.44 (492.00–600.00)546.12 ± 31.22 (494.67–593.67)0.419
 OLCR544.64 ± 30.90 (491.67–597.33)544.45 ± 32.49 (486.33–592.00)0.832
Differences between two seasons
 OCT13.15 ± 10.80 (0.00–38.00)12.32 ± 12.61 (0.00–51.33)0.349
 Specular microscopy4.71 ± 4.29 (1.00–19.67)3.93 ± 3.66 (0.00–16.66)0.531
 Corneal topography3.89 ± 3.01 (0.00–12.33)3.63 ± 3.01 (0.33–15.34)0.721
 OLCR3.85 ± 3.13 (0.34–12.33)3.90 ± 3.76 (0.33–17.34)0.302

Bold values are statistically significant measurements

OCT optic coherence tomography, OLCR optical low-coherence reflectometry

Table 4

Correlation between both session to the values of the researchers in the same device (r: Pearson correlation coefficient)

MethodsFirst examinerSecond examiner r value p value
First session (μm)
 OCT514.18 ± 25.60502.06 ± 22.330.876<0.001
 Specular microscopy525.54 ± 34.64524.53 ± 34.600.986<0.001
 Corneal topography547.46 ± 32.96546.69 ± 32.140.994<0.001
 OLCR544.25 ± 31.16544.81 ± 30.960.993<0.001
Second session (μm)
 OCT512.60 ± 20.85499.24 ± 22.290.668<0.001
 Specular microscopy523.21 ± 34.51524.13 ± 33.810.993<0.001
 Corneal topography546.61 ± 31.44546.12 ± 31.220.995<0.001
 OLCR544.64 ± 30.90544.45 ± 32.490.988<0.001

OCT optic coherence tomography, OLCR optical low-coherence reflectometry

Statistical differences between the two session’s measurement values of the examiners in the same method Bold values are statistically significant measurements OCT optic coherence tomography, OLCR optical low-coherence reflectometry Correlation between both session to the values of the researchers in the same device (r: Pearson correlation coefficient) OCT optic coherence tomography, OLCR optical low-coherence reflectometry A significant difference was also found between the OCT measurements and CT and OLCR measurements of the first session by both investigators (for the measurements of the first investigator: p < 0.001, p = 0.002 and for the measurements of second investigator p < 0.001, p < 0.001, respectively). The measurements of the first investigator using OCT and SM equipment demonstrated no difference (p = 0.972), except for between the measurements of the second investigator (p = 0.029), (Table 5).
Table 5

A comparison (P values) of the first session of the researchers measurements (Bonferroni multiple comparison)

DevicesOCTSpecular microscopyCorneal topographyOLCR
First examiner
 OCT (514.18 ± 25.60 µm)0.972<0.0010.002
 Specular microscopy (525.54 ± 34.64 µm)0.9720.0460.134
 Corneal topography (547.46 ± 32.96 µm)<0.0010.0461.000
 OLCR (544.25 ± 31.16 µm)0.0020.1341.000
Second examiner
 OCT (502.06 ± 22.33 µm)0.029<0.001<0.001
 Specular microscopy (524.53 ± 34.60 µm)0.0290.0330.066
 Corneal topography (546.69 ± 32.14 µm)<0.0010.0331.000
 OLCR (544.81 ± 30.96 µm)<0.0010.0661.000

OCT optic coherence tomography, OLCR optical low-coherence reflectometry

A comparison (P values) of the first session of the researchers measurements (Bonferroni multiple comparison) OCT optic coherence tomography, OLCR optical low-coherence reflectometry When the SM and CT measurements by both investigators were compared, significant differences were found between the measurements (p = 0.046 and p = 0.033 for the first and second investigators, respectively (Table 5). No significant difference was found between the SM and OLCR measurements (p = 0.134 for the first investigator, p = 0.066 for the second investigator) (Table 5). No significant differences were also found between the CT and OLCR measurements by both investigators (p = 1.0) (Table 5). When the correlations of the results of the different methods were analyzed, the first measurements by both investigators were statistically significant (in all: p < 0.001) (Table 6).
Table 6

The researchers compared in each of the first measurements of all devices (r: Pearson Correlation Coefficient)

DevicesOCTSpecular microscopyCorneal topographyOLCR
r p r p r p r p
First researcher
 OCT0.655<0.0010.627<0.0010.644<0.001
 Specular microscopy0.655<0.0010.966<0.0010.974<0.001
 Corneal topography0.627<0.0010.966<0.0010.984<0.001
 OLCR0.644<0.0010.974<0.0010.984<0.001
Second researcher
 OCT0.654<0.0010.678<0.0010.632<0.001
 Specular microscopy0.654<0.0010.976<0.0010.984<0.001
 Corneal topography0.678<0.0010.976<0.0010.980<0.001
 OLCR0.632<0.0010.984<0.0010.980<0.001

OCT optic coherence tomography, OLCR optical low-coherence reflectometry

The researchers compared in each of the first measurements of all devices (r: Pearson Correlation Coefficient) OCT optic coherence tomography, OLCR optical low-coherence reflectometry The results of SM, CT, and OLCR measurements were highly correlated with each other, while the results of OCT were intermediately correlated with the other three methods (Table 6).

Discussion

Until recently, the single method used in the measurement of corneal thickness was an ultrasound-based method. However, this method has some disadvantages such as relatively variable results due to the difficulties in centralization and accommodation, the necessity for topical anesthesia, the possibility of epithelial damage due to contact between the probe and the cornea, and the risk of infection [4, 5]. CCT was reported to show a ±10-μm variation following topical anesthesia application [6, 7]. Advances in the diagnostic technologies and increased interest in corneal refractive surgical techniques resulted in the development of less invasive and noncontact methods for the measurement of the corneal thickness [8]. Since CCT measurements might be performed using many types of noncontact equipment, different types of equipments are available in different clinics. For the follow-up of the patients, a correlation of the results of the measurements performed by the equipment among each other and within a single piece of equipment is still important. In the study of Borrego-Sanz L et al. which included 76 healthy patients using Pentacam, Ultrasound Pachymetry, Specular Microscopy, and Optic Biometer Lenstar LS 900 equipment, the CCT measurement results were reported as 565.36 ± 36.34, 549.61 ± 32.64, 537.24 ± 36.34, and 443.06 ± 34.28 μm, respectively. They found the highest correlation between Lenstar and UP [18]. Similarly, Tai et al. also found a high correlation between Lenstar and UP in their study that was performed on 184 eyes [9]. O’Donnell et al. [10] and Zhao et al. [11] found a high correlation between CCT measurements performed by Pentacam and Lenstar; however, they reported that these two devices could not be used in place of each other in clinical applications. Although a good correlation has been reported in UP and Pentacam in many studies, Tai et al. [19] found that Pentacam values were 10 μm higher than the UP values. They suggested such a difference may have originated from the fact that lacrima is also measured during measurement by Pentacam, and that UP applies pressure on the cornea. Moreover, Almubrad et al. [6] reported that CCT values measured by SM were lower than the values of UP (28.17 ± 19.20 μm), and they concluded that the two methods could not be used in place of each other. Huang et al. [12] measured CCT in 66 healthy corneas and reported the results as follows: 538.82 ± 26.46 μm with Pentacam, 542.14 ± 27.12 μm with Sirius, 548.10 ± 26.41 μm with Galilei, and 532.81 ± 26.24 μm with RTVue. Repeatability and reproducibility were very high in these four devices, and it was suggested to be adequate for clinical use. Although the measurements of these devices were not correlated with each other since they were different, they emphasized that Galilei yielded better results. Additionally, using similar optic methods (Pentacam and RTVue), Nam et al. [13] and Chen et al. [14] achieved high intraobserver repeatability and interobserver reproducibility. Since there are no studies of intra- and interobserver comparisons using the equipment we used in this study, it was not possible to perform a one-to-one comparison with other studies. The present study revealed that interobserver repeatability of OCT was poor, and the differences between the measurements were significant. Interobserver repeatability of other methods was good. We found interobserver and intraobserver repeatability of the measurements performed by OLCR and CT devices, and this value was less than 1 µm. Milla et al. [15] and Savini et al. [16] found high repeatability in CCT measurements performed by Sirius and Galilei in the same order. Also, Khoramnia et al. [17] and Milla et al. [15] reported good repeatability in the same order using Pentacam and Sirius. Savini et al. [18] found the repeatability of Sirius slightly poorer than Galilei and slightly better than Pentacam. Many investigators [19, 20] found CCT measurement repeatability of Pentacam to be poorer compared to FD-OCT. A probable cause of this was reported to be the more rapid screening capacity of OCT. In the present study, the results of measurements by OCT were lower than the values obtained with other equipment. Furthermore, while OCT measurement results were intermediate–high correlated with the results of other equipment, the results of other equipment were highly correlated with each other. Current and some previous studies have suggested that different devices for CCT evaluation cannot be used in place of each other [21-23]. However, there are also studies reporting correlations from intermediate to excellent levels between Galilei and Orbscan II, and between Pentacam and Orbscan II [16, 24, 25]. In one study, the results of Orbscan II were markedly lower than the measurements performed using Pentacam and Galilei [26]. On the contrary, Sedaghat et al. [27] found the CCT values measured with Orbscan II to be slightly higher than the values obtained using Pentacam. Bueh et al. found no significant differences between the measurements of these two pieces of equipment [28]. However, in another study by Amano et al., no significant difference was found between the measurements of Orbscan II and Pentacam [29]. Kim et al. found that the CCT measurements prior to photorefractive keratectomy by Orbscan II in myopic patients were similar with the results of Pentacam; however, they reported that postoperative early and late measurements by Orbscan II resulted in thinner values [30]. In conclusion, values obtained in the present study by CT and OLCR methods taken by different individuals and the measurements obtained by the same individuals were almost similar and highly correlated. The difference between the results of these two measurement methods was approximately 1 µm. Since a significant difference was present in the results of the measurements of two different individuals performed by OCT, the reliability of this method in interobserver measurements in the evaluation of CCT is lower compared to the other methods according to our study. In addition, OCT measurements yield the lowest values compared to the other methods. We consider that the presence of a difference of at least 30 μm between the measurement values of CT and OCLR, and measurements of OCT will affect the calculations for corrected intraocular pressure and plan the operations in refractive surgery.
  29 in total

1.  Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound.

Authors:  D Gherghel; S L Hosking; S Mantry; S Banerjee; S A Naroo; S Shah
Journal:  J Cataract Refract Surg       Date:  2004-06       Impact factor: 3.351

2.  Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound.

Authors:  Birgit Lackner; Gerald Schmidinger; Stefan Pieh; Martin A Funovics; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-10       Impact factor: 1.973

3.  Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.

Authors:  Shiro Amano; Norihiko Honda; Yuki Amano; Satoru Yamagami; Takashi Miyai; Tomokazu Samejima; Miyuki Ogata; Kazunori Miyata
Journal:  Ophthalmology       Date:  2006-06       Impact factor: 12.079

4.  Reliability of peripheral corneal pachymetry with the Oculus Pentacam.

Authors:  Hema Shankar; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-01       Impact factor: 3.351

5.  Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes.

Authors:  Giacomo Savini; Michele Carbonelli; Piero Barboni; Kenneth J Hoffer
Journal:  J Cataract Refract Surg       Date:  2011-02       Impact factor: 3.351

6.  A comprehensive comparison of central corneal thickness measurement.

Authors:  Jinhai Huang; Konrad Pesudovs; Ayong Yu; Thomas Wright; Daizong Wen; Ming Li; Ye Yu; Qinmei Wang
Journal:  Optom Vis Sci       Date:  2011-08       Impact factor: 1.973

7.  Central corneal thickness measurements with different imaging devices and ultrasound pachymetry.

Authors:  Lai-Yong Tai; Keat-Ween Khaw; Choung-Min Ng; Visvaraja Subrayan
Journal:  Cornea       Date:  2013-06       Impact factor: 2.651

8.  A Comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Chao Pan; Yifan Feng; Dan Cheng; Yanjun Hua; Xiuli Hu; Qinmei Wang
Journal:  Ophthalmology       Date:  2013-05-11       Impact factor: 12.079

9.  Comparison of three methods of measuring corneal thickness and anterior chamber depth.

Authors:  Wolf Buehl; Danijela Stojanac; Stefan Sacu; Wolfgang Drexler; Oliver Findl
Journal:  Am J Ophthalmol       Date:  2006-01       Impact factor: 5.258

10.  Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter.

Authors:  Turki M Almubrad; Uchechukwu L Osuagwu; Ibrahim Alabbadi; Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2011-06-24
View more
  1 in total

1.  Comparison of central corneal thickness measurements obtained by community optometrists to those obtained in secondary care.

Authors:  Laura Hughes; Cheng Yi Loo; Brian Hogan; Huai Ling Tan; Ling Shan Tang; Andrew J Tatham
Journal:  Eye (Lond)       Date:  2018-07-23       Impact factor: 3.775

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.