Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique.

Nirav Y Raval1, Pierre Squara, Michael Cleman, Kishore Yalamanchili, Michael Winklmaier, Daniel Burkhoff.   

Abstract

OBJECTIVES: Bioreactance, the analysis of intrabeat variations in phase of a transthoracic voltage in response to an applied high frequency transthoracic current, was recently introduced for noninvasive cardiac output measurement (NICOM). We evaluated NICOM compared to thermo- dilution (TD) in several clinical settings.
METHODS: 111 patients with a clinical indication for TD cardiac output (CO) measurement were recruited at five centers, including patients in cardiac catheterization laboratories, cardiac care units and intensive care units. CO measurements were made simulta- neously with TD and the bioreactance method and compared by regression analysis.
RESULTS: For studies in the intensive care units, TD-based CO and NICOM were highly correlated (r = 0.78, P < 0.0001) and did not differ significantly from each other (P = 0.55). Results in the cardiac catheterization laboratory were similar (r = 0.71, P < 0.001; P = 0.28 NICOM versus TD). In subsets of patients, NICOM was shown to be better corre- lated with TD-CO than CO obtained with the Fick method or with standard bioimpedance-based measurements of CO.
CONCLUSIONS: On average, compared to TD, bioreactance- based NICOM has acceptable accuracy in challenging clinical environments. Availability of such a tool may allow clinicians to have information about CO in patients where this information is not currently available to help diagnosis and guide therapy.

Entities:  

Mesh:

Year:  2008        PMID: 18340540     DOI: 10.1007/s10877-008-9112-5

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  20 in total

1.  Comparison of cardiac output determined by bioimpedance, thermodilution, and the Fick method.

Authors:  Milo Engoren; Daniel Barbee
Journal:  Am J Crit Care       Date:  2005-01       Impact factor: 2.228

2.  Cardiac output and cardiopulmonary responses to exercise in heart failure: application of a new bio-reactance device.

Authors:  Jonathan Myers; Pradeep Gujja; Suresh Neelagaru; Daniel Burkhoff
Journal:  J Card Fail       Date:  2007-10       Impact factor: 5.712

3.  Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients.

Authors:  E E Frezza; H Mezghebe
Journal:  Am Surg       Date:  1998-02       Impact factor: 0.688

4.  [Non-invasive versus invasive cardiovascular monitoring. Determination of stroke volume and pulmonary hydration using a new bioimpedance monitor].

Authors:  J Boldt; D Kling; A Thiel; G Hempelmann
Journal:  Anaesthesist       Date:  1988-04       Impact factor: 1.041

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Cardiac output measurement by the thermodilution method: an in vitro test of accuracy of three commercially available automatic cardiac output computers.

Authors:  A Rubini; D Del Monte; V Catena; I Attar; M Cesaro; D Soranzo; G Rattazzi; G L Alati
Journal:  Intensive Care Med       Date:  1995-02       Impact factor: 17.440

7.  Colonization and infection of pulmonary artery catheter in cardiac surgery patients: epidemiology and multivariate analysis of risk factors.

Authors:  G Kac; E Durain; C Amrein; E Hérisson; A Fiemeyer; A Buu-Hoï
Journal:  Crit Care Med       Date:  2001-05       Impact factor: 7.598

8.  A comparative evaluation of thermodilution and partial CO2 rebreathing techniques for cardiac output assessment in critically ill patients during assisted ventilation.

Authors:  Monica Rocco; Gustavo Spadetta; Andrea Morelli; Donatella Dell'Utri; Patrizia Porzi; Giorgio Conti; Paolo Pietropaoli
Journal:  Intensive Care Med       Date:  2003-12-03       Impact factor: 17.440

9.  Estimation of left ventricular systolic function by single transpulmonary thermodilution.

Authors:  Alain Combes; Jean-Baptiste Berneau; Charles-Edouard Luyt; Jean-Louis Trouillet
Journal:  Intensive Care Med       Date:  2004-04-23       Impact factor: 17.440

10.  Hemodynamic response to exercise as measured by the solar IKG impedance cardiography module and correlation with metabolic variables.

Authors:  Stephan Ziegeler; Ulrich Grundmann; Oliver Fuerst; Alexander Raddatz; Sascha Kreuer
Journal:  J Clin Monit Comput       Date:  2006-11-04       Impact factor: 1.977

View more
  56 in total

Review 1.  Cardiac output monitoring devices: an analytic review.

Authors:  Jahan Porhomayon; Ali El-Solh; Peter Papadakos; Nader Djalal Nader
Journal:  Intern Emerg Med       Date:  2011-12-07       Impact factor: 3.397

2.  Fluid responsiveness predicted by noninvasive bioreactance-based passive leg raise test.

Authors:  Brahim Benomar; Alexandre Ouattara; Philippe Estagnasie; Alain Brusset; Pierre Squara
Journal:  Intensive Care Med       Date:  2010-07-28       Impact factor: 17.440

3.  The effect of head up tilting on bioreactance cardiac output and stroke volume readings using suprasternal transcutaneous Doppler as a control in healthy young adults.

Authors:  Jie Zhang; Lester A H Critchley; Daniel C W Lee; Kim S Khaw; Shara W Y Lee
Journal:  J Clin Monit Comput       Date:  2016-02-12       Impact factor: 2.502

4.  Impact of dexmedetomidine versus propofol on cardiac function of children undergoing laparoscopic surgery.

Authors:  Hongbin Gu; Jinfen Liu; Chi Wu
Journal:  Int J Clin Exp Med       Date:  2014-12-15

5.  Applications of minimally invasive cardiac output monitors.

Authors:  Jahan Porhomayon; Gino Zadeii; Samuel Congello; Nader D Nader
Journal:  Int J Emerg Med       Date:  2012-04-24

6.  Peak cardiac power measured noninvasively with a bioreactance technique is a predictor of adverse outcomes in patients with advanced heart failure.

Authors:  Hannah Rosenblum; Stephen Helmke; Paula Williams; Sergio Teruya; Margaret Jones; Daniel Burkhoff; Donna Mancini; Mathew S Maurer
Journal:  Congest Heart Fail       Date:  2010-10-19

Review 7.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

8.  Noninvasive continuous cardiac output and cerebral perfusion monitoring in term infants with neonatal encephalopathy: assessment of feasibility and reliability.

Authors:  Eva Forman; Colm R Breatnach; Stephanie Ryan; Jana Semberova; Jan Miletin; Adrienne Foran; Afif El-Khuffash
Journal:  Pediatr Res       Date:  2017-08-02       Impact factor: 3.756

9.  The passive leg raise test to predict fluid responsiveness in children--preliminary observations.

Authors:  Guo-ping Lu; Gangfeng Yan; Yang Chen; Zhu-jin Lu; Lin-en Zhang; Niranjan Kissoon
Journal:  Indian J Pediatr       Date:  2013-12-11       Impact factor: 1.967

10.  Comparison of monitoring performance of Bioreactance vs. pulse contour during lung recruitment maneuvers.

Authors:  Pierre Squara; Dominique Rotcajg; Dominique Denjean; Philippe Estagnasie; Alain Brusset
Journal:  Crit Care       Date:  2009-07-28       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.