Remzi Karadag1, Murat Unluzeybek2, Ozgur Cakici2, Ayse Yagmur Kanra2, Huseyin Bayramlar3.
Abstract
AIM: The aim of the study is to compare the measurements of central corneal thickness (CCT) performed by two examiners with four different methods at different times inter- and intra-individually.Entities:
Keywords: Central corneal thickness; Corneal topography; Optic coherence tomography; Optical low-coherence reflectometry; Pachymetry; Specular microscopy
Mesh:
Year: 2016 PMID: 27723007 PMCID: PMC5517583 DOI: 10.1007/s10792-016-0369-2
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.031
Correlation between researchers of both the first and second measurements in the same device (r: Pearson correlation coefficient)
| Methods | First session (μm) | Second session (μm) |
|
|
|---|---|---|---|---|
| First examiner | ||||
| OCT | 514.18 ± 25.60 | 512.60 ± 20.85 | 0.747 | <0.001 |
| Specular microscopy | 525.54 ± 34.64 | 523.21 ± 34.51 | 0.985 | <0.001 |
| Corneal topography | 547.46 ± 32.96 | 546.61 ± 31.44 | 0.989 | <0.001 |
| OLCR | 544.25 ± 31.16 | 544.64 ± 30.90 | 0.987 | <0.001 |
| Second examiner | ||||
| OCT | 502.06 ± 22.33 | 499.24 ± 22.29 | 0.691 | <0.001 |
| Specular microscopy | 524.53 ± 34.60 | 524.13 ± 33.81 | 0.988 | <0.001 |
| Corneal topography | 546.69 ± 32.14 | 546.12 ± 31.22 | 0.989 | <0.001 |
| OLCR | 544.81 ± 30.96 | 544.45 ± 32.49 | 0.986 | <0.001 |
OCT optic coherence tomography, OLCR optical low-coherence reflectometry
Statistical differences between researchers of both the first and second measurement values of the same unit
| Devices | First session mean (range) | Second session mean (range) |
|
|---|---|---|---|
| First examiner (μm) | |||
| OCT | 514.18 ± 25.60 (455.67–553.00) | 512.60 ± 20.85 (469.00–548.00) | 0.617 |
| Specular microscopy | 525.54 ± 34.64 (462.33–579.00) | 523.21 ± 34.51 (460.00–573.00) |
|
| Corneal topography | 547.46 ± 32.96 (497.33–602.33) | 546.61 ± 31.44 (492.00–600.00) | 0.355 |
| OLCR | 544.25 ± 31.16 (494.33–594.00) | 544.64 ± 30.90 (491.67–597.33) | 0.672 |
| Second examiner | |||
| OCT | 502.06 ± 22.33 (459.33–546.67) | 499.24 ± 22.29 (460.67–551.33) | 0.387 |
| Specular microscopy | 524.53 ± 34.60 (466.00–581.00) | 524.13 ± 33.81 (456.67–573.67) | 0.689 |
| Corneal topography | 546.69 ± 32.14 (492.33–597.33) | 546.12 ± 31.22 (494.67–593.67) | 0.521 |
| OLCR | 544.81 ± 30.96 (494.00–598.00) | 544.45 ± 32.49 (486.33–592.00) | 0.716 |
| Differences between two examiners | |||
| OCT | 14.37 ± 9.52 (0.33–35.00) | 16.87 ± 14.16 (1.66–58.00) | 0.759 |
| Specular microscopy | 3.86 ± 4.30 (0.00–20.00) | 3.30 ± 2.52 (0.33–10.34) | 0.474 |
| Corneal topography | 2.81 ± 2.35 (0.00–9.34) | 2.60 ± 1.99 (0.00–6.66) | 0.725 |
| OLCR | 2.80 ± 2.29 (0.00–9.00) | 3.58 ± 3.59 (0.00–16.66) | 0.933 |
Bold values are statistically significant measurements
OCT optic coherence tomography, OLCR optical low-coherence reflectometry
Statistical differences between the two session’s measurement values of the examiners in the same method
| Methods | First examiner mean (range) | Second examiner mean (range) |
|
|---|---|---|---|
| First session (μm) | |||
| OCT | 514.18 ± 25.60 (455.67–553.00) | 502.06 ± 22.33 (459.33–546.67) |
|
| Specular microscopy | 525.54 ± 34.64 (462.33–579.00) | 524.53 ± 34.60 (466.00–581.00) | 0.341 |
| Corneal topography | 547.46 ± 32.96 (497.33–602.33) | 546.69 ± 32.14 (492.33–597.33) | 0.255 |
| OLCR | 544.25 ± 31.16 (494.33–594.00) | 544.81 ± 30.96 (494.00–598.00) | 0.405 |
| Second session (μm) | |||
| OCT | 512.60 ± 20.85 (469.00–548.00) | 499.24 ± 22.29 (460.67–551.33) |
|
| Specular microscopy | 523.21 ± 34.51 (460.00–573.00) | 524.13 ± 33.81 (456.67–573.67) | 0.227 |
| Corneal topography | 546.61 ± 31.44 (492.00–600.00) | 546.12 ± 31.22 (494.67–593.67) | 0.419 |
| OLCR | 544.64 ± 30.90 (491.67–597.33) | 544.45 ± 32.49 (486.33–592.00) | 0.832 |
| Differences between two seasons | |||
| OCT | 13.15 ± 10.80 (0.00–38.00) | 12.32 ± 12.61 (0.00–51.33) | 0.349 |
| Specular microscopy | 4.71 ± 4.29 (1.00–19.67) | 3.93 ± 3.66 (0.00–16.66) | 0.531 |
| Corneal topography | 3.89 ± 3.01 (0.00–12.33) | 3.63 ± 3.01 (0.33–15.34) | 0.721 |
| OLCR | 3.85 ± 3.13 (0.34–12.33) | 3.90 ± 3.76 (0.33–17.34) | 0.302 |
Bold values are statistically significant measurements
OCT optic coherence tomography, OLCR optical low-coherence reflectometry
Correlation between both session to the values of the researchers in the same device (r: Pearson correlation coefficient)
| Methods | First examiner | Second examiner |
|
|
|---|---|---|---|---|
| First session (μm) | ||||
| OCT | 514.18 ± 25.60 | 502.06 ± 22.33 | 0.876 | <0.001 |
| Specular microscopy | 525.54 ± 34.64 | 524.53 ± 34.60 | 0.986 | <0.001 |
| Corneal topography | 547.46 ± 32.96 | 546.69 ± 32.14 | 0.994 | <0.001 |
| OLCR | 544.25 ± 31.16 | 544.81 ± 30.96 | 0.993 | <0.001 |
| Second session (μm) | ||||
| OCT | 512.60 ± 20.85 | 499.24 ± 22.29 | 0.668 | <0.001 |
| Specular microscopy | 523.21 ± 34.51 | 524.13 ± 33.81 | 0.993 | <0.001 |
| Corneal topography | 546.61 ± 31.44 | 546.12 ± 31.22 | 0.995 | <0.001 |
| OLCR | 544.64 ± 30.90 | 544.45 ± 32.49 | 0.988 | <0.001 |
OCT optic coherence tomography, OLCR optical low-coherence reflectometry
A comparison (P values) of the first session of the researchers measurements (Bonferroni multiple comparison)
| Devices | OCT | Specular microscopy | Corneal topography | OLCR | |
|---|---|---|---|---|---|
| First examiner | |||||
| OCT (514.18 ± 25.60 µm) | 0.972 | <0.001 | 0.002 | ||
| Specular microscopy (525.54 ± 34.64 µm) | 0.972 | 0.046 | 0.134 | ||
| Corneal topography (547.46 ± 32.96 µm) | <0.001 | 0.046 | 1.000 | ||
| OLCR (544.25 ± 31.16 µm) | 0.002 | 0.134 | 1.000 | ||
| Second examiner | |||||
| OCT (502.06 ± 22.33 µm) | 0.029 | <0.001 | <0.001 | ||
| Specular microscopy (524.53 ± 34.60 µm) | 0.029 | 0.033 | 0.066 | ||
| Corneal topography (546.69 ± 32.14 µm) | <0.001 | 0.033 | 1.000 | ||
| OLCR (544.81 ± 30.96 µm) | <0.001 | 0.066 | 1.000 | ||
OCT optic coherence tomography, OLCR optical low-coherence reflectometry
The researchers compared in each of the first measurements of all devices (r: Pearson Correlation Coefficient)
| Devices | OCT | Specular microscopy | Corneal topography | OLCR | ||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| |
| First researcher | ||||||||
| OCT | 0.655 | <0.001 | 0.627 | <0.001 | 0.644 | <0.001 | ||
| Specular microscopy | 0.655 | <0.001 | 0.966 | <0.001 | 0.974 | <0.001 | ||
| Corneal topography | 0.627 | <0.001 | 0.966 | <0.001 | 0.984 | <0.001 | ||
| OLCR | 0.644 | <0.001 | 0.974 | <0.001 | 0.984 | <0.001 | ||
| Second researcher | ||||||||
| OCT | 0.654 | <0.001 | 0.678 | <0.001 | 0.632 | <0.001 | ||
| Specular microscopy | 0.654 | <0.001 | 0.976 | <0.001 | 0.984 | <0.001 | ||
| Corneal topography | 0.678 | <0.001 | 0.976 | <0.001 | 0.980 | <0.001 | ||
| OLCR | 0.632 | <0.001 | 0.984 | <0.001 | 0.980 | <0.001 | ||
OCT optic coherence tomography, OLCR optical low-coherence reflectometry