Comparison of anterior segment measurements with LenStar and Pentacam in patients with newly diagnosed glaucoma.

Emine Sen1, Merve Inanc2, Ufuk Elgin2, Pelin Yilmazbas2.   

Abstract

PURPOSE: To compare anterior segment measurements obtained using the Pentacam; Oculus, HR and the LenStar LS 900 in patients with newly diagnosed glaucoma.
METHODS: Patients with ocular hypertension and primary open-angle glaucoma who had been treated with PGA were included in the study. Anterior segment measurements including central corneal thickness (CCT), keratometry, anterior chamber depth (ACD) and white-to-white (WTW) corneal diameter obtained with the optic low-coherent reflectometer (LenStar LS-900, Haag-Streit AG, Switzerland) and with the Scheimpflug system (Pentacam; Oculus, HR) were compared. In order to compare LenStar and Pentacam measurements, paired sample t test and 'Bland-Altman' plot were used in the statistical analysis.
RESULTS: Twenty-two female (59.5%) and 15 male (40.5%) totally 37 cases with newly diagnosed glaucoma were included in the study. Anterior segment parameter measurements obtained with both the LenStar and the Pentacam were significantly correlated for right and left eyes, so the right eye values were used in statistical analysis. WTW and ACD values measured with LenStar were statistically significantly higher than those measured with Pentacam (p: 0.0001, p: 0.0001, respectively). There was no statistically significant difference between the CCT values measured by the two devices (p: 0.217).
CONCLUSION: There was a statistically significant difference between the mean values of ACD and WTW measured with the LenStar and Pentacam. These biometric devices should not be used interchangeably. It should be appropriate to use the same device on follow-up of glaucoma patients.

Entities:  

Keywords:  Anterior segment; Biometry; Glaucoma; LenStar; Pentacam; Scheimpflug

Mesh:

Year:  2017        PMID: 28108902     DOI: 10.1007/s10792-016-0440-z

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  16 in total

1.  Anterior chamber depth measurement as a screening tool for primary angle-closure glaucoma in an East Asian population.

Authors:  J G Devereux; P J Foster; J Baasanhu; D Uranchimeg; P S Lee; T Erdenbeleig; D Machin; G J Johnson; P H Alsbirk
Journal:  Arch Ophthalmol       Date:  2000-02

2.  Measurement of accommodation after implantation of an accommodating posterior chamber intraocular lens.

Authors:  Achim Langenbucher; Stefan Huber; Nhung X Nguyen; Berthold Seitz; Gabriele C Gusek-Schneider; Michael Küchle
Journal:  J Cataract Refract Surg       Date:  2003-04       Impact factor: 3.351

3.  Comparison of central corneal thickness and anterior chamber depth measured using LenStar LS900, Pentacam, and Visante AS-OCT.

Authors:  Clare OʼDonnell; Andreas Hartwig; Hema Radhakrishnan
Journal:  Cornea       Date:  2012-09       Impact factor: 2.651

4.  Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography.

Authors:  Haitao Li; Christopher Kai Shun Leung; Lee Wong; Carol Yim Lui Cheung; Chi Pui Pang; Robert Neal Weinreb; Dennis Shun Chiu Lam
Journal:  Ophthalmology       Date:  2007-10-04       Impact factor: 12.079

Review 5.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma.

Authors:  Mae O Gordon; Julia A Beiser; James D Brandt; Dale K Heuer; Eve J Higginbotham; Chris A Johnson; John L Keltner; J Philip Miller; Richard K Parrish; M Roy Wilson; Michael A Kass
Journal:  Arch Ophthalmol       Date:  2002-06

8.  Central corneal thickness as a risk factor for advanced glaucoma damage.

Authors:  Leon W Herndon; Jennifer S Weizer; Sandra S Stinnett
Journal:  Arch Ophthalmol       Date:  2004-01

9.  Changes in ocular biometry and anterior chamber parameters after pharmacologic mydriasis and peripheral iridotomy in primary angle closure suspects.

Authors:  Mohammad Reza Razeghinejad; Hamid Lashkarizadeh; Mohammad Hossein Nowroozzadeh; Mohammad Yazdanmehr
Journal:  J Optom       Date:  2016-02-28

10.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

View more
  1 in total

1.  Agreement analysis of Lenstar with other four techniques of biometry before cataract surgery.

Authors:  Jinling Zhang; Yong Wang
Journal:  Int Ophthalmol       Date:  2022-05-11       Impact factor: 2.029

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.