Comparison of noninvasive continuous arterial waveform analysis (Nexfin) with transthoracic Doppler echocardiography for monitoring of cardiac output.

Anna G E van der Spoel1, Albertus J Voogel, Anja Folkers, Christa Boer, R Arthur Bouwman.   

Abstract

STUDY
OBJECTIVES: To compare the Nexfin cardiac output (CO) with the CO obtained from transthoracic Doppler echocardiography (TTE) during routine cardiac function screening.
DESIGN: Observational clinical study.
SETTING: Echocardiography laboratory. PATIENTS: 40 ASA physical status 1 and 2 patients scheduled for routine TTE examination.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: In 40 patients scheduled for routine TTE examination, we obtained simultaneous CO measurements with Doppler ultrasound and derived from Nexfin blood pressure measurements. Correlation and level of agreement between Nexfin and TTE were analyzed using Pearson correlation coefficient and Bland-Altman plots. The Pearson correlation coefficient for Nexfin versus TTE was 0.68 (CI: 0.46 - 0.82, P < 0.0001). Bland-Altman analysis showed a bias of 0.51 ± 1.1 L/min and limits of agreement of -1.6 to 2.6 L/min, with a percentage error of 39%.
CONCLUSIONS: Considering limits of precision of CO measurements with Doppler echocardiography (± 30%), the agreement between noninvasive CO measurement with the Nexfin and TTE is reasonable.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22608585     DOI: 10.1016/j.jclinane.2011.09.008

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  19 in total

Review 1.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

Review 2.  Non-invasive continuous blood pressure monitoring: a review of current applications.

Authors:  Elena Chung; Guo Chen; Brenton Alexander; Maxime Cannesson
Journal:  Front Med       Date:  2013-01-23       Impact factor: 4.592

Review 3.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

Review 4.  Accuracy of non-invasive and minimally invasive hemodynamic monitoring: where do we stand?

Authors:  Issa Pour-Ghaz; Theodore Manolukas; Nathalie Foray; Joel Raja; Aranyak Rawal; Uzoma N Ibebuogu; Rami N Khouzam
Journal:  Ann Transl Med       Date:  2019-09

5.  The Feasibility of a Novel Index From a Wireless Doppler Ultrasound Patch to Detect Decreasing Cardiac Output in Healthy Volunteers.

Authors:  Jon-Émile S Kenny; Andrew M Eibl; Matthew Parrotta; Bradley F Long; Joseph K Eibl
Journal:  Mil Med       Date:  2021-01-25       Impact factor: 1.437

6.  Accuracy and trending ability of finger plethysmographic cardiac output monitoring in late pregnancy.

Authors:  Paul Bonnin; Benjamin Constans; Alain Duhamel; Maéva Kyheng; Anne-Sophie Ducloy-Bouthors; Max Gonzalez Estevez; Benoit Tavernier; Alexandre Gaudet
Journal:  Can J Anaesth       Date:  2022-08-04       Impact factor: 6.713

7.  The haemodynamic effects of intravenous paracetamol (acetaminophen) in healthy volunteers: a double-blind, randomized, triple crossover trial.

Authors:  Elizabeth Chiam; Laurence Weinberg; Michael Bailey; Larry McNicol; Rinaldo Bellomo
Journal:  Br J Clin Pharmacol       Date:  2016-01-25       Impact factor: 4.335

Review 8.  Noninvasive continuous hemodynamic monitoring.

Authors:  Jasper Truijen; Johannes J van Lieshout; Wilbert A Wesselink; Berend E Westerhof
Journal:  J Clin Monit Comput       Date:  2012-06-14       Impact factor: 2.502

9.  The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challenge.

Authors:  Xavier Monnet; Fabien Picard; Elsa Lidzborski; Malcie Mesnil; Jacques Duranteau; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2012-10-29       Impact factor: 9.097

10.  Non-invasive assessment of cardiac hemodynamics in patients with advanced cancer and with chronic heart failure: a pilot feasibility study.

Authors:  Stephan von Haehling; Mitja Lainscak; Thomas Kung; Larissa Cramer; Susann Fülster; Uwe Pelzer; Bert Hildebrandt; Anja Sandek; Joerg C Schefold; Mathias Rauchhaus; Wolfram Doehner; Stefan D Anker
Journal:  Arch Med Sci       Date:  2013-04-22       Impact factor: 3.318

View more