Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present.

Kazuya Tachibana1, Hideaki Imanaka, Muneyuki Takeuchi, Yuji Takauchi, Hiroshi Miyano, Masaji Nishimura.   

Abstract

BACKGROUND: Although evaluation of cardiac output by the partial carbon dioxide rebreathing technique is as accurate as thermodilution techniques under controlled mechanical ventilation, it is less accurate at low tidal volume. It is not clear whether reduced accuracy is due to low tidal volume or low minute ventilation. The effect of spontaneous breathing on the accuracy of partial carbon dioxide rebreathing measurement has not been fully investigated. The objectives of the current study were to investigate whether tidal volume or minute ventilation is the dominant factor for the accuracy, and the accuracy of the technique when spontaneous breathing effort is present.
METHODS: The authors enrolled 25 post-cardiac surgery patients in two serial protocols. First, the authors applied three settings of controlled mechanical ventilation in random order: large tidal volume (12 ml/kg), the same minute ventilation with a small tidal volume (6 ml/kg), and 50% decreased minute ventilation with a small tidal volume (6 ml/kg). Second, when the patient recovered spontaneous breathing, the authors applied three conditions of partial ventilatory support in random order: synchronized intermittent mandatory ventilation-pressure support ventilation, pressure support ventilation with an appropriately adjusted rebreathing loop, and pressure support ventilation with the shortest available loop. After establishing steady state conditions, the authors measured cardiac output using both partial carbon dioxide rebreathing and thermodilution methods. The correlation between the data yielded by the two methods was determined by Bland-Altman analysis and linear regression.
RESULTS: Cardiac output with the carbon dioxide rebreathing technique correlated moderately with that measured by thermodilution when minute ventilation was set to maintain normocapnia, regardless of tidal volumes. However, when minute ventilation was set low, the carbon dioxide rebreathing technique underreported cardiac output (y = 0.70x; correlation coefficient, 0.34; bias, -1.73 l/min; precision, 1.27 l/min; limits of agreement, -4.27 to +0.81 l/min). When there was spontaneous breathing, the correlation between the two cardiac output measurements became worse. Carbon dioxide rebreathing increased spontaneous tidal volume and respiratory rate (20% and 30%, respectively, during pressure support ventilation) when the rebreathing loop was adjusted for large tidal volume.
CONCLUSIONS: During controlled mechanical ventilation, minute ventilation rather than tidal volume affected the accuracy of cardiac output measurement using the partial carbon dioxide rebreathing technique. When spontaneous breathing is present, the carbon dioxide rebreathing technique is less accurate and increases spontaneous tidal volume and respiratory rate.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12657842     DOI: 10.1097/00000542-200304000-00007

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  10 in total

Review 1.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

Review 2.  Minimally invasive monitoring of cardiac output in the cardiac surgery intensive care unit.

Authors:  Jamal A Alhashemi; Maurizio Cecconi; Giorgio della Rocca; Maxime Cannesson; Christoph K Hofer
Journal:  Curr Heart Fail Rep       Date:  2010-09

3.  The Inspired Sinewave Technique: A Comparison Study With Body Plethysmography in Healthy Volunteers.

Authors:  Phi Anh Phan; Cathy Zhang; Daniel Geer; Federico Formenti; Clive E W Hahn; Andrew D Farmery
Journal:  IEEE J Transl Eng Health Med       Date:  2017-10-27       Impact factor: 3.316

4.  Improved accuracy of cardiac output estimation by the partial CO2 rebreathing method.

Authors:  Yoshifumi Kotake; Takashige Yamada; Hiromasa Nagata; Takeshi Suzuki; Ryohei Serita; Nobuyuki Katori; Junzo Takeda; Hideyuki Shimizu
Journal:  J Clin Monit Comput       Date:  2009-03-20       Impact factor: 2.502

5.  The dynamic pattern of end-tidal carbon dioxide during cardiopulmonary resuscitation: difference between asphyxial cardiac arrest and ventricular fibrillation/pulseless ventricular tachycardia cardiac arrest.

Authors:  Katja Lah; Miljenko Križmarić; Stefek Grmec
Journal:  Crit Care       Date:  2011-01-11       Impact factor: 9.097

Review 6.  Cardiac output monitoring: an integrative perspective.

Authors:  Jamal A Alhashemi; Maurizio Cecconi; Christoph K Hofer
Journal:  Crit Care       Date:  2011-03-22       Impact factor: 9.097

7.  Effects of reduced rebreathing time, in spontaneously breathing patients, on respiratory effort and accuracy in cardiac output measurement when using a partial carbon dioxide rebreathing technique: a prospective observational study.

Authors:  Kazuya Tachibana; Hideaki Imanaka; Muneyuki Takeuchi; Tomoyo Nishida; Yuji Takauchi; Masaji Nishimura
Journal:  Crit Care       Date:  2005-09-07       Impact factor: 9.097

Review 8.  Perioperative cardiovascular monitoring of high-risk patients: a consensus of 12.

Authors:  Jean-Louis Vincent; Paolo Pelosi; Rupert Pearse; Didier Payen; Azriel Perel; Andreas Hoeft; Stefano Romagnoli; V Marco Ranieri; Carole Ichai; Patrice Forget; Giorgio Della Rocca; Andrew Rhodes
Journal:  Crit Care       Date:  2015-05-08       Impact factor: 9.097

9.  Invasive and noninvasive cardiovascular monitoring options for cardiac surgery.

Authors:  Dominic P Recco; Nathalie Roy; Alexander J Gregory; Kevin W Lobdell
Journal:  JTCVS Open       Date:  2022-04-11

10.  Less invasive methods of advanced hemodynamic monitoring: principles, devices, and their role in the perioperative hemodynamic optimization.

Authors:  Christos Chamos; Liana Vele; Mark Hamilton; Maurizio Cecconi
Journal:  Perioper Med (Lond)       Date:  2013-09-17
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.