Literature DB >> 22324797

A comparison of the Nexfin® and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgery.

O Broch1, J Renner, M Gruenewald, P Meybohm, J Schöttler, A Caliebe, M Steinfath, M Malbrain, B Bein.   

Abstract

The newly introduced Nexfin(®) device allows analysis of the blood pressure trace produced by a non-invasive finger cuff. We compared the cardiac output derived from the Nexfin and PiCCO, using transcardiopulmonary thermodilution, during cardiac surgery. Forty patients with preserved left ventricular function undergoing elective coronary artery bypass graft surgery were studied after induction of general anaesthesia and until discharge to the intensive care unit. There was a significant correlation between Nexfin and PiCCO before (r(2) = 0.81, p < 0.001) and after (r(2) = 0.56, p < 0.001) cardiopulmonary bypass. Bland-Altman analysis demonstrated the mean bias of Nexfin to be -0.1 (95% limits of agreement -0.6 to +0.5, percentage error 23%) and -0.1 (-0.8 to +0.6, 26%) l.min(-1).m(-2), before and after cardiopulmonary bypass, respectively. After a passive leg-raise was performed, there was also good correlation between the two methods, both before (r(2) = 0.72, p < 0.001) and after (r(2) = 0.76, p < 0.001) cardiopulmonary bypass. We conclude that the Nexfin is a reliable method of measuring cardiac output during and after cardiac surgery. Anaesthesia
© 2012 The Association of Anaesthetists of Great Britain and Ireland.

Entities:  

Mesh:

Year:  2012        PMID: 22324797     DOI: 10.1111/j.1365-2044.2011.07018.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  47 in total

1.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

2.  Reliability of a new 4th generation FloTrac algorithm to track cardiac output changes in patients receiving phenylephrine.

Authors:  Fuhai Ji; Jian Li; Neal Fleming; David Rose; Hong Liu
Journal:  J Clin Monit Comput       Date:  2014-09-30       Impact factor: 2.502

3.  Influence of menopause status and age on integrated central and peripheral hemodynamic responses to subsystolic cuffing during submaximal exercise.

Authors:  Erik H Van Iterson; Courtney Gramm; Nicholas R Randall; Thomas P Olson
Journal:  Am J Physiol Heart Circ Physiol       Date:  2016-10-07       Impact factor: 4.733

Review 4.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

Review 5.  Non-invasive continuous blood pressure monitoring: a review of current applications.

Authors:  Elena Chung; Guo Chen; Brenton Alexander; Maxime Cannesson
Journal:  Front Med       Date:  2013-01-23       Impact factor: 4.592

6.  Interleukin-1 Blockade in Recently Decompensated Systolic Heart Failure: Results From REDHART (Recently Decompensated Heart Failure Anakinra Response Trial).

Authors:  Benjamin W Van Tassell; Justin Canada; Salvatore Carbone; Cory Trankle; Leo Buckley; Claudia Oddi Erdle; Nayef A Abouzaki; Dave Dixon; Dinesh Kadariya; Sanah Christopher; Aaron Schatz; Jessica Regan; Michele Viscusi; Marco Del Buono; Ryan Melchior; Pranav Mankad; Juan Lu; Robin Sculthorpe; Giuseppe Biondi-Zoccai; Edward Lesnefsky; Ross Arena; Antonio Abbate
Journal:  Circ Heart Fail       Date:  2017-11       Impact factor: 8.790

7.  Less invasive hemodynamic monitoring in critically ill patients.

Authors:  Jean-Louis Teboul; Bernd Saugel; Maurizio Cecconi; Daniel De Backer; Christoph K Hofer; Xavier Monnet; Azriel Perel; Michael R Pinsky; Daniel A Reuter; Andrew Rhodes; Pierre Squara; Jean-Louis Vincent; Thomas W Scheeren
Journal:  Intensive Care Med       Date:  2016-05-07       Impact factor: 17.440

Review 8.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

Review 9.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

10.  An autocalibrating algorithm for non-invasive cardiac output determination based on the analysis of an arterial pressure waveform recorded with radial artery applanation tonometry: a proof of concept pilot analysis.

Authors:  Bernd Saugel; Agnes S Meidert; Nicolas Langwieser; Julia Y Wagner; Florian Fassio; Alexander Hapfelmeier; Luisa M Prechtl; Wolfgang Huber; Roland M Schmid; Oliver Gödje
Journal:  J Clin Monit Comput       Date:  2013-12-10       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.