X Monnet1, N Anguel, M Jozwiak, C Richard, J-L Teboul. 1. Hôpitaux universitaires Paris-Sud, Hôpital de Bicêtre, service de réanimation médicale, 78, rue du Général Leclerc, Le Kremlin-Bicêtre F-94270, France. xavier.monnet@bct.aphp.fr
Abstract
BACKGROUND: The ability of the third-generation FloTrac/Vigileo software to track changes in cardiac index (CI) induced by volume expansion and norepinephrine in critically ill patients is unknown. METHODS: In subjects with circulatory failure, we administered volume expansion (20 subjects) and increased (20 subjects) or decreased (20 subjects) the dose of norepinephrine. We measured arterial pressure waveform-derived CI provided by the third-generation FloTrac/Vigileo device (CI(pw)) and transpulmonary thermodilution CI (CI(td)) before and after therapeutic interventions. RESULTS: Considering the pairs of measurements performed before and after all therapeutic interventions (n=60), a bias between the absolute values of CI(pw) and CI(td) was 0.26 (0.94) litre min(-1) m(-2) and the percentage error was 54%. Changes in CI(pw) tracked changes in CI(td) induced by volume expansion with moderate accuracy [n=20, bias=-0.11 (0.54) litre min(-1) m(-2), r(2)=0.26, P=0.02]. When changes in CI(td) were induced by norepinephrine (n=40), a bias between CI(pw) and CI(td) was 0.01 (0.41) litre min(-1) m(-2) (r(2)=0.11, P=0.04). The concordance rates between changes in CI(pw) and CI(td) induced by volume expansion and norepinephrine were 73% and 60%, respectively. The bias between changes in CI(pw) and CI(td) significantly correlated with changes in total systemic vascular resistance (r(2)=0.41, P<0.0001). CONCLUSIONS: The third-generation FloTrac/Vigileo device was moderately reliable for tracking changes in CI induced by volume expansion and poorly reliable for tracking changes in CI induced by norepinephrine.
BACKGROUND: The ability of the third-generation FloTrac/Vigileo software to track changes in cardiac index (CI) induced by volume expansion and norepinephrine in critically illpatients is unknown. METHODS: In subjects with circulatory failure, we administered volume expansion (20 subjects) and increased (20 subjects) or decreased (20 subjects) the dose of norepinephrine. We measured arterial pressure waveform-derived CI provided by the third-generation FloTrac/Vigileo device (CI(pw)) and transpulmonary thermodilution CI (CI(td)) before and after therapeutic interventions. RESULTS: Considering the pairs of measurements performed before and after all therapeutic interventions (n=60), a bias between the absolute values of CI(pw) and CI(td) was 0.26 (0.94) litre min(-1) m(-2) and the percentage error was 54%. Changes in CI(pw) tracked changes in CI(td) induced by volume expansion with moderate accuracy [n=20, bias=-0.11 (0.54) litre min(-1) m(-2), r(2)=0.26, P=0.02]. When changes in CI(td) were induced by norepinephrine (n=40), a bias between CI(pw) and CI(td) was 0.01 (0.41) litre min(-1) m(-2) (r(2)=0.11, P=0.04). The concordance rates between changes in CI(pw) and CI(td) induced by volume expansion and norepinephrine were 73% and 60%, respectively. The bias between changes in CI(pw) and CI(td) significantly correlated with changes in total systemic vascular resistance (r(2)=0.41, P<0.0001). CONCLUSIONS: The third-generation FloTrac/Vigileo device was moderately reliable for tracking changes in CI induced by volume expansion and poorly reliable for tracking changes in CI induced by norepinephrine.
Authors: Martin Petzoldt; Carsten Riedel; Jan Braeunig; Sebastian Haas; Matthias S Goepfert; Hendrik Treede; Stephan Baldus; Alwin E Goetz; Daniel A Reuter Journal: Intensive Care Med Date: 2013-01-04 Impact factor: 17.440