PURPOSE: To compare central corneal thickness (CCT) values obtained by Lenstar (LE), Pentacam (PC), specular microscopy (SM) and ultrasound pachymetry (UP) in healthy corneas and study their influence on intraocular pressure (IOP) readings determined by Goldmann applanation tonometry (GAT).
METHODS: CCT was measured in 76 healthy subjects by LE, PC, SM and UP. We established Lin's concordance correlation coefficient (ρ-C) between different techniques. The influence of CCT on GAT was established through univariate linear regression models, IOP being the dependent variable.
RESULTS: The highest ρ-C was found between LE and SM at 0.94 (95% CI: 0.91-0.96) and between LE and UP at 0.95 (95% CI: 0.94-0.97). IOP readings showed less variability when CCT was determined using LE (7.7%, B = 0.16; 95% CI: 0.004-0.28).
CONCLUSIONS: Although CCT values obtained with UP, PC, SM and LE show good correlation, these methods are not completely interchangeable. The amount of IOP variation differs when CCT is determined using LE or SM.
© 2014 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2014        PMID: 24642898     DOI: 10.1159/000356724

Source DB:  PubMed          Journal:  Ophthalmologica        ISSN: 0030-3755            Impact factor:   3.250


  3 in total

Review 1.  The thick and thin of the central corneal thickness in glaucoma.

Authors:  Graham W Belovay; Ivan Goldberg
Journal:  Eye (Lond)       Date:  2018-02-15       Impact factor: 3.775

2.  Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices.

Authors:  Remzi Karadag; Murat Unluzeybek; Ozgur Cakici; Ayse Yagmur Kanra; Huseyin Bayramlar
Journal:  Int Ophthalmol       Date:  2016-10-08       Impact factor: 2.031

3.  Corneal thickness measurement by secondary speckle tracking and image processing using machine-learning algorithms.

Authors:  Aviya Bennett; Elnatan Davidovitch; Yafim Beiderman; Sergey Agadarov; Yevgeny Beiderman; Avital Moshkovitz; Uri Polat; Zeev Zalevsky
Journal:  J Biomed Opt       Date:  2019-12       Impact factor: 3.170

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.