Literature DB >> 22265900

Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in critically ill patients.

X Monnet1, N Anguel, M Jozwiak, C Richard, J-L Teboul.   

Abstract

BACKGROUND: The ability of the third-generation FloTrac/Vigileo software to track changes in cardiac index (CI) induced by volume expansion and norepinephrine in critically ill patients is unknown.
METHODS: In subjects with circulatory failure, we administered volume expansion (20 subjects) and increased (20 subjects) or decreased (20 subjects) the dose of norepinephrine. We measured arterial pressure waveform-derived CI provided by the third-generation FloTrac/Vigileo device (CI(pw)) and transpulmonary thermodilution CI (CI(td)) before and after therapeutic interventions.
RESULTS: Considering the pairs of measurements performed before and after all therapeutic interventions (n=60), a bias between the absolute values of CI(pw) and CI(td) was 0.26 (0.94) litre min(-1) m(-2) and the percentage error was 54%. Changes in CI(pw) tracked changes in CI(td) induced by volume expansion with moderate accuracy [n=20, bias=-0.11 (0.54) litre min(-1) m(-2), r(2)=0.26, P=0.02]. When changes in CI(td) were induced by norepinephrine (n=40), a bias between CI(pw) and CI(td) was 0.01 (0.41) litre min(-1) m(-2) (r(2)=0.11, P=0.04). The concordance rates between changes in CI(pw) and CI(td) induced by volume expansion and norepinephrine were 73% and 60%, respectively. The bias between changes in CI(pw) and CI(td) significantly correlated with changes in total systemic vascular resistance (r(2)=0.41, P<0.0001).
CONCLUSIONS: The third-generation FloTrac/Vigileo device was moderately reliable for tracking changes in CI induced by volume expansion and poorly reliable for tracking changes in CI induced by norepinephrine.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22265900     DOI: 10.1093/bja/aer491

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  30 in total

1.  The impact of systemic vascular resistance on the accuracy of the FloTrac/Vigileo™ system in the perioperative period of cardiac surgery: a prospective observational comparison study.

Authors:  Yohei Sotomi; Katsuomi Iwakura; Yoshiharu Higuchi; Kazuo Abe; Junko Yoshida; Takafumi Masai; Kenshi Fujii
Journal:  J Clin Monit Comput       Date:  2013-06-08       Impact factor: 2.502

2.  Agreement between stroke volume measured by oesophageal Doppler and uncalibrated pulse contour analysis during fluid loads in severe aortic stenosis.

Authors:  Lars Øivind Høiseth; Ingrid Elise Hoff; Ove Andreas Hagen; Svein Aslak Landsverk; Knut Arvid Kirkebøen
Journal:  J Clin Monit Comput       Date:  2015-02-01       Impact factor: 2.502

3.  Tissue microcirculation measured by vascular occlusion test during anesthesia induction.

Authors:  Tae Kyong Kim; Youn Joung Cho; Jeong Jin Min; John M Murkin; Jae-Hyon Bahk; Deok Man Hong; Yunseok Jeon
Journal:  J Clin Monit Comput       Date:  2015-03-08       Impact factor: 2.502

4.  Cardiac output monitoring in septic shock: evaluation of the third-generation Flotrac-Vigileo.

Authors:  Sophie Marqué; Antoine Gros; Loic Chimot; Arnaud Gacouin; Sylvain Lavoué; Christophe Camus; Yves Le Tulzo
Journal:  J Clin Monit Comput       Date:  2013-01-30       Impact factor: 2.502

Review 5.  Newer methods of cardiac output monitoring.

Authors:  Yatin Mehta; Dheeraj Arora
Journal:  World J Cardiol       Date:  2014-09-26

Review 6.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

Review 7.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

8.  Evaluation of the use of the fourth version FloTrac system in cardiac output measurement before and after cardiopulmonary bypass.

Authors:  Sheng-Yi Lin; An-Hsun Chou; Yung-Fong Tsai; Su-Wei Chang; Min-Wen Yang; Pei-Chi Ting; Chun-Yu Chen
Journal:  J Clin Monit Comput       Date:  2017-10-16       Impact factor: 2.502

9.  Assessment of method agreement between two minimally invasive hemodynamic measurements in septic shock patients on high doses of vasopressor drugs. A preliminary study.

Authors:  Oana Antal; Mihai Mărginean; Natalia Hagău
Journal:  Rom J Anaesth Intensive Care       Date:  2017-10

10.  Stroke volume determination using transcardiopulmonary thermodilution and arterial pulse contour analysis in severe aortic valve disease.

Authors:  Martin Petzoldt; Carsten Riedel; Jan Braeunig; Sebastian Haas; Matthias S Goepfert; Hendrik Treede; Stephan Baldus; Alwin E Goetz; Daniel A Reuter
Journal:  Intensive Care Med       Date:  2013-01-04       Impact factor: 17.440