Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care.

B Bataille1, M Bertuit, M Mora, M Mazerolles, P Cocquet, B Masson, P E Moussot, J Ginot, S Silva, J Larché.   

Abstract

BACKGROUND: The esCCO monitor (ECG- estimated Continuous Cardiac Output, Nihon Kohden(®)) is a new non-invasive tool for estimating cardiac output (CO). It derives CO from the pulse wave transit time (PWTT) estimated by the ECG and the plethysmographic wave. An initial calibration is needed to refine the relation linking pulse pressure (measured by arterial pressure cuff) to PWTT. To assess the accuracy and reliability of the esCCO system, we performed an analysis of agreement of CO values obtained by transthoracic echocardiography (TTE).
METHODS: Thirty-eight intensive care unit patients were prospectively included. CO was determined simultaneously using esCCO (CO(esCCO)) and TTE (CO(TTE)) as our reference method.
RESULTS: A total of 103 paired readings from 38 patients were collected. The correlation coefficient between CO(esCCO) and CO(TTE) was 0.61 (P<0.001). The Bland and Altman analysis corrected for repeated measures showed a bias of -1.6 litre min(-1) and limits of agreement from -4.7 to +1.5 litre min(-1), with a percentage error (2 sd/mean CO) of 49%. The correlation for CO changes was significant (R=0.63, P<0.001), but the concordance rate was poor (73%). Polar plot analysis showed an angular bias of -9° with radial limits of agreement from -54° to +36°. The bias appeared to correlate with systemic vascular resistance (R=-0.45, P<0.001).
CONCLUSIONS: In critically ill patients, the performance of the esCCO monitor was not clinically acceptable, and this monitor cannot be recommended in this setting. Moreover, the esCCO failed to trend CO data reliably.

Entities:  

Mesh:

Year:  2012        PMID: 22907340     DOI: 10.1093/bja/aes298

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  12 in total

1.  Comparison of the ability of two continuous cardiac output monitors to measure trends in cardiac output: estimated continuous cardiac output measured by modified pulse wave transit time and an arterial pulse contour-based cardiac output device.

Authors:  Takashi Terada; Ayano Oiwa; Yumi Maemura; Samuna Robert; Sayaka Kessoku; Ryoichi Ochiai
Journal:  J Clin Monit Comput       Date:  2015-09-14       Impact factor: 2.502

2.  Validation of electrical velocimetry in resuscitation of patients undergoing liver transplantation. Observational study.

Authors:  Ahmed M Mukhtar; Mohamed Elayashy; Amr H Sayed; Gihan M Obaya; Akram A Eladawy; Mai A Ali; Hisham M Dahab; Dina Z Khalaf; Mostafa A Mohamed; Amr H Elfouly; Gad M Behairy; Amr A Abdelaal
Journal:  J Clin Monit Comput       Date:  2019-04-19       Impact factor: 2.502

Review 3.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

4.  Estimated continuous cardiac output based on pulse wave transit time in off-pump coronary artery bypass grafting: a comparison with transpulmonary thermodilution.

Authors:  Alexey A Smetkin; Ayyaz Hussain; Evgenia V Fot; Viktor I Zakharov; Natalia N Izotova; Angelika S Yudina; Zinaida A Dityateva; Yanina V Gromova; Vsevolod V Kuzkov; Lars J Bjertnæs; Mikhail Y Kirov
Journal:  J Clin Monit Comput       Date:  2016-03-07       Impact factor: 2.502

5.  Continuous Estimation of Cardiac Output in Critical Care: A Noninvasive Method Based on Pulse Wave Transit Time Compared with Transpulmonary Thermodilution.

Authors:  Ulrike Ehlers; Rolf Erlebach; Giovanna Brandi; Federica Stretti; Richard Valek; Stephanie Klinzing; Reto Schuepbach
Journal:  Crit Care Res Pract       Date:  2020-07-20

6.  Japanese Acupuncture: A Complementary Approach to the Meridian Balance Method.

Authors:  Benoît Bataille; Philipp Mitariu
Journal:  Med Acupunct       Date:  2020-08-13

7.  Pulse Wave Transit Time Measurements of Cardiac Output in Septic Shock Patients: A Comparison of the Estimated Continuous Cardiac Output System with Transthoracic Echocardiography.

Authors:  Marc Feissel; Ludwig Serge Aho; Stefan Georgiev; Romain Tapponnier; Julio Badie; Rémi Bruyère; Jean-Pierre Quenot
Journal:  PLoS One       Date:  2015-06-30       Impact factor: 3.240

8.  Pilot Study: Estimation of Stroke Volume and Cardiac Output from Pulse Wave Velocity.

Authors:  Yurie Obata; Maki Mizogami; Daniel Nyhan; Dan E Berkowitz; Jochen Steppan; Viachaslau Barodka
Journal:  PLoS One       Date:  2017-01-06       Impact factor: 3.240

9.  Less invasive methods of advanced hemodynamic monitoring: principles, devices, and their role in the perioperative hemodynamic optimization.

Authors:  Christos Chamos; Liana Vele; Mark Hamilton; Maurizio Cecconi
Journal:  Perioper Med (Lond)       Date:  2013-09-17

10.  Relationship between stroke volume and pulse pressure during blood volume perturbation: a mathematical analysis.

Authors:  Ramin Bighamian; Jin-Oh Hahn
Journal:  Biomed Res Int       Date:  2014-05-20       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.