Comparison of central corneal thickness and anterior chamber depth measured using LenStar LS900, Pentacam, and Visante AS-OCT.

Clare OʼDonnell1, Andreas Hartwig, Hema Radhakrishnan.   

Abstract

PURPOSE: To compare central corneal thickness (CCT) and anterior chamber depth (ACD) measured using 3 different techniques.
METHODS: CCT and ACD were measured in one eye of 27 healthy subjects (age, mean ± SD: 34 ± 7 years), using the LenStar LS900, Pentacam, and Visante AS-OCT. The agreement between the measurement techniques was assessed using the coefficient of agreement (CoA) and the 95% limits of agreement (LoA).
RESULTS: Mean ± SD values for CCT using the LenStar, Pentacam, and Visante were 550.7 ± 37.4 μm, 542.7 ± 37.9 μm, and 556.7 ± 44.4 μm, respectively (P = 0.005). Plots of differences against means displayed relatively good agreement between the Pentacam and LenStar (CoA, 19.97 μm; LoA, 15.53 μm to -24.40 μm), poorer agreement between the LenStar and Visante (CoA, 31.26 μm; LoA, 40.78 μm to -21.74 μm), and agreement was poor between the Pentacam and Visante (CoA, 37.36 μm; LoA, 25.61 μm to -49.11 μm). Mean ± SD values for ACD using the LenStar, Pentacam, and Visante were 2.93 ± 0.30 mm, 2.96 ± 0.32 mm, and 3.03 ± 0.29 mm, respectively (P < 0.001). Relatively good agreement for ACD measures was observed between the Pentacam and LenStar (CoA, 0.08 mm; LoA, 0.12 to -0.04 mm), poor agreement was obtained between the Visante and LenStar (CoA, 0.13 mm; LoA, 0.21 to -0.06 mm), and agreement was also poor between the Pentacam and Visante (CoA, 0.14 mm; LoA, 0.11 to -0.17 mm).
CONCLUSIONS: Our data show that the LenStar, Pentacam, and Visante provide measurements that are in agreement with published values for CCT and ACD in human subjects. Although reasonable agreement for CCT and ACD was found between the Pentacam and LenStar, agreement was poorer between the Visante and LenStar and between the Visante and Pentacam. Thus, CCT and ACD measures from these instruments should not be used interchangeably.

Entities:  

Mesh:

Year:  2012        PMID: 22367044     DOI: 10.1097/ICO.0b013e31823f8e2f

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  10 in total

1.  Lens thickness assessment: anterior segment optical coherence tomography versus A-scan ultrasonography.

Authors:  Nikoo Hamzeh; Sasan Moghimi; Golshan Latifi; Massood Mohammadi; Nassim Khatibi; Shan C Lin
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

Review 2.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

3.  Comparison of anterior segment measurements with LenStar and Pentacam in patients with newly diagnosed glaucoma.

Authors:  Emine Sen; Merve Inanc; Ufuk Elgin; Pelin Yilmazbas
Journal:  Int Ophthalmol       Date:  2017-01-21       Impact factor: 2.031

4.  Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness.

Authors:  Xuepei Li; Yijing Zhou; Charlotte Aimee Young; Aiming Chen; Guangming Jin; Danying Zheng
Journal:  Ann Transl Med       Date:  2020-07

5.  Species Differences in the Geometry of the Anterior Segment Differentially Affect Anterior Chamber Cell Scoring Systems in Laboratory Animals.

Authors:  Sara M Thomasy; J Seth Eaton; Matthew J Timberlake; Paul E Miller; Steven Matsumoto; Christopher J Murphy
Journal:  J Ocul Pharmacol Ther       Date:  2015-11-05       Impact factor: 2.671

6.  Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth.

Authors:  Jinhai Huang; Weicong Lu; Giacomo Savini; Hao Chen; Chengfang Wang; Xinxin Yu; Fangjun Bao; Qinmei Wang
Journal:  J Ophthalmol       Date:  2016-06-14       Impact factor: 1.909

7.  Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices.

Authors:  Remzi Karadag; Murat Unluzeybek; Ozgur Cakici; Ayse Yagmur Kanra; Huseyin Bayramlar
Journal:  Int Ophthalmol       Date:  2016-10-08       Impact factor: 2.031

8.  Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation.

Authors:  Ting Wan; Houfa Yin; Yi Yang; Fang Wu; Zhiyi Wu; Yabo Yang
Journal:  BMC Ophthalmol       Date:  2019-08-13       Impact factor: 2.209

9.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

10.  Comparison of corneal thickness in patients with dry eye disease using the Pentacam rotating Scheimpflug camera and anterior segment optical coherence tomography.

Authors:  Keiichi Fujimoto; Takenori Inomata; Yuichi Okumura; Nanami Iwata; Kenta Fujio; Atsuko Eguchi; Ken Nagino; Hurramhon Shokirova; Maria Karasawa; Akira Murakami
Journal:  PLoS One       Date:  2020-02-03       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.