| Literature DB >> 24472443 |
Christos Chamos1, Liana Vele, Mark Hamilton, Maurizio Cecconi.
Abstract
The monitoring of the cardiac output (CO) and other hemodynamic parameters, traditionally performed with the thermodilution method via a pulmonary artery catheter (PAC), is now increasingly done with the aid of less invasive and much easier to use devices. When used within the context of a hemodynamic optimization protocol, they can positively influence the outcome in both surgical and non-surgical patient populations. While these monitoring tools have simplified the hemodynamic calculations, they are subject to limitations and can lead to erroneous results if not used properly. In this article we will review the commercially available minimally invasive CO monitoring devices, explore their technical characteristics and describe the limitations that should be taken into consideration when clinical decisions are made.Entities:
Year: 2013 PMID: 24472443 PMCID: PMC3964331 DOI: 10.1186/2047-0525-2-19
Source DB: PubMed Journal: Perioper Med (Lond) ISSN: 2047-0525
Figure 1The area under the curve (AUC) of the systolic phase is indicative of the SV. The SV is the shaded area under the systolic portion of the arterial pulse waveform. It is calculated as=[area under systolic phase+ aortic compliance] x shape of pressure curve.
Figure 2The transpulmonary thermodilution curve, with the characteristic delay in the peak temperature change compared to the PAC thermodilution. Based on the same principle as the PAC thermodilution, the transpulmonary method obviates the need for a PAC.
Figure 3Correlation of fluid responsiveness with SVV/PPV variables. As a general rule, the higher the SVV/PPV values, the more fluid responsive the patient will be. The correlation can only be applied to mechanically ventilated patients.
Overview and classification of minimally invasive CO monitors
| Pulse pressure analysis | | | |
| -Uncalibrated devices | Flo/Trac-Vigileo | Standard arterial line | GEDV, EVLW |
| Pulsioflex | Standard arterial line | ||
| LidCO rapid | Standard arterial line | ||
| -Calibrated devices | LidCO plus | Standard arterial line | |
| PiCCOplus | Standard arterial line | ||
| | Central venous line | ||
| Volumeview | Standard arterial line | ||
| | Central venous line | GEDV, EVLW, GEF | |
| Doppler | | | |
| -Transesophageal | CardioQ/CardioQ-OCM | Transesophageal probe | FTc |
| -Transthoracic | USCOM | Transthoracic probe | |
| Echocardiography | ClariTEE | Miniaturized transesophageal probe | |
| Gas rebreathing | NICO | Disposable rebreathing circuit | |
| Bioimpedance/ Bioreactance | | | |
| -Thoracic electrical bioimpedance | |||
| - Bioreactance | BioZ | Specific electrodes | |
| NICOM | Specific electrodes | ||
| Non-invasive monitors | | | |
| Nexfin | Finger probe | Hb | |
| Masimo | Specific pulse oximeter | PVI | |
| esCCO | ECG/ specific pulse oximeter | PWTT |